...
- Obtain the IW sampling interval from the Curator and take the sample(s) when the core comes on deck.
- If possible, have the geochemist assist you in determining where to take the IW sample. Never take a sample that includes a change in lithology.
- Initially, 5 cm whole rounds are taken; as the water yield decreases, go to 10 cm whole rounds. Larger whole rounds may be taken after speaking to the SAC (Curator, EPM, and co-chiefs).
- If possible, take the IW sample from the bottom of the same section every time. As recovery decreases you will have to take the IW sample from wherever you can.
- A typical IW sampling plan is for 0 – 100 mbsf, take one IW per core and > 100 mbsf, take one IW every third core. Check with the Curator on the current plan.
- Note the core, section and interval of the sample and write it on the Curator's logsheet. The interval will be entered into SampleMaster by the Curator/ALO.
...
- Alkalinity/pH - If there is ample water, allocate 5 mL for shipboard analyses (alkalinity, salinity, IC, SPEC). Speak with the scientists about which analyses they would or wouldn't like performed (ie. Make a priority list). Extra water may be used for repeat analysis or stored (sent back to the repository).
- Scientific party - split and store the remaining water into the requested aliquots as outlined by the sampling plan. Each sample's storage/preservation is up to its requestor.
- ICP – 1 - 2 ml; must be acidified with trace metal-grade nitric acid (~ 30 10 µL).
- Remove the plunger from the syringe, rinse with DI water and place in the HCl bath for twelve hours. Rinse the components with DI water and bake dry at a low temperature. When completely dry, reassemble syringes and store them in ziploc bags.
...