Introduction

The Section Half Multisensor Core Logger succeeded the Archive-Half Multisensor Track (AMST) in the physical properties laboratory. The SHMSL simultaneously measures spectral reflectance and magnetic susceptibility on core section halves. Data generated from these sensors are used to augment the core descriptions.

Reflectance Spectrometry

–Percent reflectance plotted against depth supports lithology descriptions.
–Color parameters can provide a detailed time series of relative changes in the composition of the core material and can be used to correlate sections from core to core or hole to hole and to analyze cyclicity of lithologic changes.
–Spectral data can be used to estimate the abundances of certain compounds.

  • Visible range provides semiquantitative estimates of hematite and goethite with better sensitivity than XRD.
  • Near-infrared or near-ultraviolet ranges allow estimates of carbonate, opal, organic matter, chlorite, and some combinations of clay minerals.

Magnetic Susceptibility

Magnetic Susceptibility can be used to confirm whole-round core section magnetic susceptibility measurements. The SHMSL can measure magnetic susceptibility at a similar sampling point spacing to the whole-round measurements, or the user can select a completely different frequency of analysis.
MS data is used for correlation with other age-depth proxy measurements.

Theory of Operation

A split and scraped core section in a half-core liner is placed on the core track, where a barcode scanner is used to scan the section ID (from the end cap) and imports sample information from the LIMS. The electronics platform moves along a track above the core section, recording the sample height in the core liner using a laser sensor. The laser establishes the location of bottom of the section. When this step is finished, the track will pull up out of the way of the user in order to facilitate the covering of the core section with GLAD® Plastic Wrap. When the user covers the core section and answers the prompt, the platform reverses the direction of movement, moving from bottom to top while recording point magnetic susceptibility and spectral reflectance data at user-specified data acquisition intervals (generally 2–10 cm).

Reflectance Spectrometry

–Measured from 380 to 900 nm at 2 nm intervals using both an LED and a halogen light source, covering a wavelength range through the visible spectrum and slightly into the infrared domain.
–Scanning the entire wavelength range takes ~5 s per data acquisition offset.
–Data are generated using the CIELAB L*a*b* color system:

  • L* represents lightness, where 0 yields black and 100 indicates diffuse white;
  • a* represents magenta to green tinting, where negative numbers indicate red/magenta shading and positive numbers indicate green shading; and
  • b* represents yellow to blue tinting, where negative numbers indicate yellow shading and positive number indicate blue shading.

–Data are also stored and returned as CIELAB Tristimulus XYZ values. The definition of X, Y, and Z is complex and the reader should refer to reference materials for an explanation. Refer to the related documentation section below.
–Finally data are stored and returned as RGB values to facilitate comparison with the imaging logger RGB values; note however that the sampling interval is quite different between the imaging logger and the integration sphere.

Magnetic Susceptibility

–Measured at the same data acquisition rate as spectral reflectance using a contact probe with a flat 15 mm diameter sensor.
–The sensor can be configured for different integration times (1 Hz or 0.1 Hz) and different numbers of replicate measurements. Our standard conditions are 3 measurements at 1 Hz measurement frequency for each offset. These three results are averaged and uploaded to the database. Thermal drift is effectively eliminated by zeroing the meter before each section.
–Data are reported in dimensionless instrument units (SI). In order to use these data as SI magnetic susceptibility units, the appropriate volume correction must be applied, which varies by sensor type. The user should not use the cgs setting so that the data set is consistent with previous measurements and with the whole-round logger results.


Hardware

The SHMSL system consists of the following hardware components (Figure 1):

  • Bartington MS2K or MS2E (Note: MS2E is not generally recommended because of its very small read area) magnetic susceptibility contact probe
  • Bartington MS2 meter, run in serial communications mode
  • Ocean Optics QE Pro visible spectrum spectrophotometer
  • Ocean Optics HL-2000 Halogen Light Source with Attenuator and TTL-Shutter
  • Ocean Optics Multi-LED BluLoop Light Source
  • Ocean Optics 30 mm integrating sphere, with a 10 mm aperture
  • Acuity AccuRange AR700 Laser
  • Barcode scanner
  • Instrument platform
  • Hardware abort switch


Figure 1. SHMSL Hardware

Bartington MS2K Probe Susceptibility Contact Probe

Specification

Value

Dimensions (mm)

Area of Response

165 x 145 x 50mm
15mm x 33mm diameter
25.4mm diameter

Operating frequency (kHz)

0.930

Depth of Response

50% at 3mm
10% at 8mm

Measurement period:
X 1 range
X 0.1 range


1.2 s SI
12s SI

Temperature induced drift

<2 x 10-5 SI in 5 min

Calibration (tip contact)

K x 10-5 SI (1cc)


Bartington MS2 Probe Susceptibility Meter

Specification

Value

Dimensions (mm)

241 × 158 × 50

Operating temperature range (°C)

–10 to +40

Linearity

1% to 9999


Ocean Optics QE Pro Hamamatsu S7031-1006S Detector

Specification

Value

Dimensions (mm)

182 x 110 x 47

Detector

TE Cooled, 1044x64 element CCD array

Detector range (nm)

185 – 1100 nm

Dynamic range:
Typical
Single Integration Period
100 Averages
10,000 Averages
Readout Noise


~85000:1
85,000:1(min)
850,000:1(min)
8,500,000:1(min)
2.5 counts RMS (typical)

Corrected linearity (%)

0.5% nonlinearity (max)

Pixel size (µm)

1024 active

Integration Time

8 ms – 60 minutes

Signal to noise ratio

1,000:1 (typical)

Operating Temperature

0 – 50° C

Temperature stability

<0.1° C


Ocean Optics HL-2000 Halogen Light Source with Attenuator and TTL-Shutter

Specification

Value

Dimensions (mm)

58 × 59 × 140

Wavelength range (nm)

360–1700

Stability (%)

0.5

Drift (%/h)

<0.1

Bulb life (h)

2000

Bulb color temperature (K)

3.000

Operating temperature (°C)

5°–35

Operating humidity (%)

5–95 at 40°C


Ocean Optics Multi-LED BluLoop Light Source

Specification

Value

Dimensions (mm)

62 x 66 x 150

Wavelength range (nm)

395 - 750

Stability (%)

< 0.15%

Time to Stabilize (min)

15

Drift (%/h)

<0.01%

Bulb life (h)

>10,000

Bulb color temperature (K)

3.000

Room temperature (°C)

5°–35°

Operating humidity (%)

5–95 at 40°C

Lamp Power (W)

<12

Ocean Optics ISP-30-6-R-GT Integrating Sphere

Specification

Value

Body diameter (mm)

59

Body height (mm)

58 (plus trap)

Specular Trap

Yes (SCE measurements)

Aperture (mm)

10

Internal Surface

PFTE

Spectral Range (nm)

200-2500

PFTE reflectivity

>98% (400-1500 nm)
>95% (200-2500 nm)

Fibre-optic connector

SMA 905


Acuity AccuRange 700 Triangulating Laser Displacement Sensor

Specification

Value

Laser (nm)

670 nm

Resolution

±0.03% of full scale

Operating temperature (°C)

0–60

Linearity/accuracy (%)

±0.2

Laboratory Apparatus

  • GLAD® Plastic Wrap (crystal clear polyethylene)

Note: This analysis requires clear polyethylene film such as GLAD® Wrap. Other types of plastic wrap (e.g., Saran Wrap™) will not serve.

Standards

Color Reflectance

  • Labsphere, Inc. Spectralon® white standard, certified at 99% reflectivity
  • Labsphere, Inc. diffuse color reflectance targets:
    • Grayscale: 99%, 50%, 20%, 2% reflectance
    • Color: Red, Orange, Yellow, Green, Cyan, Blue, Purple, Violet

Magnetic Susceptibility

The MS standards are used to check for drift, not to calibrate the meter or sensor. The JRSO puck is used to check drift instead of the Bartington standard because it is less sensitive to centering the sensor.

  • Bartington-issued MS puck (note: user must center this standard carefully to get the correct reading)
  • JRSO-created verification puck (~48 SIx10-6; mounted on track)


Launching the IMS- SHMSL application

The IMS-SHMSL software can be launched from the Windows Start menu or from a desktop icon (Figure 2).



