Analytical Gas Monitoring System:User Guide


User Guide Information




Author(s):

Chris Bennight

Revision:

D. Houpt

Editor(s):

L. Peters

Manager Approval (Name, Title, Date):

D.J. Houpt

Audience:

Scientists, Laboratory Technicians

Origination date:

8/10/2010

Current version:

V371T

July 2017

Revised:

July 2017

Domain:

Chemistry


User Guide Contents




Topic

See page…

Introduction

Quick Start

Hardware

Software

Data Reporting Services

...


...

Introduction


The analytical gas monitoring system is a collection of hardware, software, and infrastructure designed to record, report, and alert interested parties in the current and historical status of various analytical gases. The monitoring system currently comprises 4 primary parts:

...

  • Hardware
  • Data Logging Software
  • Database/Web services AnchorRTF37383336353a203248656164RTF37383336353a203248656164
  • Reporting/Viewing software

...


Quick Start


Load the web page: {*}{_}http://web.ship.iodp.tamu.edu/tasapps/gasstatus_* on any computer on the ship to view the current status of gas supply (Figure 1).

...


Image Modified

...


Figure 1. Current gas status.


This screen provides a quick overview of the pressure (left and right banks) of each gas attached to the system, as well as an alarm setting. If the alarm button is checked, when the pressure of both banks falls below the threshold shown the alarm box an audible alert will sound. The alarm can be silenced by unchecking the alarm box.
This screen can be left minimized on any computer on the ship and will alert when a low pressure situation occurs. When the ICP is run the software is typically left running on the ICP computer.

...

...


Hardware


This system comprises four pieces of hardware purchased from DataQ Instruments ({*}{_}http://_*www.dataq.com).


Part Number

Description

Quantity

DI-718B-E

Ethernet data logger/data acquisition system

1

DI-8B38-05

Isolated signal conditioning module, strain, 10 VDC, 2 mV/V

3

FSH00827

Female port pressure sensor (PFP350 Series) 0-3000psi

3

FSH01785

Thirty foot cable with LEMO connector.

3


DI-718B-E Ethernet data logger


The data acquisition system is an input-conditioned intrinsically safe Ethernet-based data logger (Figure 2) and data acquisition board (Figure 3). It has 8 input channels, of which we currently use 3 (Figure 4). The device collects data from the module installed in each of these inputs and sends it over an Ethernet network to client software (either WinDaq or client software written using the DataQ SDK).
The device is currently installed in the lower pallet stores, against the back wall (Figure 5).
The Mac address is 00:29:4A:C4:59:E8 (Figure 6). It connects via Ethernet to port 39 in the pallet stores (Figure 7).
Image Modified

...

Figure 2. Data logger.
Image Modified

...


Figure 3. Data acquisition board.
Image Modified

...


...

Figure 4. Three acquisition channels in use.
Image Modified

...

Figure 5. Data acquisition device installed against the back wall in pallet stores.
Image Modified

...


...

Figure 6. Data acquisition device labeling.
Image Modified

...

Figure 7. Ethernet connection to acquisition device.


8B38 Isolated signal conditioning module


Three of these signal conditioning modules are currently installed in the data logger (Figure 8). They provide electrical isolation and signal conditioning necessary to convert the output of the pressure sensors to a 0–5 V final output range.
One of these units is required for each sensor connected to the data logger, up to a maximum of 8.

...


Image Modified

...


...

Figure 8. Isolated signal conditioning module.


FSH00827 0–3000 psi Futek pressure sensor

...

The pressure sensor measures pressure in the 0–3000 psi range (Figure 9). It is supplied an excitation of 10 V and has an LEMO connector. Connector pin out is listed on the side of the sensor, with an output positive and negative and excitation positive and negative.
Three of these sensors are currently installed, one on each (of 2) argon banks and one on the single helium bank. These require no maintenance or special considerations, other than taking care not to physically crush them when working near them.

...


Image Modified

...


...

Figure 9. 0–3000 psi pressure sensor.


30 foot LEMO cables


These 30 foot braided steel shielded cables have a LEMO connector on one end to plug directly into the sensor and 4 bare wire terminals on the other end to connect to the data logger (Figure 10).

...


Image Modified

...


Figure 10. LEMO connector and cable mounted on pressure sensor.

...

...


Software


Data Logging Software


The data logging portion of this system is composed of two parts:

...

