- Switch the cell to On on the main coulometer unit.
- Allow the cell to equilibrate for 30-45 minutes before continuing. The %T should be 29.6 and steady.
- Open the Coulometer software and log in.
- Heat the vial with the weighed standard (100% CaCO3) in the heating unit for 2-3 minutes before measuring.
- Attach the top of the reaction cell to the sample vial. Bubbles should appear in the pre and post scrubber traps. If not, make sure all connections are tight.
- Select the sample from the list you wish to measure on the Coulometer software (Figure 9).
...