Make sure the probe storage container is filled with 3 M KCl solution and the reservoir is filled with 0.1 M HCl solution.

Ensure no air bubbles are present in the acid dispensing line. Press DOS on the body of the titrator to push acid through the line and remove air bubbles.

Change out the Drierite trap when ~50% of the color turns from blue to pink.

If you are hearing In case of a bubbling noise, top up your temperature-controlled water reservoirbath. Tap Use tap water is fine. Check titration vessel for air bubble and eliminate if present.

...