I. Introduction


Thermal conductivity is the coefficient of proportionality relating conductive heat flow to a thermal gradient. The Teka Berlin TK04 system determines thermal conductivity based on a transient heat flow method. A line source is heated with constant power while recording source temperature. Thermal conductivity is calculated from the resulting heating curve.
The TK04 uses two types of probes: the full-space (VLQ) needle probe for soft sediments and the half-space (HLQ) probe for hard rock samples. Measuring a single point in a section takes ~54 min per sample, allowing for 3 replicates to be taken. A self

TK04 Thermal Conductivity: User Guide

...

Author:

...

M. Hastedt, Y.-G. Kim, M.A. Kominz

...

Reviewer(s):

...

T. Gorgas, M. Vasilyev, R. Wilkens, K. Milliken, H. Barnes, S. Hermann; T. Cobb

...

Management Approval (Name, Title, Date):

...

D.J. Houpt (Supervisor of Analytical Systems)

...

Audience:

...

Scientists, Technicians

...

Origination date:

...

2/14/2009; 11/14/2009

...

Current version:

...

V378P | 372

...

Revised:

...

372|V371T|no change 03/18

...

Domain:

...

Physical Properties

...

System:

...

Thermal Conductivity

...

Keywords:

...

Full-space needle probe, half-space needle probe, TK04

Content

Table of Contents
maxLevel1

I. Introduction

II. Procedures

A. Preparing the Instrument

B. Instrument Calibration

C. Set Measurement Parameters

D. Preparing Sections// Preparing Samples

E. Making a Measurement

F. Evaluating your Measurement

G. IMS Utilities

III. Uploading Data to LIMS

A. Data Upload Procedure

B. View and Verify Data

C. Retrieve Data from LIMS

IV. Important Notes

V. Appendix

A.1 Health, Safety & Environment

A.2 Maintenance and Troubleshooting

B.1 IMS Program Structure

a) IMS Program Structure

b) Communication and Control Setup

B.2 Motion Control Setup

C.1 Hardware

VI. Credits

VII. Archived Versions

Introduction

Thermal conductivity is the coefficient of proportionality relating conductive heat flow to a thermal gradient. The Teka Berlin TK04 system determines thermal conductivity based on a transient heat flow method. A line source is heated with constant power while recording source temperature. Thermal conductivity is calculated from the resulting heating curve.
The TK04 uses two types of probes: the full-space (VLQ) needle probe for soft sediments and the half-space (HLQ) probe for hard rock samples. Measuring a single point in a section takes ~54 min per sample, allowing for 3 replicates to be taken. A self-test including a drift study is conducted at the beginning of each cycle. To measure thermal conductivity the heater circuit is closed and the temperature rise in the probe is recorded. Thermal conductivity is calculated from the rate of temperature rise while the heater current is flowing. The thermal conductivity of each sample is the average of three repeated measurements for the full-space method and three to six repeated measurements for the half-space method.
Precision of the method is better than 2%, based on extended evaluation of the method; accuracy is about 5% because of random variations of thermal conductivity in natural materials.

Theory of method


Thermal conductivity is measured by transient heating of an isotropic material with a known heating power generated from a source of known geometry and measuring the temperature change with time. The needle probe contains a heater wire and calibrated thermistor. It is assumed to be a perfect conductor because it is much more conductive than unconsolidated sediments. With this assumption, the temperature of the probe has a linear relationship with the natural logarithm of the time after initiation of heating:

...

Thermal conductivity is an intrinsic material property for which the values depend on chemical composition, porosity, density, structure, and fabric of the material. Thermal conductivity profiles are used along with in situ temperature measurements to determine heat flow, which is an indicator of age of ocean crust and fluid circulation processes.


Available Probe Types


All probes consist of a source (i.e., a metal needle with an embedded heating wire and a temperature sensor), a handle or body (depending on the probe type), and a connection cable.
Full-space probes (VLQ) are needle probes equipped with a handle at one end of the source. They are completely inserted into the sample.
Half-space probes (HLQ), or "Pucks", are placed on top of the sample. The source is embedded into the bottom side of a puck-like probe body and has on-site contact with the material/sample.

