Prepare reagents
Prepare the following reagents according to the recipes in the user guides. Remember to put your initials and the date on all reagents made.
- Alkalinity/pH/chloride titrations
- Coulometer
- IC
- ICP SPEC
- HCl solution, KOH and AgNO3 if needed
- IC
- all solutions except for anion eluent, which should be made shortly before analysis
- pourable calibration standards
- ICP
- 2% and 3% TM nitric acid solutions, also 10% if doing hard rock
- internal standard
- primary in-house cocktail
- pourable in-house and IAPSO calibration standards
- SPEC
- Ammonium
- Ammonium standard and calibration standards, alkaline solution
- Phosphate
- Phosphate primary standard (dry monobasic potassium phosphate in oven overnight) + calibration standards
- Sulfuric acid solution
- Antimony potassium tartrate solution
- Ammonium molybdate solution
- Ammonium
Check gas bottles
Check the pressure of the oxygen (ChemLab), helium and argon bottles (Tween Pallet Stores). Ensure the bottles are above 200 psi; otherwise, swap them out with fresh bottles. When a Pallet Stores bottle array is empty, tear the yellow tags to EMPTY, disconnect the manifold and connect it to full bottles. It is good practice to always have a full gas bottle array ready to go.
...