- Text ID: e.g., WDGE11258831 (Note: capital letter matters)
- Analysis: LOI
- Replicate: 1 (usually) leave blank, unless true replication (not just reanalyzing) of a sample, then begin replicate numbering with '0', then '1', then '2'.
- crucible_number: Crucible ID (e.g., A, B, F...)
- crucible_preignition_mass: crucible mass only (small inner crucible) (component unit: grams)
- crucible_and_preignition_sample_mass: inner crucible with fresh sample before ignition (component unit: grams)
- crucible_and_postignition_sample_mass: inner crucible with ignited sample after ignition (component unit: grams)
- loi_percent: LOI in % (component unit: none; the value is already a percentage)
- reference_mass: sample mass (calculated difference) (component unit: grams)
- comment: specify temperature and time used for LOI (e.g., 950 deg C for 4 hours; used during Exp 391)
Text ID | Analysis | Replicate | Instrument | Display Flag | Component Name | Component Value | Component Unit | Component Name | Component Value | Component Unit | Component Name | Component Value | Component Unit | Component Name |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
QRND12345678 | LOI1 | crucible_number | NONE | crucible_preignition_mass | GRAMS | crucible_and_preignition_sample_mass | GRAMS | crucible_and_postignition_sample_mass |
...
- Wash the crucibles with DI water and a small piece of a scouring pad (no soap).
- Rinse several times with DI water.
- Place crucibles in a 10% HNO3 bath for 12 hr. If in urgent need, the quartz crucibles can be soaked a minimum of 2 hours, but extra flux blanks will need to be made and added to batches that use these crucibles.
- Rinse the quartz crucibles 3 times with DI water after the acid bath. For the ceramic crucibles, rinse 3 times with DI and then soak them in a large beaker of DI water for 4-6 hours to dilute the acid absorbed by the crucibles.
- Dry the either type of crucibles in the oven at a maximum temperature of 60°C. For the ceramic crucibles, heat them to 800°C in the muffle furnace for 2 hours and allow to cool to room temperature before removing.
...