Figure 2. IMS Icon


At launch, the program begins the following initialization process:

  • Tests all instrument communications
  • Reloads configuration values
  • Homes the X and Y axis

After successful initialization, the main IMS- SHMSL window will appear (Figure 3).




Figure 3. IMS Application Main Screen.

A Quick Introduction to the IMS Program Structure

IMS is a modular program. Individual modules are as follows:

  • INST plug-in: code for each of the instruments
  • MOTION plug-in: code for the motion control system
  • DAQ Engine: code that organizes INST and MOTION plug-ins into a track system

The SHMSL system, specifically, is built with three INST modules (MS2, QEPro, AR700), two MOTION modules (X-axis, Y-axis), and one DAQ Engine module.

The IMS Main User Interface (IMS-UI) calls these modules, instructs them to initialize, and provides a user interface to their functionality.

Each module manages a configuration file that opens the IMS program at the same state it was when previously closed and provides utilities for the user to edit or modify the configuration data and calibration routines.


The four buttons on the IMS-UI window provide access to utilities/editors via dropdown menus as shown in Figure 4.

Figure 4. SHMSL IMS Menu Options

Initial Instrument Setup

SHMSL Configuration

Each instrument has a separate setup menu.  Configuration values should be set during initial setup and configuration by the technician and scientists.  The general instrument setup should not change unless there have been changes to the hardware.


QEPro Configuration

Select Instruments> QEPro: Setup (Figure 4). The QEPro Parameters window will open (Figure 5).  This window is divided into 4 sections: General Setup, Acquistion Parameters, White and Dark Standards, and Integration Time.

Image Removed

Figure 5

.

QEPro Setup Window


General Setup

  • Instrument offset: The x-axis distance of the laser where the center of the sensor is over the benchmarks zero edge.
  • Sensor width:  Physical size of the sensor area
  • Contact width: Physical size of the contact surface
  • Analysis name: LIMS analysis component (must match LIMS component exactly).
  • Instrument group: LIMS instrument group logger name (i.e., SHMSL).
  • Model: model name of the sensor (from manufacturer).
  • S/N: serial number of the sensor (from manufacturer).
  • Menu name: value that appears as the instrument's menu name.
  • Full name: value that appears in instrument dialog boxes.


Acquisition Parameters section

  • Hi cut-off:  Used to define the upper limit of the region of interest (ROI) in the spectrum.  The spectrum above this value will not be saved or used in calculations. (Figure 6)

  • Low cut-off: Used to define the lower limit of the region of interest (ROI) in the spectrum.  The spectrum below this value will not be saved or used in calculations. (Figure 6)
  • Median Filter Rank: Number of adjacent channels used to filter noise from the signal.
  • Bin size: The width (nm) of each bin that the spectrum is recorded in. (Figure 6)
  • Illuminant: The Daylight Illuminant Standard for L*a*b calculations;  International Commission on Illumination (CIE) standard illuminant used in color calculations (also called daylight illuminant)
  • Geometry: References integration sphere measurement technique: illumination from a diffuse light source viewed 8° from normal (d/8); includes a gloss trap to exclude spectral reflections.
Image Removed
Figure 6. Illustration of high and low cut-off and bin size
  • .


White and Dark calibration Section

  • White Label ID: LIMS label_ID component value (STND-White).
  • White Text ID: LIMS text_ID component value (WHITE).
  • Dark Label ID name: LIMS label_ID component value (STND-Dark).
  • Dark Text ID: LIMS text_ID component value (DARK).
  • Standard X-offset: track position (X-axis; in cm) of laser when the center of the integration sphere sensor is over the center of the white standard.
  • Standard Y-offset: lift position (Y-axis) when the integration sphere is in contact with the White or Dark standard (same value for both).
  • Calibration expire: Time in hours between calibrations.  The user will be given a warning when the calibration time has expired.
  • Previous Timestamp: Time and date of last calibration (read only).
  • Remaining time: Clock displays of the time remaining until the next calibration is needed (read only).

Integration time section

  • Integration time: time the CCD is exposed; if too short, noise will be high; if too long, the sensor will saturate and clip data.

  • Starting integration time: beginning time (in seconds) value for WHITE calibration; set to 0.5 for new bulbs and up to 1 for aged bulbs. This setting has no effect on data from subsequent measurements.

  • Start Time: The integration time used to determine the white standard saturation point
  • % Saturation: Used by WHITE calibration to determine sample integration time by calculating the integration time required to reach the specified saturation (70%–95%). Note: if cores are light colored, increase throughput by using a low saturation level; if cores are dark colored, use a high saturation level to improve signal-to-noise ratio.
  • Step Time: The time step used to determine the White Standard's saturation point
  • White: Stack: The number of measurements to stack for the final White calibration
  • Dark: Stack:  The number of measurements to stack for the final Dark calibration

Image Added

Figure 5. QEPro Setup Window


Image Added

Figure 6. Illustration of high and low cut-off and bin size.

MS2 Configuration

Select Instruments> MS2: Setup (Figure 4). The QEPro Parameters window will open (Figure 7).  This window is divided into 3 sections: General Setup, MS Correction Factor, and Select MS Control.

Image RemovedFigure 7: MS2 Setup Window


General Setup

  • Track and Sensor Type?: Select the type of MS sensor and the track (SHMSL- Point)
  • Instrument offset: The x-axis distance of the laser where the center of the sensor is over the benchmarks zero edge.
  • Contact width: Physical size of the contact surface
  • Units: This value must match the front panel of the MS meter.  Set to SI per IODP standard practice
  • Range: This value must match the range setting on the MS meter, 1 or 10
  • Analysis name: LIMS analysis component (must match LIMS component exactly).
  • Instrument group: LIMS instrument group logger name (i.e., SHMSL).
  • Model: model name of the sensor (from manufacturer).
  • S/N: serial number of the sensor (from manufacturer).
  • Manufacturer's Name: The name of the manufacturer of the MS sensor
  • Menu name: value that appears as the instrument's menu name.
  • Full name: value that appears in instrument dialog boxes.

MS Correction Factor

These values are used when attempting to match data between multiple sensors. Default values are 1.

  • Correction Factor: A correction factor provided by the Bartington
  • Matching Constant: Value used to correc the sensor's output to a standard value

Select MS Control (Point Only)

  • Standards: Name of the standard being used
  • Label ID: The label ID used to identify the standard
  • Text ID: Unique identifier for the standard in the database
  • Standard's X offset: track position (X-axis) of the laser when the center of the MS sensor probe is over the center of the standard.
  • Standard's Y offset: lift position (Y-axis) when the MS sensor probe is in contact with the standard.
  • Standard's Value:  The standard's accepted/given value.  Value established by IODP is 48 +/- 2 SI

Image Added

Figure 7: MS2 Setup Window

Laser Configuration

Select Instruments> Laser: Setup (Figure 4). The QEPro Parameters window will open (Figure 8).  Image Removed

Figure 8: AR700

General Setup

Window

  • Analysis name: LIMS analysis component (must match LIMS component exactly).
  • Instrument group: LIMS instrument group logger name (i.e., SHMSL).
  • Model: model name of the sensor (from manufacturer).
  • S/N: serial number of the sensor (from manufacturer).
  • Manufacturer's Name: The name of the manufacturer of the laser.
  • Full name: value that appears in instrument dialog boxes.

Magnetic Susceptibility Utility

The SHMSL software also has a utility feature (shown in Figure 7, below) that allows the user to test the MS performance. The MS utility allows the user to zero the meter and to set it to continuous measurement.
Image Removed

Figure 7. MS2K Utility

Reflectance Utility

The software has a utility for the QE Pro that allows the user to run instantaneous or continuous measurements of the spectrum being acquired by the spectrometer. Different spectra (e.g., light standard, dark standard) can be overlaid and the user has a number of other options, as shown in Figure 10.
Image Removed

Figure 10. QE Pro Utility

Reflectance Standard Set Editor

The standards used for the calibration of the QE Pro (99% white Spectralon® standard and the "dark" standard, achieved by turning off the lights) are defined using the tool shown in Figure 11. Do not use this screen if you are not trained!
Image Removed

Figure 11. QE Pro Standard Editor

The AR700 configuration screen, shown in Figure 13, show the options available for the laser. These parameters are loaded to the AR700's on board memory and their values are returned in the AR700 Config Report panel (lower left). The utility can also be used to measure distances of a known metric to ensure that the laser is returning true distance values.
Image Removed

Figure 13. AR700 Configuration

Setting up the M-Drive Motion Control

This section is for advanced users only!

