TK04 Thermal Conductivity: User Guide



 

 

Author:

M. Hastedt, Y.-G. Kim, M.A. Kominz

Reviewer(s):

T. Gorgas, M. Vasilyev, R. Wilkens, K. Milliken, H. Barnes, S. Hermann; T. Cobb

Management Approval (Name, Title, Date):

D.J. Houpt (Supervisor of Analytical Systems)

Audience:

Scientists, Technicians

Origination date:

2/14/2009; 11/14/2009

Current version:

V378P | 372

Revised:

372|V371T|no change 03/18

Domain:

Physical Properties

System:

Thermal Conductivity

Keywords:

Full-space needle probe, half-space needle probe, TK04

...

Process

Time (min)

Comments

1. Obtain a whole-round core section from the core rack

0.3

See Preparing Samples for Analysis

 


2. Locate the appropriate probe for the sample type

0.5

3. Verify sample identification in software

0.5

See Configuring Measurement Program

 


4. Configure measurement program

0.3

5. Perform drift control

5

See Measuring Samples

 

 

 




6. Heat and measure sample

2

7. 10 minute pause between measurements

10

8. Repeat steps 5-7 for 2 additional measurements (3 total)

34

9. Upload results to LIMS

0.2

See Uploading Data to LIMS and Verifying Data in LIMS 

 



10. Check results in LIMS

1

11. Remove the section and deliver to splitting room

0.2

Total Time per sample:

54 (max)

 


Measuring the Standard

  1. Prior to initial testing of received cores at a site, the TK04 system should be tested and calibrated to ensure that there are no potential mechanical or software issues. Additional tests using the Standards should be run as part of the troubleshooting process if you experience issues during actual testing (See Troubleshooting).
  2. The Macor Standard for the Standard VLQ consists of its black holding shell, while the Macor Standard for both the Standard HLQ and Mini HLQ is a white disc. Calibration tests for any of the available probe types should provide results of TC=1.626-1.637±2%. The Macor standard drift calculations are based on a Macor standard (1.637 ± 0.033 W/mK) because its properties are closest to basalt cores (See Appendix: TK04 Recommended Heating Power for information).

    Standard VLQ- MACOR Standard

    Standard MACOR Disc

    Probe Test TC Value Expected Results:
    TC= 1.626±2% W/mK

    Probe Test TC Value Expected Results:
    TC= 1.637±2% W/mK


  3. To conduct a probe test, scan the STND MACOR disc TCON (H) label kept above the testing apparatus and ensure that the appropriate heating time and drift control (DCL) settings are input under the Configuration settings (See Configuring the Measurement Program).

    Example 1: A Standard HLQ properly positioned on the Standard MACOR Disc.


  4. Once the proper settings are confirmed, you can test the probes on the standard as if it were a normal sample. For the Standard VLQ this consists of leaving the probe needle in the MACOR standard, while for the HLQ probes you will need to attach it to the Standard MACOR Disc with a rubber band (See Example 1).

...

LIMS Component descriptions


Analysis

Component

Unit

Definition

TCON

Bottom_depth

m

Location of bottom of measurement, measured from the top of the hole

Comment

None

Comment about the run

Contact_value

None

Measure of contact quality between probe and sample

End_time

s

Elapsed time for end of analysis window

Heating_power

W/m

Power applied to needle during heating

Length_of_time

s

Elapsed time, start to finish, of analysis

Log_extreme_time

s

LET, used in calculation algorithm

Method

None

Data reduction method: SAM or TCON

Needle_name

None

Full-space or half-space

Number_of_solutions

None

Number of solutions found by the software

Offset

cm

Location of measurement from top of section

Start_time

s

Elapsed time into experiment for start of analysis window

Therm_con_average

W/(m·K)

Mean thermal conductivity result

Therm_con_number

None

Number of measurements in the population

Therm_con_result

W/(m·K)

Individual thermal conductivity result

Therm_con_stdev

W/(m·K)

Standard deviation (n-1) of measurement population

Top_depth

m

Location of top of measurement from top of hole


Troubleshooting


Drift phase takes too long:

...

Note: for loose sediments, use a lower heating power to avoid convective heat transport of pore fluids.

Material

 


Thermal Conductivity (W/m·K)

 


Recommended Heating Power (W/m) 


Mean

Range

VLQ

HLQ

Wood

0.21

0.1–0.35

0.15–1.3

Coal

0.29

0.1–1.5

0.15–5.4

Concrete

1.00

0.75–1.4

1.0–5.0

0.5–2.2

Fused silica

1.40

1.33–1.46

1.8–5.2

0.8–2.3

Clay

1.40

1.2–1.7

1.6–6.1

0.7–2.6

Silt

1.60

1.4–2.1

1.9–7.5

0.8–3.2

Basalt

1.95

1.4–5.4

1.9–19.0

0.8–7.6

Siltstone

2.04

0.6–4.0

0.8–14.0

0.4–5.7

Limestone

2.29

0.5–4.4

0.7–16.0

0.4–6.3

Syenite

2.31

1.3–5.3

1.7–19.0

0.8–7.5

Amphibolite

2.46

1.4–3.9

1.9–14.0

0.8–5.6

Claystone

2.46

1.6–3.4

2.1–12.0

0.5–9.3

Lava

2.47

0.2–4.5

0.3–16.0

0.2–6.4

Gabbro

2.50

1.6–4.1

2.1–15.0

0.9–5.9

Dolerite (Diabase)

2.64

1.6–4.4

2.1–16.0

0.5–6.3

Granodiorite

2.65

1.3–3.5

1.7–13.0

0.8–5.0

Quartz sand (wet)

2.70

2.4–3.1

3.2–11.0

1.3–4.5

Marble

2.80

2.1–3.5

1.8–13.0

1.2–5.0

Porphyrite

2.82

 


3.8–10.0

1.5–4.2

Boulder clay

2.90

2.5–3.3

3.4–12.0

1.4–4.8

Diorite

2.91

1.7–4.2

2.3–15.0

1.0–6.0

Slate (perpendicular)

2.91

1.5–3.9

2.0–14.0

0.9–5.6

Gneiss

2.95

1.2–4.7

1.6–17.0

0.7–6.7

Granite

3.05

1.2–4.5

1.6–16.0

0.7–6.4

Eclogite

3.10

2.4–3.4

3.2–12.0

1.3–4.9

Andesite

3.20

1.6–4.7

2.1–17.0

1.0–6.7

Dolomite

3.62

1.6–6.6

2.1–20.0

1.0–9.3

Slate (parallel)

3.80

2.2–5.2

3.0–19.0

1.2–7.4

Peridotite

3.81

 


5.0–14.0

2.0–5.5

Anhydrite

4.05

1.0–6.0

1.3–20.0

0.6–8.5

Pyroxenite

4.27

3.2–5.1

4.3–18.0

1.7–7.2

Dunite

4.41

3.5–5.2

4.7–19.0

1.9–7.4

Quartzite

4.55

3.1–>8

4.2–20.0

1.7–11.0

Quartz

9.50

6.5–12.5

8.7–20.0

3.5–17.0