Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.



Table of Contents

Table of Contents
maxLevel3
excludeTable of Contents

Image Added

The P-wave Velocity station.

 

I. Introduction

The P-wave velocity gantry measures the speed at which ultrasonic sound waves pass through materials that are placed between its transducers. The three orthogonal sets of piezoelectric transducers allow the velocity to be determined in the X-, Y-, and Z-directions (Figure 1) on working-half split-core sections. The P-wave bayonets (PWBs) measure the velocity along the core (Z-direction) and across the split-core face (Y-direction), and a P-wave caliper (PWC) measures the velocity perpendicular to the split-core face (X-direction). A

P-Wave Velocity Gantry: User Guide

Manual Information

...

Author(s):

...

T. Cobine

...

Reviewer(s):

...

H. Barnes, T. Gorgas

...

Management Approval:

...

D.J. Houpt (Supervisor of Analytical Systems)

...

Audience:

...

Scientists, Marine Laboratory Specialists

...

Origination date:

...

1/17/09

...

Current version:

...

V2.0

...

Revised:

...

V2.1

...

Domain:

...

Physics

...

System:

...

P-Wave Analysis

User Guide Contents

Table of Contents
maxLevel1

Introduction

...

laser measures the position of the top of the

...

split-core section and calculates the position

...

at which the velocity was measured (this is recorded as Offset in the LIMS Database).

...

For discrete sample cubes, the velocity is measured along each of the three axes separately using the PWC. Mini-cores are measured along the axis of the cylinder (X-direction) using the PWC. For discrete samples cubes and mini-cores, all sample information, including offset, is entered by the user when the sample was taken.



Image Added
Figure 1. Section

...

half measurement directions.

Method Theory

Measurement of P-wave velocity requires an estimate of the

...

travel time and an accurate measurement of the ultrasonic P-wave path length through the sample.
Velocity is defined as follows:
velocity = pathlength/traveltime, or
v = dS/dt.
Traveltime measurement is estimated by an algorithm for graphical first arrival pick. An ultrasonic pulser generates a high-impulse voltage, which is applied to the ultrasonic transmitter and thereby induces oscillation of the crystal element within the transducer-specific frequency band. A trigger pulse from the pulser is then applied to the oscilloscope to record the waveform from the receiving transducer.
By measuring the acoustic traveltime of the waveform through

...

a standard of known pathlength and velocity, the difference in expected travel time and actual travel timeprovides the instrumentation-specific time delay. Subtraction of the system delay time (+ liner material propagation time, if required) from the total traveltime gives the traveltime for the ultrasonic pulse through the sample.
Precise

...

thickness of the sample

...

, or path length, is derived from the readout of

...

an Acuity AR700 laser displacement sensor. The laser offset correction is determined during the calibration process and requires that the system fully close the transducer caliper when the software is opened and activated. Therefore, do not place a core section underneath the transducers when the software is started.
The chisel (bayonet) transducers are fixed at

...

82.32 mm for the z-axis (downhole) and

...

31.70 mm for the y-axis (

...

see figure 1).

...

As mentioned above, the x-axis caliper separation is derived from

...

an Acuity AR700 laser.
Traveltimes for

...

samples are calculated as follows:
x-axis = total traveltime – x-system delay time – liner traveltime (section-halves only)
y-axis = total traveltime – y-system delay time
z-axis = total traveltime – z-system delay time
Liner traveltime is calculated as the liner thickness (typically 2.7 mm) divided by the published liner material velocity (cellulose butyrate = 2140 m/s).

...

P-Wave Measurement

A temperature-equilibrated split core section in a half-core liner is placed on the core track. A barcode reader records all relevant sample information, which is used for the data upload into LIMS. A laser sensor measures the distance to the section end and determines the sampling interval from the difference between the position of the sensors and the end of the section. The user positions the section half under the sensor and triggers the measurement from the software control panel. The measurement is taken for a predetermined amount of time, generally 5 s, with the recorded result representing several thousand determinations.

Caliper Measurement

  • The sample is placed between 2 flat, 1 inch diameter sensors that squeeze firmly onto the specimen to ensure good contact.
  • One sensor acts as a transducer and the other as a receiver to record velocity measurements at a rate of 0.5 MHz.
  • To measure discrete samples, the sample is measured along each (x, y, and z-) axis between the caliper sensors.
  • A series of polycarbonate standards of different thickness are measured to obtain a linear regression transit time vs. distance for calibration. Their industry-calibrated standard sonic velocity is 2.750 m/s.

Bayonet Measurement

  • Two pairs of piezoelectric transducers set at 90° to each other are inserted into the unconsolidated or semisoft section-half materials until the sensors are buried in the material to be measured.
  • One of each pair of sensors acts as transducer and the other as receiver to measure velocity in two directions simultaneously.
  • Calibration is performed by inserting the probes into a container filled with distilled water of known temperature, and therefore known velocity.

Data Quality

Velocity data quality is affected by several variables:

  • Quality of the acoustic coupling between the core material and the sensor transducers. Note: Use water to increase the quality of the contact.
  • Quality of the coupling between both the transducer and the core liner and between the core liner and sample. Note: Use water to increase the quality of the contact.
  • Consolidation of the sediments; noncohesive sediments containing microcracks or gas voids cannot be measured accurately.

Apparatus, Reagents, & Materials

Hardware

The velocity track system consists of the following components (Figure 2):

  • Caliper transducers
  • Chisel-type transducers (bayonet transducers)
  • Linear actuator
  • Laser sensor
  • Barcode reader

...

Caliper Transducers: Panametrics-NDT Microscan Delay Line Transducers

...

Specification

...

Value

...

Frequency (MHz)

...

0.5

...

Element diameter (mm)

...

25

...

Part number

...

M2008

Bayonet Transducers

Custom

Exlar Linear Actuator

...

Specification

...

Value

...

Maximum radial load (lb)

...

15

...

Resolution (revolution)

...

0.001

...

Accuracy (revolution)

...

±0.010

...

Operating temperature (°C)

...

0–55

...

Model Number

...

TLM20-0601-1-IFM-1BS-50-AR

...

Voltage (VDC)

...

48

...

Current (A)

...

5000 @ 10 rpm

Acuity 1000 Laser Distance Sensor

...

Specification

...

Value

...

Distance (m)

...

30

...

Laser

...

650 nm, 1 mW visible red

...

Accuracy

...

±0.12 inch (3.05 mm)

...

Resolution

...

0.004 inch (0.10 mm)

...

Operating temperature (°C)

...

–10–50

...

Linearity/accuracy (mm)

...

±3

Microscan MS-4 Ultracompact Imager Barcode Reader

...

Specification

...

Value

...

Dimensions (mm)

...

25.4 × 45.7 × 53.3

...

Operating temperature (°C)

...

0–40

...

Operating humidity (%)

...

Up to 90

...

Light source

...

High-output LEDs

...

Data types

...

2-D: data matrix, QR

...

 

...

Stacked: MicroPDF, PDF417, RSS

...

 

...

Linear: all standard

...

Read parameters

...

Pitch = ±30°

...

 

...

Skew = ±30°

...

 

...

Tilt = 360°

...

 

...

Decode rate = 10/second

Software

Generally, only the marine technicians should access configurations settings in the velocity software. However, there are some configuration issues that users need to be aware of.
The program icon is shown in Figure 3. Double click to open and follow the prompts to display the program home screen. DO NOT place the sample below the instruments before opening the program, the program will close all transducers upon startup, and can cause severe injury.
Image Removed
Figure 3. Opening Program Desktop icon.

...

- Wasn't this experimentally determined?- Currently we are using 2100m/s.

T he travel times for any measurement signal is based on either a user selected location on a graphical display of the first arrival wave (manual pick), or a software auto-pick feature, which attempts to determine the first arrival of the measurement signal.  The auto-pick feature will search for the first instance of a signal stronger than a user set threshold value (milli-amps).  Next it will take the absolute value of the signal and determine where the wave-form crosses the graph's x-axis the 2nd time.  It will then subtract an assumed 1/2 of wave length, and display the pick location on the same graphical display.  This should get the travel-time of the measurement signal.  Verification of the signal and pick location by the user is critical.


II. Procedures


A. Preparing the Instrument


  1. Double-click the MUT icon on the desktop (Figure 1a) and login using ship credentials. For more information on data uploading see the "Uploading Data to LIMS" section below.
  2. Double-click the IMS icon (Figure 1b). IMS initializes the instrument. Once initialized, the logger is ready to measure the first section.


Image AddedFigure 1. (a) MUT Icon. (b) IMS Icon.

At launch, IMS begins an initialization process:

  • Testing instrument communications
  • Reloading configuration values
  • Homing the caliper and bayonets. IMP: DO NOT place samples on the track before launching the software. 

After successful initialization, the IMS Control and Instruments' windows appear (Figure 2).


Image Added

Figure 2. Main Velocity window.


***CAUTION: Make sure there are no samples or body parts underneath any of the sensors when starting IMS***


When first opening the program, all 3 sensors will go through a communication initiation process where they will move down and up. The program will display a warning message (Figure 3a) but will not stop the movement if nothing is detected on the rail.  If an object is detected a warning window will pop up (Figure 3b).


Image Added

Figure 3. (a) Reminder message while opening IMS. (b) Warning message if an object is detected.



The software home screen allows the user to modify a number of setup and acquisition parameters directly, as well as control the motion of the transducers.

The main window (Figure 4) includes:

  1. Sample type and Measurement Axis selection: The user selects the type of sample, while simultaneously selecting the measurement axis via the pictured tiles, the current selection will be highlighted.
  2. Graphical Display Tabs:  The RAW, RAW STACKED, RAW ZOOM  tabs will display the measurement signal without mathematical manipulation.  They also display the MANUAL PICK location as a vertical pink line.  The ABS and ABS ZOOM tabs display the absolute value and the uncorrected auto-pick location, which should be
    between the 1st and 2nd hump.  These displays are crucial for evaluating the signal quality and pick locations.
  3. Requested Stack and Threshold sliders:
    1. Requested Stack: The number of measurement signals the software will add together with the intent of increasing the signal to noise ratio. Each time it is adjusted a new set of signals will begin stacking.  100 stacks is a typical value.
    2. Threshold:  Voltage level (y-axis of the graphical display mV) that the stacked signal must attain for consideration in a auto-pick.  If the threshold is too low, signal noise may be selected as the first arrival.  If the threshold is too high, the first arrival may be missed and included with the signal noise.
  4. Mini Graphs: VELOCITY graph displays the running average of the last 1,000 samples and the TORQUE graph display the torque applied by the actuators to the sample.  The DISTANCE graph displays the AR700 distance readings.
  5. Motion Control Buttons: These control the up-down motion of the caliper and bayonets.  The user may select to make slow, fast or automatic movements. EMERGENCY OPEN button will bring up bayonets and caliper simultaneously.
  6. Pause Button: By selecting the pause button, the user will be able to make manual picks and have the velocity of that pick displayed.
  7. Save Data: The save button is activated when an automatic velocity pick is within the velocity filter range or when a manual pick is made.
  8. IMS Control panel (Figure 5): Provides access to utilities/editors via drop-down menus.