  1. To access the motion controller setup window, select Motion > Setup (Figure 14).
Image Removed Figure 14. M-Drive Motor Control From the M-Drive Motion Setup window (Figure 15, five setup panels can be accessed:
  • Track Configure: Motor and Track Options
  • Track Configure: Limit and Home Switches
  • Track Configure: Fixed Positions
  • Motion Profiles
  • Y-Axis Setup
Image Removed Figure 15. M-Drive Motion Controls

Track Configure – Motor and Track Options

Before using this panel you should be familiar with the M-Drive hardware and command manuals. Values 1 thru 4 (shown in Figure 16, below) are used by IMS to set motion controller parameters and convert encoder pulses into ± position values in centimeters. The correct settings for the SHMSL are as follows:

X-Axis Track Motor Setup Settings

  • Encoder Pulses/rev: 2048
  • Screw Pitch: 2.00000E+0
  • Gear Ratio: 4.00000
  • Direction of Positive Motion: CCW for positive encoder counts

Y-Axis Track Motor Setup Settings

  • Encoder Pulses/rev: 2048
  • Screw Pitch: 1.00000E+0
  • Gear Ratio: 1.00000
  • Direction of Positive Motion: CW for positive encoder counts
Image Removed
Figure 16. Track Motor and Options
Once these values have been properly set you should never, never have to change them. This panel is only used for the initial setup.

Track Configure – Fixed Positions

The screen shown in Figure 17, below allows the user to configure the fixed positions of the track and the limit switches. The correct values are as follows for each axis:

X-Axis Fixed Position Settings

  • Max Section Length: 160.0 cm
  • Track Length: 200.0 cm
  • Load Position: -4.0 cm
  • Unload Position: 170.o cm
  • All "PUSH Track Only" settings = OFF (or 0.0, as appropriate)

Y-Axis Fixed Position Settings

  • Max Section Length: not configurable for this axis
  • Track Length: not configurable for this axis
  • Load Position: -3.0 cm
  • Unload Position: -3.0 cm
  • All "PUSH Track Only" settings = OFF.
Image Removed Figure 17: Track Configuration

Track Configure – Limit & Home Switches

The screen shown in Figure 18 allows the user to configure the motion logic of the track. The SHMSL X-axis (down-core motion) and Y-axis (vertical sensor motion) are configured using this screen. As shown, the X-axis is "CW Look @ CCW Edge" logic (the first option). The Y-axis is "CCW Look @ CCW Edge" logic (the third option).
Image Removed

Figure 18. Limit and Home Switch Configuration

Track Configure – Motion Profile

The following table shows the motion profile settings for the SHMSL X and Y axes controlled by the interface shown in Figure 19, below.

Axis→

X-Axis

Y-Axis

Profile↓

Speed

Accel.

Decel.

Speed

Accel.

Decel.

DAQ Move

5.0

30.0

30.0

3.0

5.0

5.0

Limit Seek

3.0

10.0

80.0

3.0

10.0

80.0

Home Final

0.5

2.0

80.0

1.0

10.0

80.0

Load/Unload

15.0

10.0

10.0

10.0

5.0

5.0

Top/Profile

5.0

30.0

30.0

N/A

N/A

N/A

User Defined

5.0

5.0

5.0

N/A

N/A

N/A

Lift-Up

N/A

N/A

N/A

5.0

5.0

5.0

Lift-Down

N/A

N/A

N/A

3.0

5.0

5.0

Lift-Down Slow

N/A

N/A

N/A

1.0

5.0

5.0

Table: Motion Profile by Axis for SHMSL

Image Removed Figure 19. Motion Profiles

Y-Axis Setup

Y-Axis Setup is found in the DAQ menu option as shown in Figure 22, below. The Y-Axis setup is used to define the working distance for the AR700 laser. The screen shown in Figure 20, below, shows a typical set of standoff values for the various parameters, but if anything is changed on the Y-axis mount, these values should be determined experimentally.

  • To determine the distance to benchmark, click the "Get Benchmark Data" button. This value is normally between 46.1 to 46.5 mm.
  • The Touch Down Offset value is 3.80 cm.
  • The Safe Clearance value (usually set to 1.0 cm) is used to prevent the sensors from running into something (e.g., a tall rock piece) that is higher than the section half liner.
  • The Touch Compression value (usually set to 0.5 cm) is used to push downward gently after the landing to flatten the sensors against the core section.

Note: For very soft sediments, it may be necessary to set the Touch Compression to 0.0 cm.
Image Removed

Figure 20. Y-Axis Lift Setup

Setting Measurement Parameters

This section is for advanced users only!
Before measurement parameters can be set or changed the IMS must be unlocked. Note: Changing setup parameters can cause problems. Make sure you know what you are doing before changing any setup parameters.
Under File, select Unlock Setup as shown in Figure 21 below, then have the laboratory specialist type the unlock code into the keypad window (Note: this only has to be done if the setup has been locked, it usually remains unlocked).
Image Removed

Figure 21. IMS Control

To open the measurement editor, select DAQ > Measurement Editor as shown in Figure 22.
Image Removed

Figure 22. DAQ Control

Measurement Editor

In the Measurement Editor window, loaded instruments are shown in the Select Instrument fields. Click on an instrument to see its measurement parameters in the Instrument Parameters field.
Click the Instrument Parameters window to open the editor for that instrument. This will invoke the screen shown in Figure 23.
Image Removed

Figure 23. Measurement Editor Screen

The MS2K and QE Pro have similar editor screens (Figures 24 and 25, respectively). From each screen you can set the following parameters:
Interval (cm): set to 0.1 to 20 cm; intervals for both the MS and reflectance should either be the same or multiples of one another and whole number divisors of the distance between the sensors. If they are not whole number divisors, an inefficient landing schedule will be developed, taking much longer to analyze a core section.
Edge (cm): how close to measure to an edge (top and bottom of a section as well as edges of voids within the section). The distance is measured from the center of the sensor (MS2K sensor width = 4 mm or 10 mm, depending on orientation; QE Pro sensor width = 8 mm). Edge width = 1.0 cm works well.
Control (on/off): whether to measure a mounted control standard at the end of every section measurement.
Online (on/off): set the instrument online or offline.

MS-Specific Parameters

Average: (MS parameter only) number of measurements to take at each position (averaged for the final reported value).

QEPro-Specific Parameters

Method: this is a selector between single or stacked (average) measurements.
Stack #: set the number of measurements to stack and average.
Image Removed

Figure 24. MS2K Measurement Parameters

Image Removed

Figure 25. QE Pro Measurement Parameters

Click OK to save settings to the instrument configuration file and return to the Measurement Editor.
Only one parameter can be set for the laser, and that is Gap Detection Offset shown in Figure 26. This is the height below the benchmark which will be tagged by the system as a gap and will therefore not be measured. For piston cores, the recommended gap offset should be set to 10 mm or less. For hard rock cores, the gap offset should be set between 20 and 30 mm.
Image Removed

Figure 26. Gap Detection Offset

Running Samples

Core-half sections are measured on the SHMSL as soon as possible after splitting so that drying and oxidation do not affect ephemeral sample properties such as color reflectance. Sample preparation includes scraping to clean the core surface and covering wet core samples with plastic wrap to prevent contamination of the contact sensors (see Figure 33, below, but note that the sample must be laser-profiled before the wrap is applied).