Name

Standard VLQ

Standard HLQ

Mini HLQ

Probe type:

Full-space

Half-space

Half-space

Dimension (source, mm):

L: 70 Γ— diameter (D): 2

L: 70 Γ— D: 2

L: 45 Γ— D: 1.5

Dimension (handle/body, mm):

L: 90 Γ— D: 16

L: 30 Γ— D: 88

L: 30 Γ— D: 50

Evaluation parameter set:

Standard VLQ (VLQ Source 70x2)

Standard HLQ (HLQ D88 Source 70x2)

Mini HLQ (HLQ D50 Source 45x1.5)

Measuring range (W/mΒ·K):

0.1–10

0.3–10

0.3–3

Accuracy (%):

Β±2

Β±2

Β±5

Duration of 1 measurement (s):

80

80

60

Min. sample size (mm):

L: 75 Γ— D: 30

L: 15 Γ— D: 80

L: ~15 Γ— D: 50

Picture:

Image Modified

Image Modified

Image Modified


Analytical process

The approximate amount of time needed per sample is as follows:

Process

Time (min)

Comments

1. Obtain a whole-round core section from the core rack

0.3

See Preparing Samples for Analysis


2. Locate the appropriate probe for the sample type

0.5

3. Verify sample identification in software

0.5

See Configuring Measurement Program


4. Configure measurement program

0.3

5. Perform drift control

5

See Measuring Samples




6. Heat and measure sample

2

7. 10 minute pause between measurements

10

8. Repeat steps 5-7 for 2 additional measurements (3 total)

34

9. Upload results to LIMS

0.2

See Uploading Data to LIMS and Verifying Data in LIMS



10. Check results in LIMS

1

11. Remove the section and deliver to splitting room

0.2

Total Time per sample:

54 (max)

0.2

Total Time per sample:

54 (max)


II. Procedures


A. Preparing the Instrument



B. Instrument CalibrationΒ 


The TK04 doesn't need to be calibrated. In order to verify that measurements taken are in the correct range of values, MACOR standards are measured at the start of each expedition.


Measuring the Standard

  1. Prior to initial testing of received cores at a site, the TK04 system should be tested and calibrated to ensure that there are no potential mechanical or software issues. Additional tests using the Standards should be run as part of the troubleshooting process if you experience issues during actual testing (See Troubleshooting).
  2. The Macor Standard for the Standard VLQ consists of its black holding shell, while the Macor Standard for both the Standard HLQ and Mini HLQ is a white disc. Calibration tests for any of the available probe types should provide results of TC=1.626-1.637Β±2%. The Macor standard drift calculations are based on a Macor standard (1.637 Β± 0.033 W/mK) because its properties are closest to basalt cores (See Appendix: TK04 Recommended Heating Power for information).

    Standard VLQ- MACOR Standard

    Standard MACOR Disc

    Image Modified

    Image Modified

    Probe Test TC Value Expected Results:
    TC= 1.626Β±2% W/mK

    Probe Test TC Value Expected Results:
    TC= 1.637Β±2% W/mK


  3. To conduct a probe test, scan the STND MACOR disc TCON (H) label kept above the testing apparatus and ensure that the appropriate heating time and drift control (DCL) settings are input under the Configuration settings (See Configuring the Measurement Program).

    Example 1: A Standard HLQ properly positioned on the Standard MACOR Disc.

    Image Modified


  4. Once the proper settings are confirmed, you can test the probes on the standard as if it were a normal sample. For the Standard VLQ this consists of leaving the probe needle in the MACOR standard, while for the HLQ probes you will need to attach it to the Standard MACOR Disc with a rubber band (See Example 1).



The user should note that in many cases, the samples are not amenable to thermal conductivity analysis, whether because of fractures that lead to circulation, poor surface conditions, or other factors. If the Macor standard gives good results and a core section or piece does not, the sample may simply not work.

