   CAUTION: DO NOT place anything, including samples, on the track below the instruments before opening the program. The program will lower all transducers upon startup, and CAN CAUSE SEVERE INJURY.

  1. Double-click the MUT icon on the desktop (Figure 1a) and login using ship credentials. For more information on data uploading see the "Uploading Data to LIMS" section below.
  2. Double-click the IMS icon (Figure 1b). IMS initialize's initializes the instrument and checks the movement of the the three sets of instruments. Once initialized, the logger is ready to measure the first sample.

...

  1. Make sure the sensors are clean.
  2. For section halfshalves, place the desired measurement location underneath the desired sensor. Make sure that the section is oriented with the blue end-cap towards the laser (Figure 2).

...

4. Place one or two drops of deionized (DI) water at any relevant at  sample/sensor interface. It will improve the signal transmission.

...