  • Make sure supplies are adequate and equipment and instruments are operational and notify Lab Officer of any problems.
  • Calibrate Pycnometer and all tracks.
  • Calibrate Mettler Toledo balances: Balance calibration should be carried out while the ship is in port and is as stable as possible.
  • Report to LO/ALO for other port call duties.

  • Provide training to all Physical Property scientists on instrument operation (WRMSL, SHMSL, ThermCon, Gantry, MAD, NGR), sample preparations, and data upload/download procedures
  • Provide safety tour to all scientists working in the Physical Property area. Ensure scientists understand all safety concern and sign document required

Site Preparation

  • Place log sheets and instrument's Quick Users Guide at each station. Organize prior Expedition's log sheets (if not already done so) for shipment back to IODP.
  • Reset all run numbers prior to calibrations; this is not necessary, but it is a good idea to make new instrument software log files and rename the old ones w/ the prior Expedition's number in the name.

  • Make sure that the motion control systems have found home and the position of the laser for all instruments using M-Drive.
  • Perform calibrations for all tracks prior to first core on deck at each Site.

...