Preparing Samples

  • Use a spatula or smear slide to clean the cut surface of the core by lightly scraping away any material that was smeared across the surface during core splitting.
  • Bring the endcap of the section half (usually A for archive half) to be measured to the SHMSL. Use the endcap rather than the label on the bottom of the section to scan the barcode for sample information, to prevent accidentally dropping the section half.
  • Place the archive section in the core tray with the blue endcap up against the benchmark. The benchmark is the white square at the head of the rails that hold the section halves (Figure 27).
    Image Removed
    Figure 27. Benchmark
  • Adjust the section so that it is as flat as possible with respect to the plane of the benchmark. There is a limit to how much the sensor heads can float when they land on a tilted section.
  • Do not wrap the cores yet. The Acuity AR700 laser cannot reliably see through the plastic to measure an accurate profile of the section half surface.
  • Measuring Samples

    Click the Start button (Figure 28) to open the Section Information screen (Figure 29).
    Image Removed
    Figure 28. Start Scan
    Image Removed
    Figure 29. Section Information Screen
  • Place the cursor in the SCAN text field (circled in black in Figure 29, above) so that the barcode information will be parsed appropriately. Pull the trigger on the barcode scanner; at the beep, the information will automatically fill for Expedition, Site, Hole, and so on.
  • (Optional) Enter any comment needed into the Comment field.
  • If material is missing from the top of the section, enter the distance in the Missing Top field (circled in red in Figure 29, above).
    Note: If material is missing at the top of the section, be sure the section half is positioned all the way up to the top of the rail. The software will add the missing top interval to all of the measurements in the database. For example, if a 150 cm section half has a missing 10 cm at the top, the yellow endcap should be placed against "0 cm" at the top of the instrument's rails. The user should enter 10.0 in the Missing Top (cm) field. The logger will measure the section, and all measurements will be placed at the correct offsets in the database.
  • Additional intervals can be specified for omission from measurement by clicking the Exclude Interval button (top left of Figure 29, above). Enter the top and bottom offsets of the areas to be excluded during the measurement pass as shown in Figure 30. The excluded intervals apply to all enabled sensors.
    Image Removed
    Figure 30. Exclude interval
  • When the required minimum sample information has been entered, the Measure button becomes active. Click it to start the measurement. If the QE Pro does not require calibration (see Calibrating the Sensors, below), proceed to the next step.
  • The sensor assembly will move down the track while the laser acquires a profile of the split surface of the section. When the software has detected a gap, it will indicate regions that will not be measured by the MS and reflectance sensors on the screen in red.
           Note: (Gap detection parameters are configurable; ask the PP tech for assistance in modifying them.) 
    Acquisition of this profile is why the section is not yet wrapped in plastic wrap—sometimes the laser will profile the wrap rather than the sediment beneath it and give incorrect heights to the gap detection routine. 
    Image Removed
    Figure31. Laser Profile
  • After the surface profile is completed, the measured section length as determined by the laser is displayed (Figure 32). Apply GLAD® Plastic Wrap to the surface of sediment cores as demonstrated by the PP Tech (Figure 33). Adjust the displayed length if needed in the Scan Length field and click GO to start the section measurement.
    Image Removed
    Figure32. Final Sample Preparation 
    IMPORTANT!!! Before pressing GO, it is necessary to cover the core section with GLAD® PlasticWrap in order to avoid damage to the integration sphere. Any mud that gets inside the sphere will ruin it!
    IMPORTANT! Do not cover the standards with GLAD® Plastic Wrap; they will give erroneous results if you do.
    Image Removed
    Figure 33. Wrapped Sample Ready to Analyze.
  • The measurement sequence begins. For efficiency, measurements begin at the bottom of the section and move upcore. Results are displayed during acquisition on instrument graphs on screen. Note that the QE Pro has several tabs that display different views of the color reflectance data. Tabs can be changed during the measurement sequence. The Normalized Spectra tab shows the corrected percent color reflectance being measured at each point.
    Image Removed
    Figure 34. Data Acquisition Screen
  • After the section has been measured, the logger takes measurements using the MS2K and WHITE standards as check standards if the control sample measurement was set up in the measurement parameters.
  • Calibrating the Sensors

    Before running the first sample, the software will check the status of each instrument's calibration. If any instrument needs calibration, a prompt will appear listing out of date calibrations. It is recommended that the user select every instrument that needs calibration and click the Calibrate button before continuing. There is an option to ignore one or all calibrations (click Cancel), but the calibration list will reappear with every run until the calibrations are completed and data run without a calibration update will be flagged as calibration invalid in LIMS.

    Color Reflectance Spectrophotometer (Auto Calibration)

    The color reflectance spectrophotometer calibrates on two spectra, pure white (Spectralon® WHITE standard) and pure black (DARK standard – lights off), mounted in the track's bench. Color calibration is required approximately every 6 hours. The control program will notify when calibration is due and automatically open the Calibrate routine.
    Note: a calibration may be run at any time, even if the mandatory period has nod elapsed.
    Image Removed
    Figure 35. WHITE Spectralon® Calibration Standard

  • If the QE Pro is out of calibration, the Instrument Calibration List screen will appear.
  • Before beginning the calibration, ensure the WHITE standard is clean (Figure 35). If the Spectralon® standard appears even the least bit gray or discolored, contact the technician to clean or replace the standard! It may be useful to compare a clean, new piece of white paper to the standard—if the paper seems whiter, the standard is quite dirty!
    1. The Spectralon® standard can be cleaned according to the Spectralon® Reflectance Standards Care and Handling Guidelines, found in the Cumulus database under SHMSL>Diffuse Color Standards.
  • Click the Calibrate ALL CHECKED Instruments button to begin the calibration process. The logger moves the QE Pro integrating sphere over the WHITE calibration standard and lowers the sensor until it touches. The sensor takes some preliminary measurements to establish the integrating time based on an 80% saturation level of the total response (see Figure 36, below). These measurements take about 0.30 to 0.55 seconds for the current dual-light source configuration, assuming the bulbs are new.
    Note: If integration times increase to more than 1 second, alert the IODP technician. The halogen bulb or the LED source may need replacing.
  • Click Accept to accept the suggested integration time. The logger will then take 20 measurements of the WHITE standard at the calculated integration time. On the screen (Figure 37):
  • Max Counts: current highest value
  • Target Counts: percent saturation value as set in the instrument parameters (range = 32,000 counts)
  • Integration Time: measurement period where the highest wavelength count is equal to or exceeds 80 percent of the spectrophotometer's range.
    Image Removed
    Figure 36. White Calibration Screen
    Image Removed
    Figure 37. Integration TimeAfter the WHITE measurements are completed. The shutters on the light sources will close and the DARK measurement will acquire 20 measurements. The dark measurement is a baseline measurement that includes thermal noise of the system. On the screen (Figure 38):
  • Temperature: should remain below 50°C; note that the TEC temperature is usually about -9¿C (accessed through the QE Pro Utilities screen.
  • Spectral Mean: mean of the entire spectrum (at the line across the display plot). This should be only about 200 counts higher than the Dark Pixel count. With plastic wrap, it should not exceed 500 counts.
  • Dark Pixels: 20 pixels that have been deliberately masked to allow no light. These counts represent the thermal noise of the spectrophotometer.
    Image Removed
    Figure 38. DARK Calibration ScreenThe final screen shows the normalization of spectra that will occur with the just-acquired calibration (note: low values are better than high values because of signal-to-noise ratio). Ideally, the graph would display a straight line at zero value. However, a normalization factor is required to be applied to the XYZ and L*a*b* color indexes. Normalization amplifies the noise as well as the signal. If core flow allows, the data quality can be increased by averaging multiple measurements (in Measurement Editor increase the Average parameter).
    Image Removed
    Figure 39. Calibration Normalization Factor


    Image Added

    Figure 8: AR700 Setup Window

    Instrument Calibration

    Color Reflectance Spectrophotometer

    The color reflectance spectrophotometer calibrates on two spectra, pure white (Spectralon® WHITE standard) (Figure 9) and pure black (DARK standard – lights off), mounted in the track's bench. The spectrometer calibraton can be triggered automatically or manually by the user.  The expiration time for the spectrometer calibration is set in the QEPro setup menu (Figure 5). Before running the first sample, the software will check the status of the spectrometer calibration. If the calibration has expired, a prompt will appear asking the user to begin an automatic calibration.  It is recommended that the user calibrate every time the software prompts the user.   

    Image Added

    Figure 9. WHITE Spectralon® Calibration Standard


    There is an option to ignore one or all calibrations (click Cancel), but the calibration list will reappear with every run until the calibrations are completed and data run without a calibration update will be flagged as calibration invalid in LIMS.