Image Added

Figure 4: Main Velocity Window with annotated sections.


Image Added

Figure 5. IMS Control Panel.



Before start measuring assure that the transducers are clean. If not, clean them with water and paper towels.

B. Instrument Calibration


The bayonets and caliper transducers must be calibrated and the system delay determined whenever a check standard is out of range, +/-2.0% of the expected velocity. There is a separate utility for each one:

  • The bayonet transducers are calibrated by back-calculating the system delay from the total travel time, for the y- and z-axes, from the transducer separation value and theoretical velocity of distilled water at the temperature of the water bath.
  • Calibration of the caliper requires the transducer and system delay to be calibrated concurrently. The transducer separation is measured with a displacement laser.

System delay is derived from the separation for the different calibration standards and travel time is derived from a travel time pick algorithm.
The calibration uses the derived velocity of the standard material to test for acceptance or rejection of the data.  Aluminum and acrylic, materials used as calibration standards for the caliper, have a published sound speed of 6295 m/s and 2730 m/s, respectively.

Caliper Calibration

The Caliper Calibration Utility calibrates the laser offset and the delay of the system at the same time.

Measurements are taken at standards with the same velocity but with different heights. The operator can choose the number of standards used and the material. A linear correction is applied to calibrate the laser offset, while the system delay and velocity is calculated from the time measurements. 

By convention a set of six acrylic standards with increasing sizes (15.07, 20.00, 30.03, 34.99, 34.90 and 45.01 mm) are used. They will be measured one by one, following next steps and the same procedure as an ordinary sample.

To open the Caliper Calibration Utility select: Stations > Caliper Calib. from the IMS panel menu (Figre 5). The Caliper Calibration window will open (Figure 6).

Image Added

Figure 6. Caliper Calibration window.


 In the Caliper Configuration window there are 7 row of calibration data which will auto populate from the previous calibration.  You can standards by scanning in new id's and/or you can update the given thickness.  Any changes will available in subsequent calibrations. If the row already has the correct ID and thickness you do need to re-scan the id.


In Figure 6 (item 1) there are 7 round buttons on the left side of each row.  Only one can be selected at a time and the one selected will be updated constantly by values being measured.  When you are satisfied you have the correct time pick, select the next standard to measure.  The previous values are frozen while you measure the remaining samples.  If you reselect that standard the values will be overwritten with the current measurement, so choose wisely.

The button on the right-hand side (Figure 6, item 2) are selected if you want to use that value in the calibration.  Remember you need at least two for calibration but use all 6.  Keep in mind that the program is always re-calculating the calibration values (Figure 9 and Figure 10) and it is ok to select and unselect value to see the affect on the calibration values.  The same with remeasuring just make sure you apply the correct time pick to the correct sample.


Quick step by step

  1. Select the standard
  2. Optional only if needed
    1. Click Scan Standard 
    2. Use bar code scanner and scan label
    3. Click OK
  3. Place the first standard in the caliper.
  4. Click AUTO CLOSE
  5. Set the threshold to pick the first arrival
  6. Set you stack to at least 100 and wait for the time pick to stabilize.  Increase the stack as necessary.
  7. When satisfied select the next standard
  8. Click AUTO OPEN
  9. Repeat steps 2 through 8 for each standard.
  10. Select the standards to use in the calibration (can be done at anytime)
  11. Click Accept Changes or Cancel Changes


To clear the rows, de-select the rows you want to clear (right-hand Boolean controls) and then click Clear All


Image Added

Figure 7. Scan Standard window.



How to verify first arrival pick:

Verify the software is picking the first arrival wave in the Raw Stacked graph tab. You can adjust the threshold scale bar to achieve the proper pick location (Figure 8). Observe the graph to verify the system is getting a clean signal.  If you are not getting a clean signal, follow the steps in the trouble shooting section of this guide. Also, verify the auto-pick location in the Absolute Value tab  (ABS or ABS Zoom).  This pick should be at the zero crossing after the first significant wave form above noise (Figure 8)

                                                                                                                                                                                                              

Image Added

Figure 8. Caliper Calibration Window:.  Raw data graph tab (left) and the ABS zoom graph tab (right).


Before you Accept


Check the standard's calculated velocity (= inverse of the slope) it should match the known material velocity as follows:

  •  Acrylic velocity= 2730 m/s ± 2.0 %
  • Aluminum velocity= 6295 m/s ± 2.0 %). 


To calculate the laser offset, the zero contact position offset was determined for the transducer and a small slope correction values (Figure 9).

Image Added

Figure 9. Laser (AR700) offset calculation.


To calculate material velocity and the system delay value the pick time is measured during the calibration (Figure 10).

Image Added

Figure 10. System Delay regression line.


Bayonet Calibration 


Because the distance between the bayonets is fixed, only the system time delay needs to be calculated. This is achieved by measuring the velocity in water of known temperature, the difference between the known and measured value is due to system delay.

To open the Bayonet calibration window (Figure 12), select Stations > Y-Bayonet Calib. or Z-Bayonet Calib. from the IMS panel menu (Figure 5).


Image Added Image AddedImage Added


                                                                                               Figure 12. Y-bayonet and Z-bayonet Calibration windows.  Y-bayonet raw graph tab (left), Z-bayonet raw graph tab (center), Y-bayonet ABS Zoom graph tab (right).


To calibrate the bayonets:

  1. Fill the bayonet calibration liner with DI water (Figure 13).  The water level must be high enough that the black transducer pads on the bayonets can be placed below the water line without touching the liner itself.  The water should be at room temperature if possible.


Image Added

Figure 12. Bayonet Water Standard


2. Place the liner below the bayonets.

3. Open the bayonet calibration window for the selected axis

4. Verify the bayonet separation value.  If the value is incorrect, go to the Velocity Setup window (Figure 13) to edit the value.

5. Use a thermometer to measure the temperature of the water in the calibration liner.

6. Enter the temperature value in the Temperature field next to the Water Velocity Calculator (lower left of the window).  The screen displays a plot of theoretical velocity of water vs. temperature. 

7. Use the Insert slow or Insert fast buttons to lower the bayonets into the water until the black transducer pads are submersed.  Use care when lowering the Y-bayonet to ensure you do not contact the core liner.

8. Verify the software is picking the first arrival wave in the Raw Stacked graph tab. You can adjust the threshold scale bar to achieve the proper pick location (Figure 12) as long as the signal is not too noisy.  Also, verify the auto-pick location in the Absolute Value tab (ABS or ABS Zoom) (Figure 12).  This pick should be at the zero crossing after the first wave form above noise (Figure 12).  An algorithm calculates the temperature-corrected velocity of the water bath and displays the result in theCorrected Velocity field. 

9. Select Determine System Delay. The computer calculates the system delay based on the separation distance and theoretical velocity of water. 

10. Compare the corrected velocity to the calculated H20 velocity value.  If the values are not within range, redo the calibration.

11. Select Accept Changes to save the calibration or select Cancel Changes to leave without saving the calibration.

12. As a check on your calibration, measure the velocity of the water to verify it is within an expected error margin, +/- 2.0%.  Ensure the proper sample type is selected when verifying the calibration.



C. Set Measurement Parameters


Configuration values should be set during initial setup and configuration by the physical properties technician.  There should be no need to change these values unless the configuration files is corrupted.  This window allows the user to view and modify the physical configuration values for the Velocity system, as well as the liner correction values and the velocity filter settings.


To open the Velocity instrument setup window (Figure 13), select Stations > Velocity Setup from the IMS panel menu (Figure 5).

  1. Ensure the values in the window are set as shown in Figure 13.
    1. Caliper and Bayonet Offsets: Physical offset from the laser zero point to the center of the transducer pair for the three stations (caliper, bayonet Y, bayonet Z). This measurement will not change unless the station location is physically changed.
    2. Axis Separation: Physical distance between the bayonet transducer pairs. This measurement should never change unless the bayonets holders are physically changed.  The X-Axis is not shown because its separation is determined during calibration and measurement by the AR700 laser.
    3. Liner Correction: The liner delay value as determined experimentally based on the liner thickness and velocity measurements on the liner material.
    4. Velocity Disable "Save" Filter: When the filter is enabled, the save feature is inactive for any automatic velocity picks outside of the velocity range set.  
  2. Click Ok to accept to save the changes and write them to the configuration file.  Click Cancel to revert to previous values.  NOTE: Only one configuration file exists in the IMS folder.  Every save will overwrite the config file.


Image Added


Figure 13. Velocity Parameters Window


D. Preparing Sections & Samples


A temperature-equilibrated split core section-half  in it's liner is placed on the core track.  The user positions the section half under the sensor, places a small piece of glad wrap on top of the core, and triggers the measurement from the software control panel. The measurement is taken continuously, with the recorded result representing the average of several thousand determinations. 

On semi-lithified and lithified samples, loose material may be present on the core cut surface.  Before placing the working half in the core tray, make sure the surface is clean (lightly brush away any material with a Kimwipe or paper towel).

A barcode reader records all relevant sample information, which is used for the data upload into LIMS. A laser sensor measures the distance to the top of the section-half and determines the sampling interval from the known sensor offsets. 

Discrete measurements are discussed in the Caliper Measurement section below.  When inserting the bayonets or closing the calipers into/on a sample, the graph should be monitored for signal quality.



Caliper Measurement

  • The sample is placed between 2 flat, 1 inch diameter sensors that squeeze firmly onto the specimen to ensure good contact.  De-Ionized (DI) Water is introduced between the sample and sensors to aid in signal propagation. For section-half measurements, place a small piece of Glad Wrap over the desired Caliper measurement location.  This will keep the transducers clean.
  • One sensor acts as a transducer and the other as a receiver to record velocity measurements at a rate of 0.5 MHz.
  • To measure discrete samples in multiple axis, the sample is rotated by the user to each axis (x, y, and z-axis) and measured separately.