Preparing Samples for Analysis

Soft-Sediment Samples- Standard VLQ (Needle Probe)

  1. Equilibrate core sections to room temperature for at least 4 hours in the core rack before bringing a target section to the thermal conductivity workstation.
  2. Select measuring points in the core
    1. Intact core: middle of section; record offset in cm
    2. Cracked core: just above/below the middle (but in any case away from the crack); record offset in cm
  3. Use the cordless drill to drill a ~2 mm hole into the core liner at the borderline between working and archive halves. If the sediment is unconsolidated, drill only through the core liner. If the sediment is semiconsolidated, drill a small hole in the sediment for the probe as well.
    Caution: it is very easy to bend the needle on the VLQ full space needles!
  4. Optional - Apply thermal joint compound to the probe unless the sample is very soft and/or moist.
  5. Carefully insert a clean full-space needle into the sediment. Avoid twisting the needle into the core. The needle must be completely inserted into the sample up to the handle. Do not attempt to force the needle into the sample if the resistance is too great; DO NOT BEND THE NEEDLE.

Image Removed
Figure 1. Thermal conductivity measurement on a soft-sediment section using full-space probe.

Soft-Sediment Samples- HLQ ("Puck" Probe)

  1. Certain sediment samples are too hard for the VLQ Needle Probe to be safely inserted without risking damage to the probe. In these cases, the samples should be split and an appropriate section of core tested with the HLQ "Puck" Probe. If the curator has not designated the core as being a hard rock sample then the ambient temperature seawater bath can be ignored in favor of normal core equilibration on the core rack for the standard four hours.
  2. Select measuring points in the core:
    1. Probe should be positioned so that the embedded needle sits flush with the split surface with no gaps. DI Water can be used to aid in creating a proper conductive connection between the probe needle and the core.
    2. Pieces must be at least 10 cm long. The sample diameter has to be at least equal to the probe diameter and the sample should be at least 2–3 cm thick. The needle on the HLQ half-space needle pucks must be in contact with the sample material on its whole length.
  3. Use rubber bands to secure the puck to the specimen and run test with appropriate configuration input (See Configuring Measurement Program section).

Hard Rock Samples

  1. Place hard rock samples in an ambient temperature seawater bath to equilibrate and saturate (4–12 hr). Keep the sample saturated until measurement. A bell jar and vacuum pump are available in the Physical Properties lab and can be used to aid in saturation.
  2. If the surface of the split core is excessively rough, use a lap plate and grit from the Thin Section Lab to prepare a smooth surface on a split-core piece. Pieces must be at least 10 cm long. The sample diameter has to be at least equal to the probe diameter and the sample should be at least 2–3 cm thick. The needle on the HLQ half-space needle pucks must be in contact with the sample material on its whole length.
  3. Equilibrate the sample and sensor needle in an insulated seawater bath for at least 15 min prior to measurement. Do not submerge the puckβ€”let the water level rise to half the depth of the puck or less.
    Caution: If the water rises to the top of the HLQ puck, it will damage the electrical fittings!
  4. Use rubber bands to secure the puck to the specimen.

Configuring Measurement Program

  1. Load ThermCon software in offline mode. Ensure that the Text_ID field is blank.
    Image Removed
  2. Scan the core label using a scanner, then click Verify Sample.
    Image Removed
  3. If login is requested, enter UserName and Password and then click OK.
    Image Removed
  4. The folder path is shown on the screen. Do not close this window during measurement.Β 
    Image Removed
  5. Run TK04 program and choose Measuring > Configuration.Β 
    Image Removed
  6. Set configuration parameters as follows (see figure below):
    1. Probe Number: serial number of probe to be used in the measurement (Note: results may be wrong by several percent if the wrong serial number is entered or by a factor of ~2 if the wrong type of probe is entered),
    2. Root Name: six characters or less; suggest Core-Type-Section (no special characters in the root name).
    3. Serial Number: number of repeat measurements at each measurement point (1–99 single measurements).
    4. Folder: path for saving data results.
    5. Heating Power: for the VLQ (needle probe), set to twice the estimated thermal conductivity value of measured sediment. For example, 2–3 is good for sediment. (See the Appendix: TK04 Recommended Heating Power for power guidance.)
    6. Measuring Time: set to at least 80 s, or for mini HLQ 60 s.
    7. Click Expert Options to configure Drift Control and Pause in Minutes (see Step 7).
    8. Enter comments.

