Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Start the Velocity application by clicking the icon (Velocity 2.0.5) on the desktop.
  2. Once the application is launched, click the Make a Measurement button.
  3. In the Scan Sample Label dialog box, place the cursor in the Scanner String field (Figure 25).



Image Removed
Figure 25. Scan Sample Dialog Box.

4. Scan the barcode on the section half label to populate the fields on the Scan Sample Label dialog box. Enter Operator (user last name) and confirm measuring Station and Mode.

Stations Bayonet Y and Bayonet Z are for soft sediments. To use a Bayonet measuring station, the Mode value must be Section w/liner (material is still in core liner). Note: The orientation assigned to the sample measurement depends on the Station selected.

5. For Section w/liner mode, a Position Sample dialog box opens (Figure 26). Confirm the sample position as measured by the laser (the laser at the end of the track measures the range to the end of the core and calculates the position in the liner based on an offset for that station). Click OK.



Image Added Section-half measurements, in the liner. (X-Axis)Section-half measurements, in the liner. (X-Axis)Discrete sample (MAD cubes) measurements. (Z-Axis)Discrete sample (MAD cubes) measurements. (Y-Axis)Discrete sample (MAD cubes) measurements. (X-Axis)Piece measurements taken out of the section-half. Consolidated material only. (Any-Axis)Section-half measurements, in the liner. (X-Axis)


Figure 26. Position Sample Dialog Box.

6. The bayonet transducers lower into the section half; ensure they are completely inserted into the sample. If contact between the transducers and sediment is poor, add distilled water around the transducers to improve contact with sediment.

7. On the JOG Bayonet screen (Figure 27), monitor the waveform and position the transducer to obtain the best trace. The plot displays the live signal with no stacking. A weak signal is fine, as long as a clean first arrival is present. Once the waveform is acceptable, click Continue to begin sample acquisition.


Figure 27. JOG Bayonet Y Screen.

8. During acquisition, the program adds multiple waveforms until the total (max peak) is greater than ±5 V. This process is repeated and the waveform is averaged.

Measuring Discrete Samples

...

  1. Start the Velocity application by clicking the icon (Velocity 2.0.5) on the desktop.
  2. Once the application is launched, click the Make a Measurement button.
  3. In the Scan Sample Label dialog box, place the cursor in the Scanner String field (Figure 28).

Image Added


Figure 28. Scan Sample Label Dialog Box.

4. Scan the barcode on the discrete sample label to populate the fields on the Scan Sample Label screen. Enter Operator (user last name) and confirm measuring Station (Caliper) and Mode (Discrete).

5. Enter information requested in the Additional Discrete Information dialog box (Figure 29): Measurement Axis: X, Y, or Z.


Image Removed

Figure 29. Additional Discrete Information Dialog Box.

6. If the material to be measured is a whole piece removed from the liner (rarely used), select Piece w/o Liner mode in Step 4 and complete the offset (position of the piece in the liner before it was removed) in the Additional Piece Information dialog box (Figure 30).

If the orientation of the piece is unknown, check the Orientation Unknown box; otherwise the measurement is assigned to the X-direction.

Figure 30. Additional Piece Information Dialog Box.

7. Place the discrete sample in the caliper and add distilled water to create a good sample-transducer contact if needed.

8. On either the Additional Discrete Information (discrete samples) or Additional Piece Information (piece samples) box, select OK - AutoClose (for hard samples) or OK – Manual Close (for softer/friable samples).

AutoClose automatically closes the caliper on the sample. The transducer moves to a predefined position or until the actuator stalls.

9. On the JOG Caliper screen (Figure 31), monitor the waveform and position the transducer to obtain the best trace. The plot displays the live signal with no stacking. A weak signal is fine, as long as a clean first arrival is present. Once the sample is properly positioned, click Continue to begin sample acquisition.


Figure 31. JOG Caliper Screen.

10. During acquisition, the program adds multiple waveforms until the total (max peak) is greater than ±5 V. This process is repeated and the waveform is averaged.

The touch-screen Open/Close commands may be sluggish in response. If there is no response, try actuating the button using the mouse.

11. Repeat the measurement process for each axis to be measured.

Accepting Results and Manual Pick

...


Figure 33.A Manual Pick Screen. 

2. Click Continue to accept the manual pick and save it to the data file or Cancel to discard the manual pick results.

Data Handling

From the Display Results screen one of 4 data handling choices can be selected:

...

The following are common problems encountered when using the P-wave Velocity Gantry and their possible causes and solutions. For information about the laser Measurement and Operation software, see the Vp GIESA Gantry Laser Operation manual.


Issue

Possible Causes

Solution

Laser sensor: no laser light/no laser range data

Sampling is turned off

Turn sampling on through laser Measurement and Operation software

 

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

Laser sensor: Serial port not responding

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

 

Baud rate incorrect or unknown

See Acuity manual 3.5.1; laser Measurement and Operation software

Laser sensor: Error code (Exx) transmitted on serial port

 

See Acuity manual 5.3; laser Measurement and Operation software

Exlar actuator: No response

I/O error

Check drive or Expert software for faults

Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

 

Power interruption

Check wiring

Exlar actuator: Behaving erratically

Drive improperly tuned

Check gain settings

Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

 

Too much load on motor

Check for load irregularity or excess load

Exlar actuator: Cannot move load

Too much load or friction

Check load and friction sources 

Excessive side load

Check side load

 

Misalignment of output rod to load

Check alignment of rod and load

 

Current limit on drive too low

Check current on drive

 

Power supply current too low

Check power supply current

Exlar actuator: Housing vibrates when shaft in motion

Loose mounting

Tighten mounting screws

 

Drive improperly tuned

Check gain settings

Exlar actuator: Output rod rotates during motion

Rod rotation prevents proper linear motion

Install anti-rotation assembly

Exlar actuator: Overheating

Ambient temperature too high

Check ambient temperature

 

Actuator operation outside continuous ratings

Check operation settings 

Amplifier poorly tuned

Check gain settings

 Actuator Utilities

The actuator utilities can be used to test the caliper and bayonet actuators (move them up and down) and view the waveform detected by the transducers. There are also utilities to test the home position of the actuators and to check the laser function. This utility is useful when configuring the station offsets.

To change the behavior and/or force of the actuators, the Exlar utility must be used.

Open the Actuator utility (Figure 35) from Main > Actuator Utility and then select one of the JOG screens, Home All, or Read Laser.

Figure 35. Actuator Utilities.

...