Bayonet Measurement

  • Two pairs of piezoelectric transducers, set at 90° to each other, are inserted into the unconsolidated or semi soft section-half sediment. The sensors (black circles) must be at least partially buried in the material.
  • One of each pair of sensors acts as transducer and the other as receiver to measure velocity in two directions simultaneously.


Data Quality

Velocity data quality is affected by several variables.  The signal quality and pick locations should be consistently monitored by the user during every measurement.

  • Quality of the acoustic coupling between the core material and the sensor transducers. Note: Use water to increase the quality of the contact.
  • Quality of the coupling between both the transducer and the core liner and between the core liner and sample. Note: Use water to increase the quality of the contact.
  • Consolidation of the sediments; non-cohesive sediments containing microcracks or gas voids cannot be measured accurately.


E. Making a Measurement

When scanning label barcodes on the gantry system, there are specific label types that work best.

For measurements with the bayonets and the caliper on a section half, user must use a section half label.  

For measurements on discrete samples we recommend label types: Mad Residue Small, PMAG Cube label (sample type must be CUBE, CYL, or OTHR), Mad Residue Large, or Sample Table labels (the large format).  Other label formats may not parse properly.  Note that the MAD label parsing expects the container number in the name field.  If the name field is populated with other text, the label parsing may fail.

For discrete samples shared between MAD and PMAG, use the PMAG Cube label or a sample table label.  Mad residue labels WILL NOT parse for shared samples.

Currently the instrument is being upgraded, some of the measurement options or sample information input are not available.

a) Caliper Measurements


The caliper can be used to measure section half and discrete cube samples. Before measuring samples, be sure the samples are properly prepared and the system is calibrated.  Retaining the correct "up" direction is critical for axis determination, and for other systems like P-mag orientation; make sure the sample has the "up" direction marked on it.

Section halves measured on the Caliper station are working half sections placed on the track with the blue end cap (top of section) toward the AR1000 laser.  The laser measures the distance to the top of the section and the software calculates the offset to the measurement based on the known section length.  

Discrete samples measured on the Caliper station are hard material that has been cut from the core as cubes, minicores, or slabs. Because the material can be measured in various orientations, the user must select the measurement axis.
Traveltime is calculated as total traveltime minus x-system delay time. Discrete sample measurements are not corrected for the core liner.  The offset recorded in LIMS is the top offset of the discrete sample.  The AR1000 laser is not used for discrete measurements. The transducer separation is measured with the displacement laser, as it is for the sample half measurement.  For discrete samples the axis of measurement is selected for each measurement.


Section Half Measurement Procedure

  1. Place section half below the caliper.  Take care to not drag the core against the bayonets while placing the core on the track.
  2. Place a drop of distilled water below the section on the transducer and on top of a piece of Glad Wrap on top of the section to improve contact between the caliper and section.  Note: Placing Glad Wrap is not required.  This is only to improve signal quality.
  3. Select the proper instrument and measurement axis button (Caliper SHLF X-axis)
  4. Close the transducer onto the core until contact is made.  Do not over close the caliper on the section half.  
  5. Verify the automatic velocity pick in the Raw Stacked graph tab (Figure 4). You can adjust the threshold scale bar to achieve the proper pick location as long as the signal is not too noisy.  The user may want to verify the auto-pick location in the Absolute Value tab (ABS or ABS Zoom). This pick should be at the zero crossing after the first wave form.
  6. Select Save Data.  A sample information window will appear (Figure 14).
  7. Scan the section half barcode label to populate the fields of the sample information box.
  8. Verify the measurement offset.  
    1. If the offset is incorrect, select CANCEL, or enter the offset manually.  The laser range finder sometimes struggles to return a proper offset if the end cap is not opaque or if the core is not flat on track. Try adding a post it note or opaque end cap to the section half if the laser returns the wrong offset.
  9. Select SAVE DATA.


This procedure can also be used for whole rounds or pieces of section halves, but this is rarely used.  The software assumes that the whole round piece or section half piece is not within a core liner and will not correct for a liner velocity.  Scan the section half label when entering the sample ID information. 


Image Added

Figure 14. Section Half Caliper Measurements Sample Information Window


Discrete Sample Measurement Procedure

For each axis to be measured:

  1. Place a small drop of water on the lower caliper transducer
  2. Place the discrete sample on the caliper and add a drop of water to the top of the sample
  3. Select the proper instrument and measurement axis button (Caliper X-Axis, Caliper Y-Axis, Caliper Z-Axis).
  4. Lower the upper caliper transducer onto the specimen.  Do not apply unnecessary force on the specimen as it may fracture, or cause physical variation in the caliper system and measurement.
  5. Verify the automatic velocity pick in the Raw Stacked graph tab (Figure 4). You can adjust the threshold scale bar to achieve the proper pick location, or pause the system to do a manual pick.  The user may want to verify the auto-pick location in the Absolute Value tab (ABS or ABS Zoom). This pick should be at the zero crossing after the first wave form.
  6. Select Save Data.  The sample information window will open (Figure 15).  Note that the measurement offset will not be populated.  The laser range finder is not used for discrete measurements.
  7. Scan the discrete sample barcode label to populate the fields of the sample information box.  The sample's top offset in the section-half will be used as the measurement location, enter it manually.
  8. Select Save Data.


If the material to be measured is a whole piece removed from the liner (rarely used), select Caliper Piece X-Axis in Step 3. The measurement offset field in the sample information window is selectable.  Scan a section half label and enter the offset of your measurement from top of the core.


Image Added

Figure 15.  Discrete Caliper Measurements Sample Information Window

b) Bayonet Measurements


Before measuring samples, be sure the samples are properly prepared and the system is calibrated.

  1. Place section half below the bayonets.  Take care to not drag the core against the bayonets while placing the core on the track.
  2. Select the proper instrument and measurement axis button (Y or Z bayonet)
  3. Lower the selected bayonets into the section half until the black transducers are below the sediment surface.  A piece of glad wrap can be placed on the cut surface before inserting the bayonets, with water then added to aid in signal propagation.  The glad wrap is to prevent water from being absorbed into the material.
  4. Verify the automatic velocity pick in the Raw Stacked graph tab (Figure 12). You can adjust the threshold scale bar to achieve the proper pick location, or pause the system to do a manual pick.  The user may want to verify the auto-pick location in the Absolute Value tab (ABS or ABS Zoom). This pick should be at the zero crossing after the first wave form above noise.
  5. Select Save Data.  A sample information window will appear (Figure 14).
  6. Scan the section half barcode label to populate the fields of the sample information box.
  7. Verify the measurement offset.  
    1. If the offset is incorrect, select Cancel and select Save Data again, or insert it manually.  The laser range finder sometimes struggles to return a proper offset if the end cap is not opaque or if the core is not flat on track.  Try adding a post it note or opaque end cap to the section half if the laser returns the wrong offset.
  8. Once the measurement offset is verified.
  9. Select Save Data.


c) Manual Pick

In some cases, a user may wish to make a manual pick.  This usually occurs because the automatic pick is unsuccessful, often due to a noisy signal. 

Often, the Save Data option will not be available for a bad automatic velocity pick, because the values are outside of the velocity filter range. Figure 16 shows an example of a situation in which adjusting the threshold value could not overcome a large peak at the start of the measurement. Using the manual pick, the user is able to override the computers pick.  Both the automatic velocity pick and manual pick are recorded in LIMS.  This option is available for all measurements on the Velocity-Gantry track. 

IMPORTANT: It is important that all users, on both shifts, agree ahead of time on the proper manual pick location.  If they do not, each user is likely to induce a small offset in velocity due to inconsistent pick locations; the zoom tools in the lower left of the graph may help better locate the wave forms. First arrival location (Figure 17) is not as straight forward as one might assume. 

The manual pick location should be relatively obvious so anyone can see it above the noise, because noise is likely to be high when auto-pick doesn't work.  Then any necessary offset to correct can be applied in post processing.  The agreed upon manual pick location and any offset correction in first arrival should be documented in the scientists' methods section. 

To make a manual pick:

  1. Place sample in the selected instrument
  2. Once a signal is visible on the graphical display, select the Pause button in the lower left corner of the main window (Figure 4).
  3. The manual pick tab will be displayed on the screen (Figure 16).
  4. Slide the red line along the x-axis to the desired pick location.  The line should be placed at the first arrival. To help select the first arrival, the plot can be zoomed in using the graph palette zoom function in the lower left corner of the graphical display.
  5. Verify the Velocity-manual value is reasonable for the material measured.
  6. Select Save Data.  The sample information window will open.
  7. Scan the sample barcode label and save the data to the LIMS database.

To exit the manual pick and return to the automatic velocity pick, press the Play button.

Image Added

Figure 16. Manual Pick Graphical Display




Image Added

Figure 17. Zoomed Manual Pick Graphical Display

F. Evaluating your Measurement


Running a Standard as a QAQC Check

You can run a standard to verify if the instrument is measuring correctly. There are three materials available as a standard: Aluminum and acrylic are used on the caliper, and a core liner with D.I. water on the bayonets.

Expected velocities for each standard: 
Aluminum 6295 m/s (+/- 63 m/s)
Acrylic 2730 m/s (+/- 27 m/s)
Water 1480 m/s (+/- 7 m/s)

Typical allowable deviation is 1% for the caliper and 0.5% for the bayonets. There will also be differences based on temperature, especially for water and aluminum. If the standard values are out of this range, ask a technician to help calibrate the instrument.


G. IMS Utilities

  

a) Motion Utilities

The available motion utilities are found under the Motion menu from the IMS panel menu (Figure 5). 

The user may home each instrument separately using the Home Caliper, Home Y-Bayonet, Home Z-Bayonet options or home both bayonets and the caliper at the same time using the Home All command.


b) AR1000 Laser Utility

To open the AR1000 utility window (Figure 18), select Motion> AR1000 Utility from the IMS panel menu (Figure 5).  This utility is useful when a user is trying to determine if the laser is correctly reading the distance to the center of each instrument.

If the mean distance shown does not agree with the instrument offset in the Velocity Setup Window (Figure 13), the offsets reported in the database will be incorrect.

The utility immediately begins reading the AR1000 laser output and averaging the values collected once the utility is opened.

To check the instrument offsets, place an object in the center of the instrument in question, i.e. the caliper.  The object surface should be at the center of the instrument.  An end cap works well as long as it is opaque. A tape measure can also be used to independently verify the offset value.


