A common program is an increase in temperature of 3°C/min to a target temperature of 900°C and a hold of one hour. Then ramp up at a rate of 3°C/min to a target temperature of 1025°C and hold for four hours before cooling. This ramp cycle is already programmed into the furnace and corresponds to 'Program 1'. 'Program 2' ramps up at a rate of 3°C/min to 950°C and holds there for 4 hours before cooling.
Oven Programs: (Note: Once a program has finished, the temperature setting returns to room temperature, ~25°C.)
- Program 1: Ramp up to 900°C at 3°C/min and hold for 1 hour, then continue to 1025°C at 3°C/min and hold for 4 hours.
- Program 2: Ramp up to 950°C at 3°C/min and hold for 4 hours.
- Program 3: Ramp up to 550°C at 3°C/min and hold for 1 hour.
- Program 4: Ramp up to 900°C at 3°C/min and hold for 1 hour, then continue to 1025°C at 3°C/min, no hold.
To check or edit a program see additional guides attached to the furnace itself. It is also possible to run the furnace manually without a ramp up cycle. Discuss with scientists their preference.
Either enter the desired temperature manually or select your program. If you are running the furnace manually, enter the desired temperature; no other buttons or steps are needed. If selecting a program, press and hold 'Run' (Figure 40).
...
This document originated from Word document ICP_HR_Prep_UG_376.doc (see Archived Versions below for a pdf copy) that was written by H. Barnes & K. Bronk; later edited by N. Lawler & A. Armstrong. Credits for subsequent changes to this document are given in the page history.
LOI LIMS Component Table
ANALYSIS | TABLE | NAME | ABOUT TEXT |
LOI | SAMPLE | Exp | Exp: expedition number |
LOI | SAMPLE | Site | Site: site number |
LOI | SAMPLE | Hole | Hole: hole number |
LOI | SAMPLE | Core | Core: core number |
LOI | SAMPLE | Type | Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X). |
LOI | SAMPLE | Sect | Sect: section number |
LOI | SAMPLE | A/W | A/W: archive (A) or working (W) section half. |
LOI | SAMPLE | text_id | Text_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples. |
LOI | SAMPLE | sample_number | Sample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table. |
LOI | SAMPLE | label_id | Label identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples. |
LOI | SAMPLE | sample_name | Sample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter. |
LOI | SAMPLE | x_sample_state | Sample state: Single-character identifier always set to "W" for samples; standards can vary. |
LOI | SAMPLE | x_project | Project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises |
LOI | SAMPLE | x_capt_loc | Captured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2 |
LOI | SAMPLE | location | Location: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2 |
LOI | SAMPLE | x_sampling_tool | Sampling tool: sampling tool used to take the sample (e.g., syringe, spatula) |
LOI | SAMPLE | changed_by | Changed by: username of account used to make a change to a sample record |
LOI | SAMPLE | changed_on | Changed on: date/time stamp for change made to a sample record |
LOI | SAMPLE | sample_type | Sample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ) |
LOI | SAMPLE | x_offset | Offset (m): top offset of sample from top of parent sample, expressed in meters. |
LOI | SAMPLE | x_offset_cm | Offset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm) |
LOI | SAMPLE | x_bottom_offset_cm | Bottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm) |
LOI | SAMPLE | x_diameter | Diameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master |
LOI | SAMPLE | x_orig_len | Original length (m): field for the original length of a sample; not always (or reliably) populated |
LOI | SAMPLE | x_length | Length (m): field for the length of a sample [as entered upon creation] |
LOI | SAMPLE | x_length_cm | Length (cm): field for the length of a sample. This is a calculated field (length, converted to cm). |
LOI | SAMPLE | status | Status: single-character code for the current status of a sample (e.g., active, canceled) |
LOI | SAMPLE | old_status | Old status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample |
LOI | SAMPLE | original_sample | Original sample: field tying a sample below the CORE level to its parent HOLE sample |
LOI | SAMPLE | parent_sample | Parent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR) |
LOI | SAMPLE | standard | Standard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T) |
LOI | SAMPLE | login_by | Login by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created]) |
LOI | SAMPLE | login_date | Login date: creation date of the sample |
LOI | SAMPLE | legacy | Legacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition |
LOI | TEST | test changed_on | TEST changed on: date/time stamp for a change to a test record. |
LOI | TEST | test status | TEST status: single-character code for the current status of a test (e.g., active, in process, canceled) |
LOI | TEST | test old_status | TEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test |
LOI | TEST | test test_number | TEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table. |
LOI | TEST | test date_received | TEST date received: date/time stamp for the creation of the test record. |
LOI | TEST | test instrument | TEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input) |
LOI | TEST | test analysis | TEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table) |
LOI | TEST | test x_project | TEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises |
LOI | TEST | test sample_number | TEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table |
LOI | CALCULATED | Top depth CSF-A (m) | Top depth CSF-A (m): position of observation expressed relative to the top of the hole. |
LOI | CALCULATED | Bottom depth CSF-A (m) | Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole. |
LOI | CALCULATED | Top depth CSF-B (m) | Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user. |
LOI | CALCULATED | Bottom depth CSF-B (m) | Bottom depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user. |
LOI | RESULT | crucible_and_postignition_sample_mass (g) | RESULT post-ignition crucible and sample mass (g): mass of the sample plus the crucible after ignition ("ashed") |
LOI | RESULT | crucible_and_preignition_sample_mass (g) | RESULT pre-ignition crucible and sample mass (g): mass of the sample plus the crucible before ignition |
LOI | RESULT | crucible_number | RESULT crucible number: number of the crucible used for this measurement |
LOI | RESULT | crucible_preignition_mass (g) | RESULT crucible preignition mass (g): mass of the empty crucible before ignition |
LOI | RESULT | loi_percent | RESULT loss on ignition (wt%): loss of material after ignition |
LOI | RESULT | reference_mass (g) | RESULT reference mass (g): field for the reference mass used on the reference balance (not always populated) |
LOI | RESULT | ssup_asman_id | RESULT spreadsheet uploader ASMAN_ID: serial number for the ASMAN link for the spreadsheet uploader file |
LOI | RESULT | ssup_filename | RESULT spreadsheet uploader filename: file name for the spreadsheet uploader file |
LOI | SAMPLE | sample description | SAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments" |
LOI | TEST | test test_comment | TEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments" |
LOI | RESULT | result comments | RESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments" |
Archived Versions
LMUG-ICPPreparation-230220-1915-158.pdf
...