    Automatic Calibration

    1. If the QE Pro is out of calibration, the Instrument Calibration List screen will appear (Figure 10).

      Image Added

    Figure 10. Instrument Calibration Prompt


    2. Before beginning the calibration, ensure the WHITE standard is clean (Figure 9). If the Spectralon® standard appears even the least bit gray or discolored, contact the technician to clean or replace the standard! It may be useful to compare a clean, new piece of white paper to the standard—if the paper seems whiter, the standard is quite dirty.

    a. The Spectralon® standard can be cleaned according to the Spectralon® Reflectance Standards Care and Handling Guidelines. Spectralon Standards Care and Handling.pdf

    3. Click the Calibrate button to begin the calibration process. The user may select Ignore Continue Measurement and skip the calibration, but the software will prompt the user for a calibration every time a measurement is started until the calibration is completed.

    a. The logger moves the QE Pro integrating sphere over the WHITE calibration standard and lowers the sensor until it touches. The sensor begins the white calibration measurements (Figure 11) based on the set saturation level of the total response (Figure 5). 

    b. The display includes:

      • Max Counts: current highest value
      • Target Counts: percent saturation value as set in the instrument parameters (range = 32,000 counts)
      • Integration Time: measurement period where the highest wavelength count is equal to or exceeds 80 percent of the spectrophotometer's range.

    4. When the white calibration is complete, the integration time will be displayed on the screen and the user will be prompted to Accept, Re-Start, or Abort the calibration (Figure 12). 

    a. The integration time should be about 0.30 to 0.55 seconds for the current dual-light source configuration, assuming the bulbs are new. 

    b.  If integration times increase to more than 1 second, alert the IODP technician. The halogen bulb or the LED source may need replacing.

    5. Click Accept to accept the white calibratoin.  (Figure 11):



    Image Added
    Figure 11. White Calibration Screen


    Image Added
    Figure 12. White Calibration Integration Time Display

    6. After the WHITE measurements are completed, the shutters on the light sources will close and the DARK measurement will acquire 20 measurements. The dark measurement is a baseline measurement that includes thermal noise of the system. On the screen (Figure 13):

      • Temperature: should remain below 50°C; note that the TEC temperature is usually about -9 degrees C; accessed through the QE Pro Utilities screen.
      • Spectral Mean: mean of the entire spectrum (at the line across the display plot). This should be only about 200 counts higher than the Dark Pixel count. With plastic wrap, it should not exceed 500 counts
      • Dark Pixels: 20 pixels that have been deliberately masked to allow no light. These counts represent the thermal noise of the spectrophotometer.



    Image Added
    Figure 13. DARK Calibration Screen

    6. When the dark calibration is complete, the user is prompted to Abort, Re-Start, or Accept the dark calibration (Figure 14).

    Image Added

    Figure 14. Dark Calibration Completed Prompt

    7. The final screen shows the normalization of spectra that will occur with the just-acquired calibration (note: low values are better than high values because of signal-to-noise ratio) (Figure 15). Ideally, the graph would display a straight line at zero value. However, a normalization factor is required to be applied to the XYZ and L*a*b* color indexes. Normalization amplifies the noise as well as the signal. If core flow allows, the data quality can be increased by averaging multiple measurements (in Measurement Editor increase the Average parameter).



    Image Added

    Figure 15. Calibration Normalization Factor

    8. Select Accept.  If this was an automatic calibration at the start of a measurement, the laser profile will start automatically, followed by the normal measurement sequence.

    Manual Calibration

    To perform calibration manually (not when specified by the software), select Instruments > QE Pro: Calibrate (Figure 5).  The QEPro will move to the white standard and begin the calibration process.  Follow the automatic procedure from step 4 to complete the calibration.


    Recognizing a Bad Calibration

    Bad Calibration with Blue Light and White Halogen on and working. 
    Image Added
    When the red end of the calibration is over the amount of counts the program will process, this might be for a number of reasons, a bad light, too much light bleed-in, or some other cause. The blue end of the spectrum is within the correct amount of counts and maintains the right curve with the spike around 400 wavelength and a dip into the 420-430 range. 


    Bad Calibration When the blue light is off. 
    Image Added
    You'll notice the left side of the graph, the blue side isn't there, whereas the red side of the graph is at 180K counts.

    Magnetic Susceptibility Meter

    The point magnetic susceptibility meter is calibrated by the manufacturer (Bartington, Ltd.). The probe is zeroed in air before the core section is measured, so drift is not a significant factor. It is not necessary to calibrate the MS2K or MS2E probes (or the MS2C loops on the whole-round loggers), but the calibration check standard can be used to demonstrate consistent results.


    Utilities

    Magnetic Susceptibility Utility

    The SHMSL software also has a utility feature (shown in Figure 7, below) that allows the user to test the MS performance. The MS utility allows the user to zero the meter and to set it to continuous measurement.
    Image Added

    Figure 7. MS2K Utility

    Reflectance Utility

    The software has a utility for the QE Pro that allows the user to run instantaneous or continuous measurements of the spectrum being acquired by the spectrometer. Different spectra (e.g., light standard, dark standard) can be overlaid and the user has a number of other options, as shown in Figure 10.
    Image Added

    Figure 10. QE Pro Utility

    Reflectance Standard Set Editor

    The standards used for the calibration of the QE Pro (99% white Spectralon® standard and the "dark" standard, achieved by turning off the lights) are defined using the tool shown in Figure 11. Do not use this screen if you are not trained!
    Image Added

    Figure 11. QE Pro Standard Editor



    The AR700 configuration screen, shown in Figure 13, show the options available for the laser. These parameters are loaded to the AR700's on board memory and their values are returned in the AR700 Config Report panel (lower left). The utility can also be used to measure distances of a known metric to ensure that the laser is returning true distance values.
    Image Added

    Figure 13. AR700 Configuration

    Setting up the M-Drive Motion Control

    This section is for advanced users only!

    1. To access the motion controller setup window, select Motion > Setup (Figure 14).


    Image Added Figure 14. M-Drive Motor Control


    From the M-Drive Motion Setup window (Figure 15, five setup panels can be accessed:

    • Track Configure: Motor and Track Options
    • Track Configure: Limit and Home Switches
    • Track Configure: Fixed Positions
    • Motion Profiles
    • Y-Axis Setup


    Image Added Figure 15. M-Drive Motion Controls

    Track Configure – Motor and Track Options

    Before using this panel you should be familiar with the M-Drive hardware and command manuals. Values 1 thru 4 (shown in Figure 16, below) are used by IMS to set motion controller parameters and convert encoder pulses into ± position values in centimeters. The correct settings for the SHMSL are as follows:

    X-Axis Track Motor Setup Settings

    • Encoder Pulses/rev: 2048
    • Screw Pitch: 2.00000E+0
    • Gear Ratio: 4.00000
    • Direction of Positive Motion: CCW for positive encoder counts

    Y-Axis Track Motor Setup Settings

    • Encoder Pulses/rev: 2048
    • Screw Pitch: 1.00000E+0
    • Gear Ratio: 1.00000
    • Direction of Positive Motion: CW for positive encoder counts


    Image Added
    Figure 16. Track Motor and Options
    Once these values have been properly set you should never, never have to change them. This panel is only used for the initial setup.

    Track Configure – Fixed Positions

    The screen shown in Figure 17, below allows the user to configure the fixed positions of the track and the limit switches. The correct values are as follows for each axis:

    X-Axis Fixed Position Settings

    • Max Section Length: 160.0 cm
    • Track Length: 200.0 cm
    • Load Position: -4.0 cm
    • Unload Position: 170.o cm
    • All "PUSH Track Only" settings = OFF (or 0.0, as appropriate)

    Y-Axis Fixed Position Settings

    • Max Section Length: not configurable for this axis
    • Track Length: not configurable for this axis
    • Load Position: -3.0 cm
    • Unload Position: -3.0 cm
    • All "PUSH Track Only" settings = OFF.



    Image Added Figure 17: Track Configuration

    Track Configure – Limit & Home Switches

    The screen shown in Figure 18 allows the user to configure the motion logic of the track. The SHMSL X-axis (down-core motion) and Y-axis (vertical sensor motion) are configured using this screen. As shown, the X-axis is "CW Look @ CCW Edge" logic (the first option). The Y-axis is "CCW Look @ CCW Edge" logic (the third option).
    Image Added

    Figure 18. Limit and Home Switch Configuration

    Track Configure – Motion Profile

    The following table shows the motion profile settings for the SHMSL X and Y axes controlled by the interface shown in Figure 19, below.

