1. | Load ThermCon software in offline mode. Ensure that the Text_ID field is blank. |
2. | Scan the core label using a scanner, then click Verify Sample. |
3. | If login is requested, enter UserName and Password and then click OK. |
4. | Run TK04 program and choose Measuring > Configuration. |
5. | Set configuration parameters as follows: –Root Name: 6 characters or less; suggest Core-Type-Section (no special characters). –Serial Number: number of repeat measurements at each point (1–99 single measurements). –Folder: path for saving data results. –Heating Power: for the VLQ (needle probe), set to twice the estimated thermal conductivity value of measured sediment. For example, 2–3 is good for sediment. –Measuring Time: set to at least 80 s (needle probe); set to 60 s for Puck. Click Expert Options to configure Drift Control and Pause in Minutes (see Step 6). |
6. | To configure Drift Control and Pause in Minutes in Expert Options: –Pause in Minutes: insert a pause between single measurements; recommended = 10 min. This parameter does not apply when conducting a single measurement on each core or each core section. |
...