Image Added

Figure 18. AR1000 Laser Utilty.


c) AR700 Displacement Laser Utility


To open the AR700 utility window (Figure 19), select Stations> AR700 Utility from the IMS panel menu (Figure 5). 

Use this utility to determine if the AR700 displacement laser is returning accurate distances.  The readings from this laser are used as the distance values in the velocity calculations.  Inaccurate readings from the laser will cause error in the velocity readings.  

The AR700 is mounted on the Caliper actuator.  The more the Caliper is closed, the smaller the laser distance returned.  It is recommended that the objects used for testing are similar in size to the samples being measured.

  Image Added

Figure 19. AR700 Displacement Laser Utility.


d) EXLAR Utility

This utility allows to control the movement and change the parameters of the three Exlar actuators: Bayonet Z, bayonet Y and Caliper.

To open it, on the main window, go to: Motion > EXLAR Utility. TRITEX Configuration Utility window will open (Figure 20).


Image Added


Figure 20.- TRITEX Configuration Utility window. CONTROL tab.



The EXLAR actuator setup is saved in three different places: The actuator, the software and the Utility. The Utility is the interface between the Actuator and the software.


ACTUATOR DOWNLOAD: Copies the information from Edit Me (Utility) to the actuator.

COPY DEFAULT to EDIT ME: Copies the information from the software to the utility.

COPY ACTUATOR to EDIT ME: Copies the information from the actuator to the utility.

SYNC ALL: Copies the information from the utility to both, sofware and actuator.

CANCEL RESTORE ORIGINAL: Returns to the software setup.

DONE: Exits utility.


CALIPER: Select it if you want to work with caliper setup.

Y-BAYONET: Select it if you want to work with Y-bayonet setup.

Z-BAYONET: Select it if you want to work with z-bayonet setup.


CONTROL: Allows to move the actuator and set upper and lower limit. Could be used for testing and troubleshooting purposes.

EDIT ME: Allows to edit the utility setup information.

ACTUATOR-VIEW ONLY: Shows the information saved on the utility. To edit it the user should go to EDIT ME tab.



CONTROL tab


MOVE 1: Used for auto clamping/ insertion move.

MOVE 2: Used for removal/ unclamp.

MOVE 3: Not used by the program.

MOVE 4: Used for Emergency opening (quick unclamp/removal).

HOME: Moves the actuator to the home position.

JOG UP: Moves the actuator up (opening).

JOG DOWN: Moves the actuator down (closing).

DEF HOME: Sets the information for the home position.

RESET FAULTS: ???

DISABLE NEG LIMIT: Disable negative limit written in the box.

DISABLE POS LIMIT: Disable positive limit, written in the box.

SET NEG LIMIT: Positions the Upper actuator limit.

SET POS LIMIT: Positions the Lower actuator limit.

Negative Limit box: Selected upper limit.

Positive Limit box: Selected lower limit.

JPG SPD box: If checked, jog movements are FAST. If unchecked movements are SLOW.

MOVE TYPE box: If checked, move buttons motion is MAINTAINED. If unchecked is MOMENTARY.

STOP: Stops de movement. ???

PAUSE: Pause the movement. ???

Position: Position of the caliper or bayonet, counting from the home position.

Velocity: Speed of the movement, measured in REV/S

Current: Measured in Amp.

Caliper Compression switch State: Indicates if the caliper is under pressure.

Position, Velocity and Current graphs: Show the evolution and characteristics of the actuator movement.



ACTUATOR STATUS tab (Figure 21)


Image Added


Figure 21: Actuator Status. EXLAR Utility.

Input_Functions: List of commands directed to the actuator

Output_Functions: Information related with the position and behavior of the actuator.

Disabled by: ???

Feedback: Setup information ???

Drive Info: Vendor's information about the actuator

ALL FAULTS: List of possible faults

HARD FAULTS: Hardward derived faults list.

SOFT FAULTS: Software derived faults list.

Serial Errors: ????

Status Info: Info of temperature, voltage and power.

Command Method: Command user to move the actuator.

HOST Disables: ???

Inputs: Controls set to monitor the actuator behavior. Ex: Input 1 switched ON, means that the actuator is at home position. Input 5 ON means that the caliper pressure limit switch is triggered.

Outputs: ???



ACTUATOR - VIEW ONLY tab (Figure 22)


Default parameters Setup for the caliper actuator are shown in Figure 22.


Image Added


Figure 22. Caliper, actuator - view only Setup. EXLAR Utility.


The ACTUATOR - VIEW ONLY diagram shows the default setup for each one of the actuators. 

All the characteristics of the different types of movements that the actuators execute are defined here.


There are different setups for each one of the actuator. See Figure 23 and 24 for Bayonets setup.


Image Added

Figure 23. Y-Bayonet actuator - view only Setup. EXLAR Utility.



Image Added


Figure 24. Z-Bayonet actuator - view only Setup. EXLAR Utility.


EDIT ME tab

It presents the same configuration as the ACTUATOR - VIEW ONLY tab. On this diagram the user can define or change the behavior of each one of the movements defined for the actuator.

In order to save the changes the information must be transferred to the actuator, the software or to both of them. For this purpose, use the buttons on the left side of the window.



III. Uploading Data to LIMS


A. Data Upload Procedure


To upload the images into the database (LIMS) MegaUploadaTron (MUT) program must be running in background.

If not already started do the following:

  1. On the desktop click the MUT icon on the bottom task bar (Figure 25) and login with ship credentials. The LIMS Uploader window will appear (Figure 26).
  2. Once activated, the list of files from the C:\DATA\IN directory is displayed. Files are marked ready for upload by a green check mark.


Image Added

Figure 25. MUT icon


Image Added

Figure 26. LIMS Uploader window


3. To manually upload files, check each file individually and clip upload. To automatically upload files, click on the Automatic Upload checkbox. The window can be minimized and MUT will be running in the background.

4. If files are marked by a purple question mark or red and white X icons, please contact a technician.
    Purple question mark: Cannot identify the file.

    Red and white X icons: Contains file errors.

5. Upon upload, data is moved to C:\DATA\Archive. If upload is unsuccessful, data is automatically moved to C:\DATA\Error. Please contact the PP technician is this occurs.


Data File Formats

The files generated by the Velocity Gantry system are:

  • Velocity.ini:  A file containing the configuration information for each instrument on the Velocity Gantry track and the most recent calibration information for each instrument
  • .pwc: A file containing an individual measurement on the Caliper or Bayonets.  Data includes distance between calipers (mm), travel time (us), the measurement offset on the section (cm) (no offset included for discretes), the calculated velocity (m/s), the time stamp of the measurement, the system delay, and the liner time correction.  The name and location of all associated files are also listed within this file.
  • .lvm: A file containing the transducer signal data for the given measurement.  The signal and time (us) are provided.


B. View and Verify Data


Data Available in LIVE

The data measured on each instrument can be viewed in real time on the LIMS information viewer (LIVE).

Choose the appropriate template (Ex: PHYS_PROPS_Summary), Expedition, Site, Hole or the needed restrictions and click View Data. The requested data will be displayed. You can travel in them by clicking on each of each core or section, which will enlarge the image.


Data Available in LORE


The data for the P-wave velocity Caliper System (PWC) and P-wave velocity Bayonets System (PWB) are each displayed under a different report in LORE.  These reports are found under the Physical Properties heading. The expanded reports include the linked original data files and more detailed information regarding the measurement.


P-Wave Velocity, Caliper System (PWC) Standard Report

  • Exp: Expedition number
  • Site: Site number
  • Hole: Hole number
  • Core: Core number
  • Type: Type indicates the coring tool used to recover the core (typical types are F, H, R, X).
  • Sect: Section number
  • A/W: Archive (A) or working (W) section half.
  • Offset on section (cm): Position of the sample being measured relative to the top of a section or section half.
  • Offset (cm): Position of observation made on a sample (typically a section half), measured relative to the top of the sample.
  • Depth CSF-A (m): Location of the observation expressed relative to the top of a hole.
  • Depth [other] (m): Location of the observation expressed relative to the top of a hole. The location is presented in a scale selected by the science party or the report user.
  • P-wave velocity x|y|z|unknown (m/s): Velocity of pressure wave through sample along the noted axis.
  • Caliper separation (mm): Displacement between the faces of the transducers in contact with the sample, associated with the velocity observation in the specified axis.
  • Sonic traveltime (µs): Traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
  • P-wave velocity x|y|z|unknown [manual] (m/s): Velocity of pressure wave through sample in the noted axis, calculated using the manually picked traveltime.
  • Sonic traveltime [manual] (µs): Manually picked traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
  • Timestamp (UTC): Point in time at which the data was uploaded on the Gantry.
  • Instrument: An abbreviation or mnemonic for the P-wave sensing device used to make these observations (CaliperPWS).
  • Instrument group: Abbreviation or mnemonic for the data collection device used to acquire this observation (Gantry).
  • Text ID: Automatically generated unique database identifier for a sample, visible on printed labels.
  • Test No: Unique number associated with the instrument measurement steps that produced these data.
  • Comments: Observer's notes about a measurement, the sample, or the measurement process.


P-Wave Velocity, Bayonet System (PWB)  Standard Report

  • Exp: Expedition number
  • Site: Site number
  • Hole: Hole number
  • Core: Core number
  • Type: Type indicates the coring tool used to recover the core (typical types are F, H, R, X).
  • Sect: Section number
  • A/W: Archive (A) or working (W) section half.
  • Offset on section (cm): Position of the sample being measured relative to the top of a section or section half.
  • Offset (cm): Position of observation made on a sample (typically a section half), measured relative to the top of the sample.
  • Depth CSF-A (m): Location of the observation expressed relative to the top of a hole.
  • Depth [other] (m): Location of the observation expressed relative to the top of a hole. The location is presented in a scale selected by the science party or the report user.
  • P-wave velocity x|y|z|unknown (m/s): Velocity of pressure wave through sample along the noted axis.
  • Caliper separation (mm): Displacement between the faces of the transducers in contact with the sample, associated with the velocity observation in the specified axis.
  • Sonic traveltime (µs): Traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
  • P-wave velocity x|y|z|unknown [manual] (m/s): Velocity of pressure wave through sample in the noted axis, calculated using the manually picked traveltime.
  • Sonic traveltime [manual] (µs): Manually picked traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
  • Timestamp (UTC): Point in time at which the data was uploaded on the Gantry.
  • Instrument: An abbreviation or mnemonic for the P-wave sensing device used to make these observations (CaliperPWS).
  • Instrument group: Abbreviation or mnemonic for the data collection device used to acquire this observation (Gantry).
  • Text ID: Automatically generated unique database identifier for a sample, visible on printed labels.
  • Test No: Unique number associated with the instrument measurement steps that produced these data.
  • Comments: Observer's notes about a measurement, the sample, or the measurement process

C. Retrieve Data from LIMS


Expedition data can be downloaded from the database using the instrument Expanded Report on Download LIMS core data (LORE).