    Axis→

    X-Axis

    Y-Axis

    Profile↓

    Speed

    Accel.

    Decel.

    Speed

    Accel.

    Decel.

    DAQ Move

    5.0

    30.0

    30.0

    3.0

    5.0

    5.0

    Limit Seek

    3.0

    10.0

    80.0

    3.0

    10.0

    80.0

    Home Final

    0.5

    2.0

    80.0

    1.0

    10.0

    80.0

    Load/Unload

    15.0

    10.0

    10.0

    10.0

    5.0

    5.0

    Top/Profile

    5.0

    30.0

    30.0

    N/A

    N/A

    N/A

    User Defined

    5.0

    5.0

    5.0

    N/A

    N/A

    N/A

    Lift-Up

    N/A

    N/A

    N/A

    5.0

    5.0

    5.0

    Lift-Down

    N/A

    N/A

    N/A

    3.0

    5.0

    5.0

    Lift-Down Slow

    N/A

    N/A

    N/A

    1.0

    5.0

    5.0

    Table: Motion Profile by Axis for SHMSL


    Image Added Figure 19. Motion Profiles

    Y-Axis Setup

    Y-Axis Setup is found in the DAQ menu option as shown in Figure 22, below. The Y-Axis setup is used to define the working distance for the AR700 laser. The screen shown in Figure 20, below, shows a typical set of standoff values for the various parameters, but if anything is changed on the Y-axis mount, these values should be determined experimentally.

    • To determine the distance to benchmark, click the "Get Benchmark Data" button. This value is normally between 46.1 to 46.5 mm.
    • The Touch Down Offset value is 3.80 cm.
    • The Safe Clearance value (usually set to 1.0 cm) is used to prevent the sensors from running into something (e.g., a tall rock piece) that is higher than the section half liner.
    • The Touch Compression value (usually set to 0.5 cm) is used to push downward gently after the landing to flatten the sensors against the core section.

    Note: For very soft sediments, it may be necessary to set the Touch Compression to 0.0 cm.
    Image Added

    Figure 20. Y-Axis Lift Setup

    Setting Measurement Parameters

    This section is for advanced users only!
    Before measurement parameters can be set or changed the IMS must be unlocked. Note: Changing setup parameters can cause problems. Make sure you know what you are doing before changing any setup parameters.
    Under File, select Unlock Setup as shown in Figure 21 below, then have the laboratory specialist type the unlock code into the keypad window (Note: this only has to be done if the setup has been locked, it usually remains unlocked).
    Image Added

    Figure 21. IMS Control


    To open the measurement editor, select DAQ > Measurement Editor as shown in Figure 22.
    Image Added

    Figure 22. DAQ Control

    Measurement Editor

    In the Measurement Editor window, loaded instruments are shown in the Select Instrument fields. Click on an instrument to see its measurement parameters in the Instrument Parameters field.
    Click the Instrument Parameters window to open the editor for that instrument. This will invoke the screen shown in Figure 23.
    Image Added

    Figure 23. Measurement Editor Screen


    The MS2K and QE Pro have similar editor screens (Figures 24 and 25, respectively). From each screen you can set the following parameters:
    Interval (cm): set to 0.1 to 20 cm; intervals for both the MS and reflectance should either be the same or multiples of one another and whole number divisors of the distance between the sensors. If they are not whole number divisors, an inefficient landing schedule will be developed, taking much longer to analyze a core section.
    Edge (cm): how close to measure to an edge (top and bottom of a section as well as edges of voids within the section). The distance is measured from the center of the sensor (MS2K sensor width = 4 mm or 10 mm, depending on orientation; QE Pro sensor width = 8 mm). Edge width = 1.0 cm works well.
    Control (on/off): whether to measure a mounted control standard at the end of every section measurement.
    Online (on/off): set the instrument online or offline.

    MS-Specific Parameters

    Average: (MS parameter only) number of measurements to take at each position (averaged for the final reported value).

    QEPro-Specific Parameters

    Method: this is a selector between single or stacked (average) measurements.
    Stack #: set the number of measurements to stack and average.

    Image Added

    Figure 24. MS2K Measurement Parameters


    Image Added

    Figure 25. QE Pro Measurement Parameters

    Click OK to save settings to the instrument configuration file and return to the Measurement Editor.
    Only one parameter can be set for the laser, and that is Gap Detection Offset shown in Figure 26. This is the height below the benchmark which will be tagged by the system as a gap and will therefore not be measured. For piston cores, the recommended gap offset should be set to 10 mm or less. For hard rock cores, the gap offset should be set between 20 and 30 mm.
    Image Added

    Figure 26. Gap Detection Offset

    Running Samples

    Core-half sections are measured on the SHMSL as soon as possible after splitting so that drying and oxidation do not affect ephemeral sample properties such as color reflectance. Sample preparation includes scraping to clean the core surface and covering wet core samples with plastic wrap to prevent contamination of the contact sensors (see Figure 33, below, but note that the sample must be laser-profiled before the wrap is applied).

    Preparing Samples

    1. Use a spatula or smear slide to clean the cut surface of the core by lightly scraping away any material that was smeared across the surface during core splitting.
    2. Bring the endcap of the section half (usually A for archive half) to be measured to the SHMSL. Use the endcap rather than the label on the bottom of the section to scan the barcode for sample information, to prevent accidentally dropping the section half.
    3. Place the archive section in the core tray with the blue endcap up against the benchmark. The benchmark is the white square at the head of the rails that hold the section halves (Figure 27).

      Image Added



      Figure 27. Benchmark

    4. Adjust the section so that it is as flat as possible with respect to the plane of the benchmark. There is a limit to how much the sensor heads can float when they land on a tilted section.
    5. Do not wrap the cores yet. The Acuity AR700 laser cannot reliably see through the plastic to measure an accurate profile of the section half surface.

    Measuring Samples

    1. Click the Start button (Figure 28) to open the Section Information screen (Figure 29).

      Image Added
      Figure 28. Start Scan


      Image Added
      Figure 29. Section Information Screen


      1. Place the cursor in the SCAN text field (circled in black in Figure 29, above) so that the barcode information will be parsed appropriately. Pull the trigger on the barcode scanner; at the beep, the information will automatically fill for Expedition, Site, Hole, and so on.
      2. (Optional) Enter any comment needed into the Comment field.
      3. If material is missing from the top of the section, enter the distance in the Missing Top field (circled in red in Figure 29, above).
        Note: If material is missing at the top of the section, be sure the section half is positioned all the way up to the top of the rail. The software will add the missing top interval to all of the measurements in the database. For example, if a 150 cm section half has a missing 10 cm at the top, the yellow endcap should be placed against "0 cm" at the top of the instrument's rails. The user should enter 10.0 in the Missing Top (cm) field. The logger will measure the section, and all measurements will be placed at the correct offsets in the database.
      4. Additional intervals can be specified for omission from measurement by clicking the Exclude Interval button (top left of Figure 29, above). Enter the top and bottom offsets of the areas to be excluded during the measurement pass as shown in Figure 30. The excluded intervals apply to all enabled sensors.
        Image Added
        Figure 30. Exclude interval

    2. When the required minimum sample information has been entered, the Measure button becomes active. Click it to start the measurement. If the QE Pro does not require calibration (see Calibrating the Sensors, below), proceed to the next step.
    3. The sensor assembly will move down the track while the laser acquires a profile of the split surface of the section. When the software has detected a gap, it will indicate regions that will not be measured by the MS and reflectance sensors on the screen in red.
             Note: (Gap detection parameters are configurable; ask the PP tech for assistance in modifying them.) 
      Acquisition of this profile is why the section is not yet wrapped in plastic wrap—sometimes the laser will profile the wrap rather than the sediment beneath it and give incorrect heights to the gap detection routine. 
      Image Added
      Figure31. Laser Profile


    1. After the surface profile is completed, the measured section length as determined by the laser is displayed (Figure 32). Apply GLAD® Plastic Wrap to the surface of sediment cores as demonstrated by the PP Tech (Figure 33). Adjust the displayed length if needed in the Scan Length field and click GO to start the section measurement.
      Image Added
      Figure32. Final Sample Preparation 

      IMPORTANT!!! Before pressing GO, it is necessary to cover the core section with GLAD® PlasticWrap in order to avoid damage to the integration sphere. Any mud that gets inside the sphere will ruin it!
      IMPORTANT! Do not cover the standards with GLAD® Plastic Wrap; they will give erroneous results if you do.