Image Removed
To configure Drift Control and Pause in Minutes in Expert Options:

  • Drift Control (DCL): limit for the range of temperature drift allowed prior to heating and measuring. A larger number allows quicker but less accurate measurement. Default (unchecked) DCL = 10; Recommended DCL = 40.
  • Pause in Minutes: insert a pause between single measurements; recommended = 10 min. This parameter does not apply when conducting a single measurement on each core or each core section.

...

fractures that lead to circulation, poor surface conditions, or other factors. If the Macor standard gives good results and a core section or piece does not, the sample may simply not work.

C. Set Measurement Parameters


Configuring Measurement Program

  1. Load ThermCon software in offline mode. Ensure that the Text_ID field is blank.
    Image Added
  2. Scan the core label using a scanner, then click Verify Sample.
    Image Added
  3. If login is requested, enter UserName and Password and then click OK.
    Image Added
  4. The folder path is shown on the screen. Do not close this window during measurement.Β 
    Image Added
  5. Run TK04 program and choose Measuring > Configuration.Β 
    Image Added
  6. Set configuration parameters as follows (see figure below):
    1. Probe Number: serial number of probe to be used in the measurement (Note: results may be wrong by several percent if the wrong serial number is entered or by a factor of ~2 if the wrong type of probe is entered),
    2. Root Name: six characters or less; suggest Core-Type-Section (no special characters in the root name).
    3. Serial Number: number of repeat measurements at each measurement point (1–99 single measurements).
    4. Folder: path for saving data results.
    5. Heating Power: for the VLQ (needle probe), set to twice the estimated thermal conductivity value of measured sediment. For example, 2–3 is good for sediment. (See the Appendix: TK04 Recommended Heating Power for power guidance.)
    6. Measuring Time: set to at least 80 s, or for mini HLQ 60 s.
    7. Click Expert Options to configure Drift Control and Pause in Minutes (see Step 7).
    8. Enter comments.

Image Added
To configure Drift Control and Pause in Minutes in Expert Options:

  • Drift Control (DCL): limit for the range of temperature drift allowed prior to heating and measuring. A larger number allows quicker but less accurate measurement. Default (unchecked) DCL = 10; Recommended DCL = 40.
  • Pause in Minutes: insert a pause between single measurements; recommended = 10 min. This parameter does not apply when conducting a single measurement on each core or each core section.

Image Added


D. Preparing Sections & Samples


Soft-Sediment Samples- Standard VLQ (Needle Probe)


  1. Equilibrate core sections to room temperature for at least 4 hours in the core rack before bringing a target section to the thermal conductivity workstation.
  2. Select measuring points in the core
    1. Intact core: middle of section; record offset in cm
    2. Cracked core: just above/below the middle (but in any case away from the crack); record offset in cm
  3. Use the cordless drill to drill a ~2 mm hole into the core liner at the borderline between working and archive halves. If the sediment is unconsolidated, drill only through the core liner. If the sediment is semiconsolidated, drill a small hole in the sediment for the probe as well.
    Caution: it is very easy to bend the needle on the VLQ full space needles!
  4. Optional - Apply thermal joint compound to the probe unless the sample is very soft and/or moist.
  5. Carefully insert a clean full-space needle into the sediment. Avoid twisting the needle into the core. The needle must be completely inserted into the sample up to the handle. Do not attempt to force the needle into the sample if the resistance is too great; DO NOT BEND THE NEEDLE.

Image Added
Figure 1. Thermal conductivity measurement on a soft-sediment section using full-space probe.

