When the shatterbox cycle is done, open the lid and remove the vessels, placing them on the counter. Open the grinding vessel and with clean tweezers take a bit of the powder and feel it against the inside of your wrist. The sample should feel like talc powder. If it does not, repeat the shatterbox cycle. A 1-3 minute cycle is enough for most of rocks.

Transfer powder into Vial

Disassemble the vessel carefully wearing Nitrile gloves. Clean off any powder on the lid or puck with a Kimwipe or Teflon spatula. Carefully remove the puck from the vessel. Pour the sample onto a clean piece of weigh paper. If any powder remains, use a cleaned Teflon spatula to dislodge it.
Never use metal to dislodge or scrap out sample material, as any grooves or scratches in the vessels will increase the risk of contamination.

...