      Image Added
      Figure 33. Wrapped Sample Ready to Analyze.
    2. The measurement sequence begins. For efficiency, measurements begin at the bottom of the section and move upcore. Results are displayed during acquisition on instrument graphs on screen. Note that the QE Pro has several tabs that display different views of the color reflectance data. Tabs can be changed during the measurement sequence. The Normalized Spectra tab shows the corrected percent color reflectance being measured at each point.
      Image Added
      Figure 34. Data Acquisition Screen
    3. After the section has been measured, the logger takes measurements using the MS2K and WHITE standards as check standards if the control sample measurement was set up in the measurement parameters.
  • If this was a mandatory calibration the laser profile will start automatically, followed by the normal measurement sequence.
  • Spectrophotometer (Manual Calibration)

    To perform calibration manually (not when specified by the software), follow these steps:

    1. From the Main control screen, select Calibration > Calibrate. Follow the instructions on the screen.
    2. Continue from Step 3 in Auto Calibration

    Please see Appendix - SHMSL User Guide- DELETE (INFO HAS BEEN ADDED TO USER GUIDE) for notes on how to recognize a "bad" calibration.

    Point Magnetic Susceptibility Meter

    The point magnetic susceptibility meter is calibrated by the manufacturer (Bartington, Ltd.). The probe is zeroed in air before the core section is measured, so drift is not a significant factor. It is not necessary to calibrate the MS2K or MS2E probes (or the MS2C loops on the whole-round loggers), but the calibration check standard can be used to demonstrate consistent results.

    Sensor Calibration Triggers

    Time: Calibration expiration time can be set in the Instrument Setup screens. Calibration expiration can be set from 1 to 24 h (default = 6 h). The time trigger is independent of number of samples run.
    Image Removed
    Figure 40. Calibration Time Trigger
  • Core number or section number change: Set this option in Track > Calibrate Setup.
  • Manual override.
  • Data Handling

    After the core section has been measured, the software generates the following files/reports, which are uploaded into the LIMS upon acceptance (see LIMS Integration):
    C:\data\IN

    • Section half L* a* b* and XYZ data
    • MS2K check standard result
    • MS2K (MSPOINT) data file
    • QE Pro white check standard result
    • Laser profile file


    C:\AUX_DATA\PROFILE
    Laser profile data in 0.1 mm steps
    C:\AUX_DATA\RSC\CALIB
    White and Dark calibration spectral data per wavelength (380–900 in 2 nm steps)
    C:\AUX_DATA\RSC\CNTRL
    Raw spectra and normalized (% color reflectance) spectral data per wavelength (380–900 in 2 nm steps) for the white check standard at the end of a run
    C:\AUX_DATA\RSC\RUN
    Raw spectra and normalized (% color reflectance) spectral data per wavelength (380–900 in 2 nm steps) for the measured section

    Quality Assurance/Quality Control

    Analytical Batch

    The analytical batch is defined by the number of samples run between each spectrometer calibration. Each sample in the batch run with the current calibration is associated with that calibration data in the LIMS. If a calibration problem is discovered, all samples in the batch can be identified and rerun.

    Accuracy

    Magnetic Susceptibility

    Measure a well-characterized magnetic susceptibility control sample and compare results with true value and/or whole-core track results.

    Reflectance

    Measure a second standard, standard reference material, or characterize a material in-house to use as a control. Measuring the BCRA-calibrated tiles ensures the desired accuracy is being maintained. This should be done periodically throughout each expedition.

    Precision

    Run samples or standard reference materials more than once (separate measurement runs) and calculate the standard deviation. This should be done every ~20-section halves allowing for a comparison between runs and estimation of uncertainty based on precision.

    LIMS Integration

    LIMS Components

    Results are stored in the LIMS database associated with an analysis code and an analysis component. Analysis codes and their components and units are listed below.

    Magnetic Susceptibility


    Analysis

    Component

    Definition

    Unit

    MSPOINT

    comment

    User entered comment

    None

    config_asman_id

    ASMAN ID number of configuration file

    None

    config_filename

    ASMAN filename of configuration file

    None

    instrument_group

    Logger sensor is mounted on

    None

    magnetic_susceptibility

    Magnetic susceptibility of sample

    SI x 10-6 (unitless)

    offset

    Offset of measurement on section half

    cm

    run_asman_id

    ASMAN ID number of runtime file

    None

    run_filename

    ASMAN filename of runtime file

    None


    AnalysisComponentDefiniton

    Magnetic Susceptibility. Point

    or Contact System (MSP)

    Expexpedition number
    Sitesite number
    Holehole number
    Corecore number
    Typetype indicates the coring tool used to recover the core (typical types are F, H, R, X).
    Sectsection number
    A/Warchive (A) or working (W) section half
    Offset (cm)position of the observation made, measured relative to the top of a section or section half
    Depth CSF-A (m)location of the observation expressed relative to the top of a hole
    Depth [other] (m)location of the observation expressed relative to the top of a hole. The location is presented in a scale selected by the science party or the report user.
    Magnetic susceptibility (SI)

    magnetic susceptibility of the sample, volume-corrected as follows:

    • MS2E sensor: magnetic susceptibility is reported in calibrated SI values, assuming the sample is a flat surface and has a depth greater than 10 mm
    • MS2K sensor: magnetic susceptibility is reported in calibrated SI values, assuming the sample is a flat surface and has a depth greater than 10 mm
    Timestamp (UTC)point in time at which an observation or set of observations was made on the logger.
    Instrumentan abbreviation or mnemonic for the sensing device used to make this observation.
    Instrument groupabbreviation or mnemonic for the data collection device (logger) used to acquire this observation (SHMSL).
    Text IDautomatically generated unique database identifier for a sample, visible on printed labels.
    Test NoUnique number associated with the instrument measurement steps that produced these data.
    Commentsobserver's notes about a measurement, the sample, or the measurement process.


    Color Reflectance Spectrophotometry


    Analysis

    Component

    Definition

    Unit

    RSC

    calibration_valid

    Y or N Boolean for valid calibration

    None

    chroma

    Chroma value of sample in hvc notation

    None

    cielab_a_star

    a*

    None

    cielab_b_star

    b*

    None

    cielab_l_star

    L*

    None

    comments

    User-entered comment

    None

    config_asman_id

    ASMAN ID number of configuration file

    None

    config_filename

    ASMAN filename of configuration file

    None

    geometry

    d/8 specular component excluded (SPE)

    None

    hue

    Hue value of sample in hvc notation

    None

    illuminant

    Illuminant standard used

    None

    instrument_group

    Logger sensor is mounted on

    None

    observer

    Observer geometry used

    None

    offset

    Offset of measurement on section half

    Cm

    rsc_norm_asman_id

    Normalized (percent reflectance) spectral file ASMAN ID

    None

    rsc_norm_filename

    Normalized (percent reflectance) spectral file name

    None

    rsc_raw_asman_id

    Raw spectral file ASMAN ID

    None

    rsc_raw_filename

    Raw spectral file name

    None

    run_asman_id

    Runtime file ASMAN ID

    None

    run_filename

    Runtime file name

    None

    sample_time

    Spectral acquisition time

    ns

    tristimulus_x

    Tristimulus X value of sample

    None

     

    tristimulus_y

    Tristimulus Y value of sample

    None

     

    tristimulus_z

    Tristimulus Z value of sample

    None


    AnalysisComponentDefinition

    Reflectance Spectroscopy and Colorimetry

    (RSC)