IV. Important Notes


Standards

Various standards are available for the P-Wave Gantry system (Figure 27).  These include:

  • Laboratory reagent water (distilled) (for the bayonet calibration), velocity = 1480 m/s at 20 degrees C
  • Aluminum block of known length and velocity; velocity = 6295 m/s (for caliper calibration)
  • Acrylic half-cylinders of variable thicknesses; the assumed Acrylic Velocity as of June, 2019 is 2730 m/s. 
  • Liner Velocity accepted value as of June 2019 is 2100 m/s with 40 m/s error


It is recommended that the check standard used be similar in size to the sample being measured. 

  Image Added

 Image Added

          Figure 27. Velocity Standards.



V. Appendix


A.1 Health, Safety & Environment


Safety

  • Keep extraneous items and body parts away from the moving parts.
  • The track system has three emergency opening buttons, one on the top of each actuator (Figure 28).  Pressing the emergency opening button will open the actuator.
  • Do not look directly into the laser light source (class 2 laser product).
  • Do not direct the laser beam at other people.
  • Do not attempt to work on the system while a measurement is in progress.
  • Do not lean over or on the track.
  • Do not stack anything on the track.
  • This analytical system does not require personal protective equipment.


Image Added

Figure 28. Emergency opening buttons on the Velocity Track

Pollution Prevention

This procedure does not generate heat or gases and requires no containment equipment.


Waste Management

Dispose all the plastic wrap in an approved waste container.



A.2 Maintenance and Troubleshooting


Common Issues


The following are common problems encountered when using the P-wave Velocity Gantry and their possible causes and solutions. For information about the laser Measurement and Operation software, read Gantry NI Max setup.docx. Vp GIESA Gantry Laser Operation manual.(updating)


Issue

Possible Causes

Solution

Laser sensor: no laser light/no laser range data

Sampling is turned off

Turn sampling on through laser Measurement and Operation software

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

Laser sensor: Serial port not responding

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

Baud rate incorrect or unknown

See Acuity manual 3.5.1; laser Measurement and Operation software

Laser sensor: Error code (Exx) transmitted on serial port



See Acuity manual 5.3; laser Measurement and Operation software

Exlar actuator: No response

I/O error

Check drive or Exlar software for faults

Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

Power interruption

Check wiring

Exlar actuator: Behaving erratically

Drive improperly tuned

Check gain settings

Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

Too much load on motor

Check for load irregularity or excess load

Exlar actuator: Cannot move load

Too much load or friction

Check load and friction sources

Excessive side load

Check side load

Misalignment of output rod to load

Check alignment of rod and load

Current limit on drive too low

Check current on drive

Power supply current too low

Check power supply current

Exlar actuator: Housing vibrates when shaft in motion

Loose mounting

Tighten mounting screws

Drive improperly tuned

Check gain settings

Exlar actuator: Output rod rotates during motion

Rod rotation prevents proper linear motion

Install anti-rotation assembly

Exlar actuator: Overheating



Ambient temperature too high

Check ambient temperature

Actuator operation outside continuous ratings

Check operation settings

Amplifier poorly tuned

Check gain settings

Caliper Freezes in placePut too much force on object and passed the current limitOpen Exlar utility and press sync to coordinate the software settings and the actuator.
Auto Open/Auto Close Buttons cause caliper to freezeThe rev/sec may be incorrect for the caliperCheck Exlar utilities window.  Change 22 rev/sec on Motion 1 and 3 to 21 rev/sec



Resetting the AR700 Laser Baud Rate


A common issue with the Caliper system is the AR700 laser baud rate.  If power is lost to the Laser it may default to an incorrect baud rate when turned back on.  See the Gantry NI Max setup.docx  or find the AR700 Vendor Manual in the Lab Notebook: Physical Properties page under Resources: General section.



Maintenance


After Every Sample

Clean contact sensors with tissue, cloth, or paper towel and water.


Start of Expedition/As Needed

Remove the caliper transducer caps (clear acrylic) from the caliper by unscrewing the aluminum retaining ring.  With a Kimwipe, remove all old Couplant fluid and water.  Thinly coat the transducer's surface, or the cap for the upper transducer, with Couplant B, and re-screw the retaining ring.  Re-calibrate after each time, paying attention to the signal quality.

Clean dust from the laser sensor lens using compressed air or delicate tissue wipes. Do not use organic cleaning solvents on the sensor lens.

Check that lower transducer for caliper is making contact with core liner as sections are measured.

Polish the transducer cap when marks are noticed. Transparent ones are polish on Thin Sections lab. Metal ones?


Annually

Coat (not pack) the following parts of the actuators with grease:

  • Angular contact thrust bearings
  • Roller screw cylinder
  • Roller screw assembly in the actuators

Examine the cable management system for abraded cables or other indications of wear.


B.1 IMS Program Structure


a) IMS Program Structure


IMS is a modular program. Individual modules are as follows:

  • Config Files: Unique for each track. Used for initialize the track and set up default parameters.
  • Documents: Important Information related with configuration setup.
  • Error: IMS error file.
  • FRIENDS: Systems that use IMS but don't use DAQ Engine.
  • IMS Common: Programs that are used by different instruments.
  • PLUG-INS: Code for each of the instruments. 
  • Projects: Main IMS libraries.
  • Resources: Other information and programs needed to run IMS.
  • UI: User Interface
  • X-CONFIG: .ini files, specifics for each instrument.
  • MOTION plug-in: Codes for the motion control system.
  • DAQ Engine: Code that organizes INST and MOTION plug-ins into a track system.

The IMS Main User Interface (IMS-UI) calls these modules, instructs them to initialize, and provides a user interface to their functionality.


The Velocity system, specifically, is built with three INST modules (Caliper, Y-Bayonet, Z-bayonet), three MOTION modules (one for each instrument actuator) .

The IMS Main User Interface (IMS-UI) calls these modules, instructs them to initialize, and provides a user interface to their functionality.

Each module manages a configuration file that opens the IMS program at the same state it was when previously closed and provides utilities for the user to edit or modify the configuration data and calibration routines.

The three buttons on the left side of the IMS-UI window provide access to utilities/editors via dropdown menus as shown in Figure 5.



b) Communication and Control Setup


Data communication and control is USB based and managed via National Instrument’s Measurement & Automation Explorer (NI-MAX). When you open NI-MAX and expand the Device and Interface section.  The correct communications setup can be found at IMS Hardware Communications Setup.

B.2 Motion Control Setup


Refer to EXLAR Utility for further information.


B.3 LIMS Component Tables


The tables below contain the LIMS components for the following ANALYSIS codes:

  • PWAVE_B - p-wave velocity by X- and Z-axis bayonet
  • PWAVE_C - p-wave velocity by caliper (X, Y, or Z axis)
ANALYSISTABLENAMEABOUT TEXT
PWAVE_BSAMPLEExpExp: expedition number
PWAVE_BSAMPLESiteSite: site number
PWAVE_BSAMPLEHoleHole: hole number
PWAVE_BSAMPLECoreCore: core number
PWAVE_BSAMPLETypeType: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
PWAVE_BSAMPLESectSect: section number
PWAVE_BSAMPLEA/WA/W: archive (A) or working (W) section half.
PWAVE_BSAMPLEtext_idText_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples.
PWAVE_BSAMPLEsample_numberSample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table.
PWAVE_BSAMPLElabel_idLabel identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples.
PWAVE_BSAMPLEsample_nameSample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter.
PWAVE_BSAMPLEx_sample_stateSample state: Single-character identifier always set to "W" for samples; standards can vary.
PWAVE_BSAMPLEx_projectProject: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
PWAVE_BSAMPLEx_capt_locCaptured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
PWAVE_BSAMPLElocationLocation: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
PWAVE_BSAMPLEx_sampling_toolSampling tool: sampling tool used to take the sample (e.g., syringe, spatula)
PWAVE_BSAMPLEchanged_byChanged by: username of account used to make a change to a sample record
PWAVE_BSAMPLEchanged_onChanged on: date/time stamp for change made to a sample record
PWAVE_BSAMPLEsample_typeSample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ)
PWAVE_BSAMPLEx_offsetOffset (m): top offset of sample from top of parent sample, expressed in meters.
PWAVE_BSAMPLEx_offset_cmOffset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm)
PWAVE_BSAMPLEx_bottom_offset_cmBottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm)
PWAVE_BSAMPLEx_diameterDiameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master
PWAVE_BSAMPLEx_orig_lenOriginal length (m): field for the original length of a sample; not always (or reliably) populated
PWAVE_BSAMPLEx_lengthLength (m): field for the length of a sample [as entered upon creation]
PWAVE_BSAMPLEx_length_cmLength (cm): field for the length of a sample. This is a calculated field (length, converted to cm).
PWAVE_BSAMPLEstatusStatus: single-character code for the current status of a sample (e.g., active, canceled)
PWAVE_BSAMPLEold_statusOld status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample
PWAVE_BSAMPLEoriginal_sampleOriginal sample: field tying a sample below the CORE level to its parent HOLE sample
PWAVE_BSAMPLEparent_sampleParent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR)
PWAVE_BSAMPLEstandardStandard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T)
PWAVE_BSAMPLElogin_byLogin by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created])
PWAVE_BSAMPLElogin_dateLogin date: creation date of the sample
PWAVE_BSAMPLElegacyLegacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition
PWAVE_BTESTtest changed_onTEST changed on: date/time stamp for a change to a test record.
PWAVE_BTESTtest statusTEST status: single-character code for the current status of a test (e.g., active, in process, canceled)
PWAVE_BTESTtest old_statusTEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test
PWAVE_BTESTtest test_numberTEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table.
PWAVE_BTESTtest date_receivedTEST date received: date/time stamp for the creation of the test record.
PWAVE_BTESTtest instrumentTEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input)
PWAVE_BTESTtest analysisTEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table)
PWAVE_BTESTtest x_projectTEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
PWAVE_BTESTtest sample_numberTEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table
PWAVE_BCALCULATEDDepth CSF-A (m)Depth CSF-A (m): position of observation expressed relative to the top of the hole.
PWAVE_BCALCULATEDDepth CSF-B (m)Depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
PWAVE_BRESULTconfig_asman_idRESULT config file ASMAN_ID: serial number of the ASMAN link for the configuration file
PWAVE_BRESULTconfig_filenameRESULT config filename: file name of the configuration file
PWAVE_BRESULTdistance_in_caliper (mm)RESULT distance in caliper (mm): distance between transducers
PWAVE_BRESULTmanual_pick_travel_time (µs)RESULT manual pick travel time (us): manually picked traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
PWAVE_BRESULTmanual_pick_velocity_y (m/s)RESULT manual pick Y-axis velocity (m/s): velocity of pressure wave through sample for the Y axis, calculated using the manually picked traveltime.
PWAVE_BRESULTmanual_pick_velocity_z (m/s)RESULT manual pick Z-axis velocity (m/s): velocity of pressure wave through sample for the Z axis, calculated using the manually picked traveltime.
PWAVE_BRESULTobserved_length (cm)RESULT observed length (cm): length of the section as recorded by the core logger track
PWAVE_BRESULToffset (cm)RESULT offset (cm): position of the observation made, measured relative to the top of a section half.
PWAVE_BRESULTrun_asman_idRESULT run file ASMAN_ID: serial number of the ASMAN link for the run (.PWC) file
PWAVE_BRESULTrun_filenameRESULT run filename: file name of the run (.PWC) file
PWAVE_BRESULTtimestampRESULT timestamp: date/time stamp for the actual measurement time
PWAVE_BRESULTtransducer_signal_asman_idRESULT transducer signal file ASMAN_ID: serial number of the ASMAN link for the transducer signal (.LVM) file
PWAVE_BRESULTtransducer_signal_filenameRESULT run filename: file name of the transducer signal (.LVM) file
PWAVE_BRESULTtravel_time (µs)RESULT travel time (us): traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis. (Calculated by software)
PWAVE_BRESULTvelocity_y (m/s)RESULT sonic velocity Y-axis (m/s): velocity of pressure wave through sample for the Y axis, calculated by the software
PWAVE_BRESULTvelocity_z (m/s)RESULT sonic velocity Z-axis (m/s): velocity of pressure wave through sample for the Z axis, calculated by the software
PWAVE_BSAMPLEsample descriptionSAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments"
PWAVE_BTESTtest test_commentTEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments"
PWAVE_BRESULTresult commentsRESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments"