Soft-Sediment Samples- HLQ ("Puck" Probe)


  1. Certain sediment samples are too hard for the VLQ Needle Probe to be safely inserted without risking damage to the probe. In these cases, the samples should be split and an appropriate section of core tested with the HLQ "Puck" Probe. If the curator has not designated the core as being a hard rock sample then the ambient temperature seawater bath can be ignored in favor of normal core equilibration on the core rack for the standard four hours.
  2. Select measuring points in the core:
    1. Probe should be positioned so that the embedded needle sits flush with the split surface with no gaps. DI Water can be used to aid in creating a proper conductive connection between the probe needle and the core.
    2. Pieces must be at least 10 cm long. The sample diameter has to be at least equal to the probe diameter and the sample should be at least 2–3 cm thick. The needle on the HLQ half-space needle pucks must be in contact with the sample material on its whole length.
  3. Use rubber bands to secure the puck to the specimen and run test with appropriate configuration input (See Configuring Measurement Program section).

Hard Rock Samples


  1. Place hard rock samples in an ambient temperature seawater bath to equilibrate and saturate (4–12 hr). Keep the sample saturated until measurement. A bell jar and vacuum pump are available in the Physical Properties lab and can be used to aid in saturation.
  2. If the surface of the split core is excessively rough, use a lap plate and grit from the Thin Section Lab to prepare a smooth surface on a split-core piece. Pieces must be at least 10 cm long. The sample diameter has to be at least equal to the probe diameter and the sample should be at least 2–3 cm thick. The needle on the HLQ half-space needle pucks must be in contact with the sample material on its whole length.
  3. Equilibrate the sample and sensor needle in an insulated seawater bath for at least 15 min prior to measurement. Do not submerge the puckβ€”let the water level rise to half the depth of the puck or less.
    Caution: If the water rises to the top of the HLQ puck, it will damage the electrical fittings!
  4. Use rubber bands to secure the puck to the specimen.


E. Making a Measurement


Measuring Samples

  1. Confirm configuration settings shown in the lower part of the TK04 screen, insert the probe into the hole drilled into the sample, and click Start Measuring.Β 
  2. Drift control (DCL) repeats until the criterion for drift (set in the Expert Options window) is met. Each drift control measurement takes 0.5 min. For DCL = 40, drift control is <10 series (~5 min). For DCL = 10, drift control is <30 (~15 min).
    1. Counter: remaining number of measurements in the current drift series
    2. Recording: currently recorded time and temperature and drift series being recorded
    3. Drift: drift value from the last drift series; indicated in blue on the slider scale when acceptable
    4. Start signal light:
      1. Yellow: drift has reached the threshold; approximately half the drift time has passed
      2. Green: drift limit has been reached and measurement can begin
  3. After satisfying drift control, sample heating and measuring begins, and temperature values are corrected automatically for the drift effect predicted from the last drift series. Elapsed measurement time is controlled by the value entered in Configuring Measurement Program > Step 6F.
    1. Recording: displays currently recorded time and temperature
    2. Data: lists recorded time/temperature values from the heating curve
    3. Heating diagram: continuously updated; can display as linear or logarithmic scale
  4. Solutions calculated by the algorithm are shown on the screen. Note the result on the handwritten log.
    1. TC: thermal conductivity
    2. LET: logarithm of the extreme time (lower limit = 4); the measurement with the largest LET is used to calculate thermal conductivity
    3. CV: contact value
    4. PC: power control is shown on the plot in the lower part of the screen (recommended PC = 2–3. If PC is out of range, adjust the Heating Power (HP) in Configuring Measurement Program > Step 6E as follows: if PC > 3, decrease HP; if PC < 2, increase HP).
    5. Mean: mean thermal conductivity
    6. Count: number of measurements used to calculate thermal conductivity

...


F. Evaluating your Measurement



G. IMS Utilities



III. Uploading Data to LIMS




A. Data Upload Procedure


  1. To upload results to LIMS, click Upload to LIMS. If upload is successful, a message like "Logged results for sample …" is shown. Close the ThermCon program.

...


B. View and Verify Data


Verifying Data in LIVE (LIMS Viewer)

...

Note: The "expanded" report shows all of the database parameters and may be confusing to a general user; use the "standard" report.