    Expexpedition number
    Sitesite number
    Holehole number
    Corecore number
    Typetype indicates the coring tool used to recover the core (typical types are F, H, R, X)
    Sectsection number
    A/Warchive (A) or working (W) section half
    Offset (cm)position of the observation made, measured relative to the top of a section or section half.
    Depth CSF-A (m)location of the observation expressed relative to the top of a hole
    Depth (other) (m)location of the observation expressed relative to the top of a hole. The location is presented in a scale selected by the science party or the report user
    Reflectance L*lightness of the sample expressed in the CIELAB color space. Expected values between 0-100
    Reflectance a*position of the color on the axis between red/magenta and green (negative = greater green character) in the CIELAB color space
    Reflectance b*Position of the color on the axis between yellow and blue (negative = greater blue character) in the CIELAB color space
    Tristimulus Xx-component of the tristimulus value
    Tristimulus Yy-component of the tristimulus value
    Tristimulus Zz-component of the tristimulus value
    Normalized spectral data

    ASCII file with normalized spectral data

    Unnormalized spectral dataASCII file with unnormalized spectral data
    Timestamp (UTC)point in time at which the observation or set of observations was made.
    Instrumentan abbreviation or mnemonic for the spectrophotometric sensing device used to make this observation (OOUSB4000V)
    Instrument groupabbreviation or mnemonic for the data collection device (logger) used to acquire this observation (SHMSL)
    Text IDautomatically generated unique database identifier for a sample, visible on printed labels.
    Test NoUnique number associated with the instrument measurement steps that produced these data.
    Commentsobserver's notes about a measurement, the sample, or the measurement process.


    Uploading Data to LIMS

    As with most laboratory data, uploading Uploading SHMSL data to the LIMS database is executed via the MUT application. Look for an icon on the bottom menu of the Desktop (shown by the red triangle below in Figure 41) or the more prominent "Puppy" MUT Icon (usually on the Desktop). Double-click on one of these application launchers. MegaUploadaTron application. 


    Figure 41. MUT Icons


    The user must log in using database credentials to use MUT. Once the application is activated, it displays a table-like list of files in the C:\DATA\IN directory as shown in Figure 42, below. Files are ready to be uploaded if they have a green check mark next to them.

    Figure 42. SHMSL MUT Screen


    Once the Upload button is clicked or the Automatic Upload option is checked, files will be transferred to the LIMS database via MUT and then moved to the archive folder. The presence of files in the Archive folder is the indication that they have been uploaded to the database.

    • Purple question marks on files in MUT indicates that MUT cannot identify them.
    • Red and white X icons on files in MUT indicate errors associated with the files.

    Please ask a technician for help if you see these indications.

    Data Management

    Once all sections for the Expedition have been sent through the track, all data needs to be placed in the appropriate folders on data1 (S:\data1). Be sure to include AUX_DATA.


    1. Copy RSC files from archive and place them in the 7.1 Petrophysics SHMSL – RSC reflectance spectrophotometry colorimetry folder. Confirm relocation. Delete all RSC files off the local drive.
    2. Copy MSPOINT files from archive an place them in the 7.2 Petrophysics SHMSL – MSPOINT point susceptibility folder. Confirm relocation. Delete all MSPOINT files off the local drive.
    3. Copy PROFILE files from archive and place them in 7.3 Petrophysics SHMSL – PROFILE split section surface profile folder. Confirm relocation. Delete all PROFILE files off the local drive.

    Health, Safety, & Environment

    Safety

    • Keep extraneous items and body parts away from the moving platform, belt, and motor.
    • The track system has a well-marked emergency stop button to halt the system if needed.
    • Do not look directly into the spectrometer light source.
    • Do not look directly into the laser light source.
    • Do not attempt to work on the system while a measurement is in progress.
    • Do not lean over or onto the track.
    • Do not stack anything on the track.
    • This analytical system does not require personal protective equipment.

    Pollution Prevention

    This procedure does not generate heat or gases and requires no containment equipment.

    Waste Management

    Dispose of soiled GLAD® Plastic Wrap in an approved waste container.

    Maintenance/Troubleshooting

    Common Issues


    Problem

    Possible Causes

    Solution

    MS sensor: baseline drift

    Temperature variations

    Ensure operating temperature is constant and preferably cool

    Make sure samples have equilibrated to room temperature

    MS sensor: systematic offset from MS2C data

    MS2K/MS2E zeroing environment not the same as that for the samples

    Check that the zeroing height is roughly equivalent to the height of a sample, and that no large metal objects have been introduced to the area

    MS sensor: reading fluctuations greater than ±1 least significant digit

    External electrical noise

    Check the operation area for:

    -         large ferrous objects

    -         heavy electrical machinery

    -         radio frequency source devices

     

    Track is “stuck”

    Gantry flag has tripped the end-of-travel limit switch

    Adjust gantry flag and run sample again

    Current limit on motors was exceeded

    Check the motor for LED error indicators. Call PP tech or ET to reset motor controller

    Torque limit on motors was exceeded.

    Call PP tech or ET to reset motor controller

    Power supply voltage too low

    Call PP tech or ET to check power supply input voltage

    RSC values nonsensical/too bright

    Ambient light level too high

    Reduce ambient light level/ensure integration sphere makes flat contact with sample

     

    Laser sensor: no laser light/no laser range data

    Configuration data lost

    Press function button to restore factory default configuration

    Calibration data lost

    Contact manufacturer

    Laser sensor: LED flashes continuously at 1 Hz

    Laser configuration incorrect

    Call PP tech or ET to reset laser sensor

    Scheduled Maintenance
    Daily

    • Keep contact sensors and laser window clean by wiping with a Kimwipe.
    • If necessary, use isopropyl alcohol to remove soil from sensors and laser window.
    • Do not use any other solvent!

    Annually

    • Remove the end covers on the linear actuators and check if the motor belts need tightening.
    • Examine the cable management system for abraded cables or other indications of wear.
    • Remove the top covers of the linear actuators and check the ball screws to see if they need cleaning or additional lubrication.

    When Needed

    • If the reflectance standard becomes nicked or soiled, it can be smoothed, flattened, and cleaned.


    Communication and Control Setup

    Data communication and control is USB based and managed via National Instrument’s Measurement & Automation Explorer (NI-MAX). When you open NI-MAX and expand the Device and Interface section.  The correct communications setup can be found at IMS Hardware Communications Setup.

    Vendor Information and Part Numbers

    Reflectance Spectrometer

    Vendor

    Ocean Optics, Inc.
    www.oceanoptics.com
    info@OceanOptics.com
    727-733-2447

    Parts:

    Spectrometer: PN QE PRO
    Light source: PN HL-2000-FHSA or HL-2000-FHSA-LL
    Spare bulb: PN HL-2000-B or HL-2000-LL-B
    Reflectance standard: PN WS-1-SL

    Magnetic Susceptibility Meter

    Vendor:

    Bartington Instruments, Ltd.
    www.bartington.com
    sales@bartington.com
    44-1993-706565

    Parts:

    Meter: PN MS2
    Sensor: PN MS2K

    Laser Displacement Sensor

    Vendor:

    Acuity Laser Measurement
    www.acuitylaser.com
    702-616-6070

    Parts:

    Sensor: PN AR200

    Related Documentation and Links

    • AccuRange 700 laser displacement sensor data sheet: ar700-data-sheet.pdf
    • Acuity Laser Measurement User Manual: ar700-users-manual.pdf
    • Ocean Optics Halogen Light Source Operation Manual: hl2000fhsa.pdf
    • Ocean Optics Fiber Optic Spectrometer Operation Manual: USB4000operatinginstructions.pdf
    • Microscan MS-4 Ultra-Compact Imager data sheet: ms4spec.pdf
    • Bartington MS2 Magnetic Susceptibility sensors data sheet: MS2 sensors D50020.pdf
    • Bartington MS2 Magnetic Susceptibility system operation manual: OM0408 MS2.pdf
    • "Environmental Magnetic Susceptibility Using the Bartington MS2 System": OM0409 J_Dearing_Handbook_iss7.pdf
    • Hunt, R. W. (1998). Measuring Colour (3rd ed.). England: Fountain Press. ISBN 0-86343-387-1.. See pgs. 39–46 for the basis in human eye physiology of three-component color models, and 54–57 for chromaticity coordinates.
    • Hunter Lab application note Vol 8. No. 1 http://www.hunterlab.com/appnotes/an04_96a.pdf