ANALYSISTABLENAMEABOUT TEXT
PWAVE_CSAMPLEExpExp: expedition number
PWAVE_CSAMPLESiteSite: site number
PWAVE_CSAMPLEHoleHole: hole number
PWAVE_CSAMPLECoreCore: core number
PWAVE_CSAMPLETypeType: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
PWAVE_CSAMPLESectSect: section number
PWAVE_CSAMPLEA/WA/W: archive (A) or working (W) section half.
PWAVE_CSAMPLEtext_idText_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples.
PWAVE_CSAMPLEsample_numberSample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table.
PWAVE_CSAMPLElabel_idLabel identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples.
PWAVE_CSAMPLEsample_nameSample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter.
PWAVE_CSAMPLEx_sample_stateSample state: Single-character identifier always set to "W" for samples; standards can vary.
PWAVE_CSAMPLEx_projectProject: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
PWAVE_CSAMPLEx_capt_locCaptured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
PWAVE_CSAMPLElocationLocation: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
PWAVE_CSAMPLEx_sampling_toolSampling tool: sampling tool used to take the sample (e.g., syringe, spatula)
PWAVE_CSAMPLEchanged_byChanged by: username of account used to make a change to a sample record
PWAVE_CSAMPLEchanged_onChanged on: date/time stamp for change made to a sample record
PWAVE_CSAMPLEsample_typeSample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ)
PWAVE_CSAMPLEx_offsetOffset (m): top offset of sample from top of parent sample, expressed in meters.
PWAVE_CSAMPLEx_offset_cmOffset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm)
PWAVE_CSAMPLEx_bottom_offset_cmBottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm)
PWAVE_CSAMPLEx_diameterDiameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master
PWAVE_CSAMPLEx_orig_lenOriginal length (m): field for the original length of a sample; not always (or reliably) populated
PWAVE_CSAMPLEx_lengthLength (m): field for the length of a sample [as entered upon creation]
PWAVE_CSAMPLEx_length_cmLength (cm): field for the length of a sample. This is a calculated field (length, converted to cm).
PWAVE_CSAMPLEstatusStatus: single-character code for the current status of a sample (e.g., active, canceled)
PWAVE_CSAMPLEold_statusOld status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample
PWAVE_CSAMPLEoriginal_sampleOriginal sample: field tying a sample below the CORE level to its parent HOLE sample
PWAVE_CSAMPLEparent_sampleParent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR)
PWAVE_CSAMPLEstandardStandard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T)
PWAVE_CSAMPLElogin_byLogin by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created])
PWAVE_CSAMPLElogin_dateLogin date: creation date of the sample
PWAVE_CSAMPLElegacyLegacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition
PWAVE_CTESTtest changed_onTEST changed on: date/time stamp for a change to a test record.
PWAVE_CTESTtest statusTEST status: single-character code for the current status of a test (e.g., active, in process, canceled)
PWAVE_CTESTtest old_statusTEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test
PWAVE_CTESTtest test_numberTEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table.
PWAVE_CTESTtest date_receivedTEST date received: date/time stamp for the creation of the test record.
PWAVE_CTESTtest instrumentTEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input)
PWAVE_CTESTtest analysisTEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table)
PWAVE_CTESTtest x_projectTEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
PWAVE_CTESTtest sample_numberTEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table
PWAVE_CCALCULATEDDepth CSF-A (m)Depth CSF-A (m): position of observation expressed relative to the top of the hole.
PWAVE_CCALCULATEDDepth CSF-B (m)Depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
PWAVE_CRESULTconfig_asman_idRESULT config file ASMAN_ID: serial number of the ASMAN link for the configuration file
PWAVE_CRESULTconfig_filenameRESULT config filename: file name of the configuration file
PWAVE_CRESULTdistance_in_caliper (mm)RESULT distance in caliper (mm): distance between transducers
PWAVE_CRESULTmanual_pick_travel_time (µs)RESULT manual pick travel time (us): manually picked traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis.
PWAVE_CRESULTmanual_pick_velocity_x (m/s)RESULT manual pick X-axis velocity (m/s): velocity of pressure wave through sample for the X axis, calculated using the manually picked traveltime.
PWAVE_CRESULTmanual_pick_velocity_y (m/s)RESULT manual pick Y-axis velocity (m/s): velocity of pressure wave through sample for the Y axis, calculated using the manually picked traveltime.
PWAVE_CRESULTmanual_pick_velocity_z (m/s)RESULT manual pick Z-axis velocity (m/s): velocity of pressure wave through sample for the Z axis, calculated using the manually picked traveltime.
PWAVE_CRESULTobserved_length (cm)RESULT observed length (cm): length of the section as recorded by the core logger track
PWAVE_CRESULToffset (cm)RESULT offset (cm): position of the observation made, measured relative to the top of a section half.
PWAVE_CRESULTrun_asman_idRESULT run file ASMAN_ID: serial number of the ASMAN link for the run (.PWC) file
PWAVE_CRESULTrun_filenameRESULT run filename: file name of the run (.PWC) file
PWAVE_CRESULTtimestampRESULT timestamp: date/time stamp for the actual measurement time
PWAVE_CRESULTtransducer_signal_filenameRESULT run filename: file name of the transducer signal (.LVM) file
PWAVE_CRESULTtravel_time (µs)RESULT travel time (us): traveltime of the sonic wave from transducer to transducer through the sample, associated with the velocity observation in the specified axis. (Calculated by software)
PWAVE_CRESULTvelocity_x (m/s)RESULT sonic velocity X-axis (m/s): velocity of pressure wave through sample for the X axis, calculated by the software
PWAVE_CRESULTvelocity_y (m/s)RESULT sonic velocity Y-axis (m/s): velocity of pressure wave through sample for the Y axis, calculated by the software
PWAVE_CRESULTvelocity_z (m/s)RESULT sonic velocity Z-axis (m/s): velocity of pressure wave through sample for the Z axis, calculated by the software
PWAVE_CSAMPLEsample descriptionSAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments"
PWAVE_CTESTtest test_commentTEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments"
PWAVE_CRESULTresult commentsRESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments"


C.1 Hardware


Hardware

The velocity track system consists of the following components (Figure 29):

  • Caliper transducers: X-axis velocity measurements
  • Chisel-type transducers (bayonet transducers): Y and Z axis velocity measurements
  • 3 Linear actuators: Each actuator controls the motion for one instrument 
  • Laser Range Finder (AR1000 laser): Used to determine offset of measurements on section halves
  • Displacement Laser (AR700): Used to determine distance between caliper transducers during measurements
  • Two Signal Amplifiers: One amplifier is common to all three instruments and is used to amplify the transducer signals.  The second amplifier is used as a pre-amplifier for the receiving signal from the Caliper only.  The Caliper returns a weaker signal than the Bayonets.  By using this pre-amplifier the Caliper signal can be amplified without affecting the Bayonet measurements.   If the signal is noisy or too weak, adjustments are made on only the gain dial.

Image AddedImage AddedImage Added

Figure 29. Bayonets, caliper and displacement laser on the Gantry track (left), signal amplifier (center), and laser range finder (right).


Caliper Transducers: Panametrics-NDT Microscan Delay Line Transducers

Specification

Value

Frequency (MHz)

0.5

Element diameter (mm)

25

Part number

M2008


Bayonet Transducers

Custom made.


Exlar Linear Actuator

Specification

Value

Maximum radial load (lb)

15

Resolution (revolution)

0.001

Accuracy (revolution)

±0.010

Operating temperature (°C)

0–55

Model Number

TLM20-0601-1-IFM-1BS-50-AR

Voltage (VDC)

48

Current (A)

5000 @ 10 rpm


Acuity 1000 Laser Distance Sensor

Specification

Value

Distance (m)

30

Laser

650 nm, 1 mW visible red

Accuracy

±0.12 inch (3.05 mm)

Resolution

0.004 inch (0.10 mm)

Operating temperature (°C)

–10–50

Linearity/accuracy (mm)

±3


Acuity AR700-4 Laser Displacement Sensor

Specification

Value

Model (P/N)AR700-4 5mW (AP7020040)
Laser class3R

Laser type

670nm 5mW visible RED diode

Span

101.6 mm

Resolution

5.1 microns

Operating temperature

0 to 50°C

Linearity/accuracy

±31 microns



VI. Credits


This document was written June 2019 by B. Novak and A. de Loach with contributions from G. Acton in the IODP wiki.  Portions of the manual were taken from the Archived P-Wave Velocity Gantry User Guide written by T. Cobine in 2009.