3. Under Select Sample Range, specify Expedition, Site, Hole, and Section image(s) to retrieve.

4. Click View data or Download data file to view results or download a CSV file.

After Verifying Data Upload

...

Once all sections for the Expedition have been sent through the track, all data needs to be placed in the appropriate folders on data1 (S:\data1).
1. Copy files from archive and place them in the 5.1 Petrophysics TCON thermcon folder. Confirm relocation. Delete all files off the local drive.

LIMS Component descriptions

...

Analysis

...

Component

...

Unit

...

Definition

...

TCON

...

Bottom_depth

...

m

...

Location of bottom of measurement, measured from the top of the hole

...

Comment

...

None

...

Comment about the run

...

Contact_value

...

None

...

Measure of contact quality between probe and sample

...

End_time

...

s

...

Elapsed time for end of analysis window

...

Heating_power

...

W/m

...

Power applied to needle during heating

...

Length_of_time

...

s

...

Elapsed time, start to finish, of analysis

...

Log_extreme_time

...

s

...

LET, used in calculation algorithm

...

Method

...

None

...

Data reduction method: SAM or TCON

...

Needle_name

...

None

...

Full-space or half-space

...

Number_of_solutions

...

None

...

Number of solutions found by the software

...

Offset

...

cm

...

Location of measurement from top of section

...

Start_time

...

s

...

Elapsed time into experiment for start of analysis window

...

Therm_con_average

...

W/(mΒ·K)

...

Mean thermal conductivity result

...

Therm_con_number

...

None

...

Number of measurements in the population

...

Therm_con_result

...

W/(mΒ·K)

...

Individual thermal conductivity result

...

Therm_con_stdev

...

W/(mΒ·K)

...

Standard deviation (n-1) of measurement population


LIMS Component descriptions


Analysis

Component

Unit

Definition

TCON

Bottom_depth

m

Location of bottom of measurement, measured from the top of the hole

Comment

None

Comment about the run

Contact_value

None

Measure of contact quality between probe and sample

End_time

s

Elapsed time for end of analysis window

Heating_power

W/m

Power applied to needle during heating

Length_of_time

s

Elapsed time, start to finish, of analysis

Log_extreme_time

s

LET, used in calculation algorithm

Method

None

Data reduction method: SAM or TCON

Needle_name

None

Full-space or half-space

Number_of_solutions

None

Number of solutions found by the software

Offset

cm

Location of measurement from top of section

Start_time

s

Elapsed time into experiment for start of analysis window

Therm_con_average

W/(mΒ·K)

Mean thermal conductivity result

Therm_con_number

None

Number of measurements in the population

Therm_con_result

W/(mΒ·K)

Individual thermal conductivity result

Therm_con_stdev

W/(mΒ·K)

Standard deviation (n-1) of measurement population

Top_depth

m

Location of top of measurement from top of hole

C. Retrieve Data from LIMS



IV. Important Notes




V. Appendix


A.1 Health, Safety & Environment

Safety

  • Keep extraneous items and body parts away from the moving platform, belt, and motor.
  • The track system has a well-marked emergency stop button to halt the system if needed.
  • Don't insert your hand between the GRA source and the detector.
  • Do not look directly into the laser light source.
  • Do not attempt to work on the system while a measurement is in progress.
  • Do not lean over or onto the track.
  • Do not stack anything on the track.
  • This analytical system does not require personal protective equipment.


Pollution Prevention

This procedure does not generate heat or gases and requires no containment equipment.

A.2 Maintenance and Troubleshooting

...

Top_depth

...

m

...


Troubleshooting


Drift phase takes too long:

  • Ambient temperature is not stable: allow sample more time to equilibrate
  • Probe and sample are not in equilibrium with ambient temperature: place sample/probe in insulated case
  • If necessary, force measurements by choosing a weaker drift limit

...

  • Water-saturated samples drying out: keep sample wet during analysis

...


C.1 TK04 Recommended Heating

...

PowerΒ 


Note: for loose sediments, use a lower heating power to avoid convective heat transport of pore fluids.


