Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Calibrating the electrode

...

Standard ratio correction

...

Drift Span

...

Cleaning up after Measurement

...

Data Handling

...

To select a saved standard ratio correction, double-click it.


Drift Span


A drift span of 30 (default) indicates that a minimum of 30 measurements will be taken after each addition of titrant (acid). The difference between the first and last measurements is compared to the drift tolerance specified in the rate program (see Creating a New Dispensing Rate Program). Generally, drift tolerance starts at 0.10 mV/s. Drift tolerance acts as follows:

...

The mV readings come from the final value reached upon satisfying the stability criteria.





Cleaning up after Measurement


1.

Remove electrode from cup.

2.

Rinse electrode with DI water in squeeze bottle.

3.

Blot the electrode dry with a Kimwipe. Do not rub the electrode, which could cause static charge buildup.

4.

Place the electrode in a storage container containing 3 M KCl.

5.

Rinse and dry the acid dispensing tip and stir bar.

6.

Pipette or pour the titrated alkalinity IW sample into a container to ship to repository or scientist.

7.

Write on label how much HCl was added during titration.

8.

Rinse the titration vessel cup with DI water and dry.


Data Handling


Editing Gran Factor Points

...



Anchor
_Ref302373996
_Ref302373996
Figure 10. Gran Factor Points (blue = Gran factor F, red = optimum mV range for linearity: 220–240 mV).


Data Reports


Data reports are not available at the instrument, but data can be viewed on the View Datalog tab (Figure 11). Also, the software appends the alkalinity results to the DAT file C:\ALKALINITY\LOG\ALKALINITYSAMPLES.DAT. The best way to view the alkalinity values to be used in graphs and reports is via LIMS Reports. It is also highly recommended to record the alkalinity and pH values in the blue laboratory notebook to protect against inadvertent data loss.

...

Analysis

Component

Unit

Description

ALKALINITY

alkalinity

mM

Ability of a solution to neutralize acid to the equivalence point of carbonate

pH

NA

Acidity or basicity of a solution (–log[H+])

acid_quantity

µL

Amount of 0.1 M HCl added to the sample during titration

correction_factor

Correction factor for non-ideal behavior of samples to adjust calibration

ALK_QAQC

alkalinity

mM

Ability of a solution to neutralize acid to the equivalence point of carbonate

pH

NA

Acidity or basicity of a solution (–log[H+])

acid_quantity

µL

Amount of 0.1 M HCl added to the sample during titration

correction_factor

Correction factor for non-ideal behavior of samples to adjust calibration

slope

Slope of the calibration equation

intercept

Intercept value of the calibration equation

corr

Rho

Correlation coefficient of the calibration


Uploading Data to LIMS


When the alkalinity titration finishes, the GRAN-METHOD screen shows the slope information, correction factors, and final alkalinity value (Figure 14).
Edit outlier Gran factor points, if necessary (see Editing Gran Factor Points). If satisfied with the results, click Ok/Save to load the values into LIMS.

...

Health, Safety, & Environment


Safety


Wear personal protective equipment including close-toed shoes, lab coat, gloves and safety glasses when working with acids of any strength.
Use a fume hood when making solutions from concentrated acids.


Pollution Prevention


Make reagent and standard solutions in quantities no larger than needed to complete sample analysis.


Waste Management


Neutralize acid solutions before disposal.

...

Maintenance/Troubleshooting

...



Make sure the probe storage container is filled with 3 M KCl solution and the reservoir is filled with 0.1 M HCl solution.

Ensure no air bubbles are present in the acid dispensing line. Press DOS on the body of the titrator to push acid through the line and remove air bubbles.

Change out the Drierite trap when ~50% of the color turns from blue to pink.

In case of a bubbling noise, top up your temperature-controlled water bath. Use tap water. Check titration vessel for air bubble and eliminate if present.


Consumables


Electrode: Metrohm 6.0234.100 or equivalent.

0.1   M HCl, Fisher CH5009

Microvalve buret tip, Metrohm 020683244, CM5129

...