All improvements to the Quick Start Guides and User Guides are a communal effort, with honorable mention to the group of LOs, ALOs, and technicians who have helped.

VII. Archived Versions

...

Configuration Editor: Station Setup Tab

The Station Setup tab (Figure 5) stores the physical configuration measurements of the caliper and bayonet transducers.

  • Axis Offsets: physical offset from the laser zero point to the center of the transducer pair for the three stations (caliper, bayonet Y, bayonet Z). This measurement will not change unless the station location is physically changed.
  • Transducer Separation: physical distance between the bayonet transducer pair. This measurement should never change unless the holder is physically changed.

Image Removed
Figure 5. Configuration Editor: Station Setup Tab.

Configuration Editor: DAQ Setup Tab

The DAQ Setup tab (Figure 6) stores settings for pick and arrival calculations.

  • Stack Max: no longer used; no setting required.
  • Stack Level: maximum ± voltage level that the stacked signal must attain for a successful first pick calculation. The acquisition stacks for 10 seconds to arrive at this value. If after 10 seconds the value has not exceeded 5 V, the measurement is aborted. If this happens, reposition the sample under the transducer. Do not change the Stack Level value.
  • Stack Iterations: number of stacks that exceed ±5 V, averaged to reduce noise. Increasing this value will increase acquisition time.
  • Threshold Level: value used to detect the initial peak and subsequent first arrival. Do not change this value.
  • Peak Width: value used to detect the initial peak and subsequent first arrival. Do not change this value.
  • 2 MHz Filter: value used to filter out electrical noise. This value is set at user's discretion.

...

Caliper Calibration Tab

This tab (Figure 7) shows the results of the latest calibration and values used to calculate the calibration; values cannot be changed.

  • Caliper System Delay, Slope, Intercept, and r-Value: transducer system delay parameters
  • LVDT Slope, Intercept, and r-Value: calibration of LVDT volts output to millimeters
  • Linear Velocity and Liner Thickness: standard values used to correct the measurement of delay time through the core liner.

...

Bayonet Calibration Tab

This tab (Figure 8) shows the results of the latest bayonet calibration. Values cannot be changed.

  • Y-Bayonet axis system delay: water velocity corrected to temperature for Y-bayonet.
  • Z-Bayonet axis system delay: water velocity corrected to temperature for Z-bayonet.

...

VISA Resources Tab

This tab (Figure 9) shows the alias locations for the serial ports configured using LabView Measurement and Automation Explorer Program. Only Application Developers can change the values on this tab.
Figure 9. Configuration Editor: VISA Resources Tab.

File Paths Tab

...

Standards

  • Laboratory reagent water (distilled)
  • Aluminum cylinder of known length and velocity; velocity = 6295 m/s
  • Acrylic (as above) half-cylinders of variable thicknesses to accommodate transducer placement on the half-core section. As of August 2010, the velocity is ~2950 m/s but subject to further verification.

Image Removed
Figure 11. P-Wave Acrylic Cylinder Standards.

Instrument Calibration

The bayonet and caliper transducers must be calibrated and the system delay determined daily.

  • The bayonet transducers are calibrated by measuring the transducer spacing and backcalculating the system delay from the total traveltime for the y- and z-axes from the transducer separation value and theoretical velocity of distilled water at the temperature of the water bath.
  • Calibration of the caliper requires the transducer and system delay to be calibrated concurrently. The transducer separation is measured with an LDVT, which requires a line fit and linear interpolation to derive the transducer separation.

System delay is derived from a separation-traveltime curve. Separation is entered by the operator for the calibration standard and traveltime is derived from a traveltime pick algorithm. As the current transducers have pliable plastic pole faces the system delay is on the order of 25 µs.
The calibration uses the derived velocity of the standard material to test for acceptance or rejection of the data. Acrylic plexiglas, the material used for the calibration standards, has a published sound speed of 2750 m/s. The calibration pieces are cut from a single rod of plexiglas at variable thicknesses between 10 and 75 mm and are a numbered set for consistency.

Calibrating the Bayonet Transducers

Because the distance between the bayonets is fixed, only the system time delay needs to be calculated. This is achieved by determining by measuring velocity in water of known temperature and calculating the velocity correction. The difference between the known and measured value is due to system delay.

Procedure

  1. Prepare a water bath by setting ~2 L of distilled water at room temperature for a couple of hours.
  2. Once at room temperature, insert the thermocouple/digital thermometer and record the temperature of the water bath.
  3. On the Main window, select Bayonet Y (or Z) > then Station > then Y (or Z) Bayonet Calibration
  4. Enter the temperature of the water bath into the Temperature field on the calibration screen. The screen displays a plot of measured velocity vs. temperature of the water bath.

Lower the bayonets into the water bath (the black transducer should be covered). An algorithm
calculates the temperature-corrected velocity of the water bath and displays the result
5. Enter Bayonet Separation in the value field ( Y – 31.50, Z – 82.50 ).
6. The computer calculates the system delay for the y- and/or z-axes and displays the ultrasonic
waveforms on the corresponding graphic display. The software then displays the velocity
measurement of the water. Compare the measured velocity to the theoretical value. When
the graph shows an acceptable waveform click Accept, or Cancel, then repeat the calibration
procedure.
7. Once accepted, the system delay times are saved to the configuration file and the calibration
indicator and date time stamps are updated. If results are not acceptable, rerun the bayonet
calibration.
8. Repeat the calibration procedure for the other bayonet.

Calibrating the CALIPER

Caliper calibration is a little bit more involved. Because the distance between the calipers is variable, the system delay is determined using a set of standards of identical velocity but various lengths. The actual velocity of the standard is irrelevant, but is good to know as a quality check. To determine the system delay time, the first arrival time is plotted against distance and the intercept with "zero distance" equals the system delay (Figure 17). As a quality check the the slope will equal the velocity of the material used. This value should be close to the known value and the r-value (fit) should equal 1 (or be very, very close). The calibration should look like the example shown below (note: the delay time will change as the transducers wear).
Image Removed
Figure 17. Caliper System Delay.
In addition to the delay time determination, the LVDT output (volts) must be calibrated to distance in millimeters (Figure 18). The known lengths of the same standards are used and both calibrations are performed at the same time. As standards are measured, the LVDT calibration improves, so previous measurements are constantly recalculated. This is one of the reasons why at least 6 measurements are required before accepting calibration.

...

Procedure

  1. To begin calibration (Figure 19) to open the Caliper Calibration window for the PWV select the Caliper Piece X-Axis > then Stations > (Figure 20). to access the drop down menu, then > Calibrate Calipers.

...

  1. Place the cursor in the Scanner String field and scan the standard's 2D bar code. Once the information is automatically entered into the box, confirm the standard length and ensure the unit is meters. This value can be entered manually if necessary.
  2. Place the standard in the caliper and click Start Measurement. During acquisition, the waveform, time pick, and calibration plots (after two measurements) will update on the Caliper Calibration window (Figure 22).

Image Removed
Figure 22. Standard Acquisition.

  1. To stop the calibration, click the Stop – Return button on the Standard Information box (Figure 21). To resume measurement, click the Measure A Standard button on the Caliper Calibration window (Figure 22). To complete calibration, at least six standards must be measured; however, additional standards can be added or any standard can be measured multiple times.

Editing Calibration Points

Calibration points can be edited if necessary. For example, the calibration curve in Figure 23 shows an obvious problem with one point: The wrong length was assigned to the standard.
Image Removed
Figure 23. Calibration Curve with an Error.

Procedure

  1. Click the Stop – Return button on the Standard Identification box (Figure 21).
  2. Locate the Index number of the point containing the wrong information on the Caliper Calibration screen and type it into the Index field (Figure 24).
  3. Click Delete by Index to delete data for that point from the calibration.
  4. Remeasure the standard using the Measure A Standard button.
  5. Other options in the Array section of the Caliper Calibration screen include deleting only the last measurement (Delete Last) or deleting all data points and starting a new calibration (Delete All Data and Start Again) (Figure 24).

Image Removed
Figure 24. Array Section of the Caliper Calibration Screen.

Accepting Calibration

To accept the calibration, click the Save button on the Caliper Calibration window (Figure 22).

Sample Preparation

Two types of samples can be measured using the velocity gantry:

  • Section-half core samples
  • Discrete samples

Sample Prep Overview

  • Before placing the working half in the core tray, make sure the surface is clean (lightly scrape away any material smeared across the cut surface during core splitting).
  • Place the core section in the tray and make sure it is as flat as possible.

Instrument Preparation

  • Clean transducers of any residue with water and paper towels.
  • Before initializing the P-wave software, ensure the caliper transducers and laser beam are not blocked. Note: This step is crucial.

Sample Analysis

Measuring Section-Half Samples

Before measuring samples, be sure the samples are properly prepared and the system is calibrated.

  1. Start the Velocity application by clicking the icon (Velocity 2.0.5) on the desktop.
  2. Once the application is launched, click the Make a Measurement button.
  3. In the Scan Sample Label dialog box, place the cursor in the Scanner String field (Figure 25).

...

  1. Scan the barcode on the section half label to populate the fields on the Scan Sample Label dialog box. Enter Operator (user last name) and confirm measuring Station and Mode.

Stations Bayonet Y and Bayonet Z are for soft sediments. To use a Bayonet measuring station, the Mode value must be Section w/liner (material is still in core liner). Note: The orientation assigned to the sample measurement depends on the Station selected.

  1. For Section w/liner mode, a Position Sample dialog box opens (Figure 26). Confirm the sample position as measured by the laser (the laser at the end of the track measures the range to the end of the core and calculates the position in the liner based on an offset for that station). Click OK.

...

Image Removed
Figure 26. Position Sample Dialog Box.

  1. The bayonet transducers lower into the section half; ensure they are completely inserted into the sample. If contact between the transducers and sediment is poor, add distilled water around the transducers to improve contact with sediment.
  2. On the JOG Bayonet screen (Figure 27), monitor the waveform and position the transducer to obtain the best trace. The plot displays the live signal with no stacking. A weak signal is fine, as long as a clean first arrival is present. Once the waveform is acceptable, click Continue to begin sample acquisition.