Material


Thermal Conductivity (W/mΒ·K)


Recommended Heating Power (W/m)


Mean

Range

VLQ

HLQ

Wood

0.21

0.1–0.35

0.15–1.3

β€”

Coal

0.29

0.1–1.5

0.15–5.4

β€”

Concrete

1.00

0.75–1.4

1.0–5.0

0.5–2.2

Fused silica

1.40

1.33–1.46

1.8–5.2

0.8–2.3

Clay

1.40

1.2–1.7

1.6–6.1

0.7–2.6

Silt

1.60

1.4–2.1

1.9–7.5

0.8–3.2

Basalt

1.95

1.4–5.4

1.9–19.0

0.8–7.6

Siltstone

2.04

0.6–4.0

0.8–14.0

0.4–5.7

Limestone

2.29

0.5–4.4

0.7–16.0

0.4–6.3

Syenite

2.31

1.3–5.3

1.7–19.0

0.8–7.5

Amphibolite

2.46

1.4–3.9

1.9–14.0

0.8–5.6

Claystone

2.46

1.6–3.4

2.1–12.0

0.5–9.3

Lava

2.47

0.2–4.5

0.3–16.0

0.2–6.4

Gabbro

2.50

1.6–4.1

2.1–15.0

0.9–5.9

Dolerite (Diabase)

2.64

1.6–4.4

2.1–16.0

0.5–6.3

Granodiorite

2.65

1.3–3.5

1.7–13.0

0.8–5.0

Quartz sand (wet)

2.70

2.4–3.1

3.2–11.0

1.3–4.5

Marble

2.80

2.1–3.5

1.8–13.0

1.2–5.0

Porphyrite

2.82


3.8–10.0

1.5–4.2

Boulder clay

2.90

2.5–3.3

3.4–12.0

1.4–4.8

Diorite

2.91

1.7–4.2

2.3–15.0

1.0–6.0

Slate (perpendicular)

2.91

1.5–3.9

2.0–14.0

0.9–5.6

Gneiss

2.95

1.2–4.7

1.6–17.0

0.7–6.7

Granite

3.05

1.2–4.5

1.6–16.0

0.7–6.4

Eclogite

3.10

2.4–3.4

3.2–12.0

1.3–4.9

Andesite

3.20

1.6–4.7

2.1–17.0

1.0–6

.7

Dolomite

3.62

1.6–6.6

2.1–20.0

1.0–9.3

Slate (parallel)

3.80

2.2–5.2

.7

Dolomite

3.62

1.6–6.6

2.1–20.0

1.0–9.3

Slate (parallel)

3.80

2.2–5.2

3.0–19.0

1.2–7.4

Peridotite

3.81


5.0–14.0

2.0–5.5

Anhydrite

4.05

1.0–6.0

1.3–20.0

0.6–8.5

Pyroxenite

4.27

3.2–5.1

4.3–18

3.0–19

.0

1.

2–7

7–7.2

Dunite

4.41

Peridotite

3.

81

5–5.2

5

4.

0–14

7–19.0

2

1.

0–5

9–7.

5

4

Anhydrite

Quartzite

4.55

3.

05

1–>8

1

4.

0–6

2–20.0

1.

3–20

7–11.0

0.6–8.5

Pyroxenite

4.27

3.2–5.1

4.3–18.0

1.7–7.2

Dunite

4.41

3.5–5.2

4.7–19.0

1.9–7.4

Quartzite

4.55

3.1–>8

4.2–20.0

1.7–11.0

Quartz

9.50

6.5–12.5

8.7–20.0

3.5–17.0

Quartz

9.50

6.5–12.5

8.7–20.0

3.5–17.0


VI. Credits



This document originated from 2009, TK04 UG v.,V378P | 372 (Revised: 372|V371T|no change 03/18Β  ), that had contributions from the authors Hastedt, Y.-G. Kim, M.A. Kominz, and the reviewers David Houpt, T. Gorgas, M. Vasilyev, R. Wilkens, K. Milliken, H. Barnes, S. Hermann; T. Cobb. Credits for subsequent changes to this document are given in the page history.