Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Place the aluminum standard between the transducers. Close the transducers. Aluminum standard width is 76.2mm. Select Laser Offset Correction. Confirm that the Distance matches the standard width.
  2. Enter the Aluminum Standard velocity. 6295.00 m/s. Select Determine System Delay. Verify that the Velocity matches the Aluminum Standard.
    Figure 9. PWave Calibration Utility. Note Liner Delay corrected.

    Turn on Liner Delay
  3. Open transducers. Remove Aluminum Standard. Position the Water Standard in place. Close transducers. Turn on Liner Delay. Liner Thickness is 0.275 cm. Determine Liner Velocity. Verify that the water velocity matches the calculated H2O velocity +/- 1% (liner velocity is calculated based on this).
  4. Open transducers. Accept Changes

Uploading data To Lims

Image Removed

To up load the images into the database MegaUploadaTron must be running in background. If not already started do the following:

  1. On the desktop click the MUT icon on the bottom task bar.
    Image Added

  2. The following will appear:

    Figure 10. LIMS Uploader

  3. Check Automatic Upload in the lower right hand corner and then click the minimize window and MUT will run in background.