  1. Start DataQ Instrument Hardware Manager.
  2. Once a logger has been detected (Figure 11), select the Start WinDaq button.
  3. WinDaq can be run in default mode (Figure 12).
  4. An ISO of the WinDaq CD is located on the ship developer box under /Volumes/Ozymandias/ConfigurationManagement/software/WinDaq/windaq_april_15_2010.iso
  5. Close the DataQ Instrument Hardware Manager screen.
  6. Once WinDaq is communicating with the data logger, start the Gas Data Collector software (Figure 13) ({*}{_}http://web.ship.iodp.tamu.edu/tasapps/gasmonitor_*).
  7. This requires the .NET 3.5 run time and WinDaq to be installed. Source code is located at {*}{_}https://shiptest.ship.iodp.tamu.edu/svn/jr/NET/GasMonitorUploader_*)
  8. Press Start, then minimize the application and leave running in the background.


...


Image Modified

...


Figure 11. DataQ Instrument Hardware Manager.

...


Image Modified

...


...

Figure 12. WinDaq showing 3 acquisition channels.

...


Image Modified

...


Figure 13. Gas Data Collector.

...

...


Database/Webservices


The data back end consists 2 parts:

...

  • A database table: LIMS.MONITOR (see below)
  • A set of Web services: resteasy-monitor




Column Name

Data Type

Nullable

MONITOR_ID

NUMBER

No

TIMELOGGED

TIMESTAMP(6)

Yes

EXPEDITION

VARCHAR2(20 BYTE)

Yes

DATA_GROUP

VARCHAR2(100 BYTE)

Yes

DATA_DURATION

VARCHAR2(100 BYTE)

Yes

DATA_INSTRUMENT

VARCHAR2(100 BYTE)

Yes

DATA_SENSOR

VARCHAR2(100 BYTE)

Yes

DATA_LOCATION

VARCHAR2(100 BYTE)

Yes

DATA_COMMENTS

VARCHAR2(4000 BYTE)

Yes

DATA_NAME

VARCHAR2(50 BYTE)

Yes

DATA_VALUE

VARCHAR2(50 BYTE)

Yes

DATA_UNITS

VARCHAR2(20 BYTE)

Yes

DATA_LOGGED_BY

VARCHAR2(50 BYTE)

Yes



The database table creation script is included in the source code for the resteasy-monitor project, located at {*}{_}https://shiptest.ship.iodp.tamu.edu/svn/jr/java/resteasy-monitor._*
The database table exists in the LIMS tablespace, and the services use the LIMS account to read/write from the table. No special user authentication is currently required to write from/read to these services. One sequence, LIMS.MONITOR_SEQUENCE, and one trigger, LIMS.MONITOR_TRIGGER exist as part of this package for the sole purpose of incrementing a unique ID (LIMS.MONITOR.MONITOR_ID ) on the table. The trigger also sets the timelogged field to the value of SYSDATE (current time in GMT per our configuration) when a new entry is created.
A .NET client for the Web services is included in the in GasDataCollector software. The client is located at {*}{_}https://shiptest.ship.iodp.tamu.edu/svn/jr/NET/GasMonitorUploader/gasmon/_*MonitorService.cs.
A Silverlight client for the services is included in the GasStatus Web page, and can be found at {*}{_}https://shiptest.ship.iodp.tamu.edu/svn/jr/Silverlight/GasStatus/trunk/gasstatus/MonitorService.cs._*

...


...



Data Reporting Services


GasStatus is the only current data reporting service for the system. Data is reported in one of two ways:

...

This display allows the user to view the actual pressure of any configured analytical gas. The check box near the Alarm label enables or disables the alarm's audible alert. Use the entry field under the Alarm label to set the pressure at which, when both banks decrease below this value, an alarm condition will exist. If enabled, an audible alarm will sound and the color of the status circle will change from green to red.


Image Modified

...

Figure 14. Overview/alarm display screen.


Gas details view


This view provides a detailed history of each gas bank. The current status is displayed for the selected gas, as well as a visual indicator (gauge). A history displays the history of gas pressures over time. Access the details view by clicking on the desired analytical gas in the top menu. This reporting page can be accessed at [{*}{_}http://web.ship.iodp.tamu.edu/tasapps/gasstatus._*


http://web.ship.iodp.tamu.edu/tasapps/gasstatus]The source code for this project is in version control located at {*}{_}https://shiptest.ship.iodp.tamu.edu/_*svn/jr/Silverlight/GasStatus


anchorRTF32343731373a204669675469


RTF32343731373a204669675469

Image Modified
Figure 15. Gas details view.