Image Removed
Figure 27. JOG Bayonet Y Screen.

  1. During acquisition, the program adds multiple waveforms until the total (max peak) is greater than ±5 V. This process is repeated and the waveform is averaged.

Measuring Discrete Samples

Discrete samples measured on the Cailper station are hard material that has been cut from the core as cubes, minicores, or slabs. Because the material can be measured in various orientations, the user must select the measurement axis (see Figure Appendix A).
Traveltime is calculated as total traveltime minus x-system delay time. Discrete sample measurements are not corrected for the core liner nor is an offset recorded in the LIMS. The transducer separation is measured with an LDVT as it is for the sample half measurement. For discrete samples the axis of measurement is selected for each measurement.
Retaining the correct "up" direction is critical for axis determination and P-mag orientation; make sure the sample has the "up" direction marked on it (see Figure Appendix A).
Before measuring samples, be sure the samples are properly prepared and the system is calibrated.

  1. Start the Velocity application by clicking the icon (Velocity 2.0.5) on the desktop.
  2. Once the application is launched, click the Make a Measurement button.
  3. In the Scan Sample Label dialog box, place the cursor in the Scanner String field (Figure 28).

...

  1. Scan the barcode on the discrete sample label to populate the fields on the Scan Sample Label screen. Enter Operator (user last name) and confirm measuring Station (Caliper) and Mode (Discrete).
  2. Enter information requested in the Additional Discrete Information dialog box (Figure 29): Measurement Axis: X, Y, or Z.

Image Removed
Image Removed
Figure 29. Additional Discrete Information Dialog Box.

  1. If the material to be measured is a whole piece removed from the liner (rarely used), select Piece w/o Liner mode in Step 4 and complete the offset (position of the piece in the liner before it was removed) in the Additional Piece Information dialog box (Figure 30).

If the orientation of the piece is unknown, check the Orientation Unknown box; otherwise the measurement is assigned to the X-direction.
Image Removed
Figure 30. Additional Piece Information Dialog Box.

  1. Place the discrete sample in the caliper and add distilled water to create a good sample-transducer contact if needed.
  2. On either the Additional Discrete Information (discrete samples) or Additional Piece Information (piece samples) box, select OK - AutoClose (for hard samples) or OK – Manual Close (for softer/friable samples).

AutoClose automatically closes the caliper on the sample. The transducer moves to a predefined position or until the actuator stalls.

  1. On the JOG Caliper screen (Figure 31), monitor the waveform and position the transducer to obtain the best trace. The plot displays the live signal with no stacking. A weak signal is fine, as long as a clean first arrival is present. Once the sample is properly positioned, click Continue to begin sample acquisition.

Image Removed
Figure 31. JOG Caliper Screen.

  1. During acquisition, the program adds multiple waveforms until the total (max peak) is greater than ±5 V. This process is repeated and the waveform is averaged.

The touch-screen Open/Close commands may be sluggish in response. If there is no response, try actuating the button using the mouse.

  1. Repeat the measurement process for each axis to be measured.

Accepting Results and Manual Pick

Once the velocity data have been acquired, the results are shown on the Display Results screen (Figure 32).
Image Removed
Figure 32. Display Results Screen.
The Manual Pick Velocity field remains blank until the user clicks the Manual Pick Arrival Time button and sets a manual pick time. The first arrival should be picked in the same manner the system was calibrated. Both Manual Pick and Auto Pick are saved with the sample data.

  1. Select the Manual Pick Arrival Time button to open the Manual Pick screen (Figure 33.A), showing the waveform. To help select the first arrival, the plot can be zoomed in using the graph palette zoom function or by entering Zoom Range Minimum and Zoom Range Maximum (27. to 40.) below the waveform plot on the Manual Pick screen. Select the Cursor Movement Tool on the graph palette, then use the mouse to drag the red cursor line to the first arrival, this should be the slope of the first positive peak at zero (Figure 33.B).

Image Removed
Figure 33.A Manual Pick Screen. 

  1. Click Continue to accept the manual pick and save it to the data file or Cancel to discard the manual pick results.

Data Handling

From the Display Results screen one of 4 data handling choices can be selected:

  • Exit Save Data: saves data and returns to the Sample Information dialog box for the next sample.
  • Cancel Without Saving Data: does not save the data but returns to the Sample Information dialog box. Use this option if there is a data quality warning or if the waveform for the sample was unacceptable.
  • Repeat Without Saving Data and Keep ID: does not save data but returns to the JOG screen where the user can reposition the transducer and take another measurement of the same sample.
  • Save Data and Repeat with Same ID: saves the data and returns to the JOG screen where the same sample can be run again. This option allows replicate measurements.

LIMS Integration

Sample/Analysis Attributes and Components

...

Analysis

...

Component

...

Unit

...

Description

...

PWAVE_C or
PWAVE_B

...

bottom_depth

...

m

...

Location of bottom of individual measurement on the sample, measured from top of hole

...

 

...

distance_in_caliper

...

mm

...

Displacement between the faces of the transducers in contact with the sample

...

 

...

instrument_group

...

...

Core logger on which the sensor is deployed

...

 

...

liner_correction

...

...

Liner correction value applied to calculation:
0: measurement on discrete sample
1: measurement on section half passing through 1 core liner

...

 

...

number_of_readings

...

...

Number of signals stacked together while obtaining this result

...

 

...

offset

...

cm

...

Location of measurement on sample measured from top

...

 

...

run_asman_id

...

...

Serial number of the raw P-wave data file in the digital asset management system (ASMAN)

...

 

...

run_filename

...

...

Filename of raw P-wave data file

...

 

...

top_depth

...

m

...

Location of top of individual measurement on the sample, measured from top of hole

...

 

...

travel_time

...

µs

...

Traveltime of the sonic wave from transducer to transducer through the sample

...

 

...

velocity

...

m/s

...

Velocity of pressure wave through sample

...

 

...

velocity_x

...

m/s

...

Velocity of pressure wave through sample in the x-axis

...

 

...

velocity_x

...

m/s

...

Velocity of pressure wave through sample in the y-axis

...

 

...

velocity_y

...

m/s

...

Velocity of pressure wave through sample in the x-axis

Data Upload Procedure

Before running a sample or completing a calibration, you must start the Data Uploader so the data uploading process will run in the background looking for files to upload and every section measured on the logger will automatically upload to LIMS.

  1. Open the LIMS Uploader icon on the desktop.
  2. Click Start Monitoring to save data files and automatically upload calibration files to LIMS.
  3. When data collection is completed, click Stop Monitoring.

Health, Safety, & Environment

Safety

  • Keep extraneous items and body parts away from the moving parts.
  • The track system has a well-marked yellow and red emergency stop button (Figure 34) to halt the system if needed.
  • Do not look directly into the laser light source (class 2 laser product).
  • Do not direct the laser beam at other people.
  • Do not attempt to work on the system while a measurement is in progress.
  • Do not lean over or on the track.
  • Do not stack anything on the track.
  • This analytical system does not require personal protective equipment.

...

Figure 34. Emergency Stop Button.

Maintenance & Troubleshooting

Common Issues

The following are common problems encountered when using the P-wave Velocity Gantry and their possible causes and solutions. For information about the laser Measurement and Operation software, see the Vp GIESA Gantry Laser Operation manual.

Issue

Possible Causes

Solution

Laser sensor: no laser light/no laser range data

Sampling is turned off

Turn sampling on through laser Measurement and Operation software

 

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

Laser sensor: Serial port not responding

Power supply voltage too low

Check power supply input voltage through laser Measurement and Operation software

 

Baud rate incorrect or unknown

See Acuity manual 3.5.1; laser Measurement and Operation software

Laser sensor: Error code (Exx) transmitted on serial port

 

See Acuity manual 5.3; laser Measurement and Operation software

Exlar actuator: No response

I/O error

Check drive or Expert software for faults
Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

 

Power interruption

Check wiring

Exlar actuator: Behaving erratically

Drive improperly tuned

Check gain settings
Use MOTION1/MOTION2/JOG commands. Use arrow up to upload parameters

 

Too much load on motor

Check for load irregularity or excess load

Exlar actuator: Cannot move load

Too much load or friction

Check load and friction sources

 

Excessive side load

Check side load

 

Misalignment of output rod to load

Check alignment of rod and load

 

Current limit on drive too low

Check current on drive

 

Power supply current too low

Check power supply current

Exlar actuator: Housing vibrates when shaft in motion

Loose mounting

Tighten mounting screws

 

Drive improperly tuned

Check gain settings

Exlar actuator: Output rod rotates during motion

Rod rotation prevents proper linear motion

Install anti-rotation assembly

Exlar actuator: Overheating

Ambient temperature too high

Check ambient temperature

 

Actuator operation outside continuous ratings

Check operation settings

 

Amplifier poorly tuned

Check gain settings

Actuator Utilities

The actuator utilities can be used to test the caliper and bayonet actuators (move them up and down) and view the waveform detected by the transducers. There are also utilities to test the home position of the actuators and to check the laser function. This utility is useful when configuring the station offsets.
To change the behavior and/or force of the actuators, the Exlar utility must be used.
Open the Actuator utility (Figure 35) from Main > Actuator Utility and then select one of the JOG screens, Home All, or Read Laser.
Image Removed
Figure 35. Actuator Utilities.

Scheduled Maintenance

After Every Sample

Clean contact sensors with tissue, cloth, or paper towel and water.

Daily

Clean dust from the laser sensor lens using compressed air or delicate tissue wipes. Do not use organic cleaning solvents on the sensor lens.

Weekly

Check set screws on piezoelectric transducers to make sure they are firmly held in position when embedding in sample.

Annually

Coat (not pack) the following parts with grease:

  • Angular contact thrust bearings
  • Roller screw cylinder
  • Roller screw assembly in the actuators

Examine the cable management system for abraded cables or other indications of wear.

Vendor Contact/Consumable Parts

Panametrics transducer

Olympus (www.olympusndt.com)

  • Transducer: NDT M2008
  • Light source: HL-2000-FHSA or HL-2000-FHSA-LL
  • Spare bulb: HL-2000-B or HL-2000-LL-B
  • Reflectance standard: WS-1-SL
Laser displacement sensor (AR1000)

Acuity Laser Measurement
www.acuitylaser.com
702-616-6070

Barcode reader (MS-4)

Microscan
www.microscan.com
helpdesk@microscan.com
800-251-7711

Actuator (TLM 20)

Tritex
www.exlar.com

...