Introduction


The principles of spectroscopic analysis rely on Beer's law. The principle of Beer's law is that passing light of a known wavelength through a sample of known thickness and measuring how much of the light is absorbed at that wavelength will provide the concentration of the unknown, provided that the unknown is in a complex that absorbs light at the chosen wavelength. 

IODP's Agilent Technologies Cary 100 double-beam UV-Vis (ultraviolet–visible) spectrophotometer is ideal for shipboard routine laboratory work. The system measures analytes in interstitial water obtained from sediment cores using standard colorimetric methodology.

The described methods are based on ODP Technical Note 15, Chemical Methods for Interstitial Water Analysis Aboard the JOIDES Resolution, Aug 1991; J.M. Gieskes, T. Gamo, and H. Brumsack.

Methods


Ammonium

Determination of ammonium concentration is of importance because this constituent is an indicator of diagenesis of organic matter in the sediments. The onset of sulfate reduction coincides with initiation of ammonium ion production. Ammonium production increases strongly in the zone of methanogenesis, presumably as a result of associated deammonification reactions. The large potential variation in ammonium concentrations, therefore, suggests that a few preliminary ammonium concentrations should be run in order to set limits to the sample dilution and range of standards to be used. Suggestions for this follow below.

The methodology is based on Solorzano (1969), originally developed to detect very small NH4+ concentrations in seawater. Although background contamination problems in seawater are enormous, the relatively high concentrations of ammonium in pore fluids (as high as 85 mM in ODP Leg 112 samples; Kastner et al., 1990) minimizes this problem when matrix blanks are run along with the samples. In areas of low sedimentation, however, very low ammonium concentrations require careful sample handling to avoid this problem. 

The ammonium method is based on diazotization of phenol and subsequent oxidation of the diazo compound by Chlorox™ to yield a blue color, measured spectrophotometrically at 640 nm.

Reagents

Reagent Solutions

Ammonium Standard

(0.10 M)

make once an expedition

Dry ammonium chloride (NH4Cl) overnight in an oven.

Dissolve 5.349 g dried NH4Cl in a

1 L

1000 mL volumetric flask. Bring to volume with nanopure water.

For
  • To make 25 mL - Dissolve 133.728 mg dried NH4Cl in a 25 mL volumetric flask. Bring to volume with nanopure water.

Alkaline

solution

make once an expedition

Dissolve 7.5 g (tri)sodium citrate tribasic dihydrate (Na3C6H5O7·2H2O) and 0.4 g sodium hydroxide (NaOH) in a 500 mL volumetric flask. Bring to volume with nanopure water.

Oxidizing

solution

make fresh daily

Add 1 mL fresh sodium hypochlorite (4% available chlorine) to 50 mL Alkaline solution. This should be adjusted for the amount of samples to be run.

  • Use sodium hypochlorite or regular strength Chlorox™ bleach. DO NOT use Chlorox that contains NaOH, fragrances, or other agents, and do not use another brand of bleach.
  • The Clorox bleach can go bad with time. Absorbance values of the ammonium analysis will be much lower when this happens. Absorbance of the 1000 uM standard falling below 0.2 or the 100 uM standard falling below 0.01 could be indications that a new bottle of bleach is needed.

Phenol

solution

make fresh daily

Add 1 mL liquid Phenol to 100 mL Absolute Ethanol 

Sodium Nitroprusside

solution

make fresh daily

Dissolve 75 mg sodium nitroprusside (Na2[Fe(CN)5NO]) (also known as sodium nitroferricyanide) in 100 mL nanopure water

Standards

Standards

Add standard to a 50 mL volumetric flask and bring to volume with nanopure water.

50 mL batches are stable for 1 month. 

concentration

(µM)

volume of ammonium standard

(mL)

volume of nanopure water

(mL)

0

0

50.000

50

0.025

47.975

100

0.050

49.950

150

0.075

49.950

200

0.100

49.900

400

0.200

49.800

600

0.300

49.700

800

0.400

49.600

1000

0.500

49.500

1500

0.750

49.250

2000

1.000

49.000

3000

1.500

48.500

Procedure

Concentrations of ammonium may differ quite a bit at different sites. Typically in areas with strong evidence of organic carbon diagenesis (e.g., organic carbon–rich sediments), high concentrations of NH4+ can be expected. In that case, sample aliquots must be made appropriately small or sample dilution may be required. The range can be established by using a sample near the alkalinity maximum. Once the range has been determined, prepare standards that cover this range. In this manner, samples and standards are treated in a similar way.

The high range of the instrument is 3.0 absorbance units (Abs). The results are linear up to this point. If the readings are expected to be higher than 3.0 Abs, use a sample volume of 30 µL instead of 100 µL and divide measured concentrations by 0.3 to arrive at the final sample concentration.

Care should be taken to use clean vials, preferably not those used for Si and PO4 determinations, in which ammonium molybdate is used as a reagent.

Note:  Use a smaller aliquot of sample if the result exceeds the linear range of the spectrophotometer, making up the volume with nanopure water. (For example, for a 100 µL aliquot of a sample, add 2.1 mL nanopure water.)

NoteThe order of dilution (below) matters, so do not change this order. Shake samples after EACH addition.

1.

Transfer 200 µL of sample/standard to

an 8 mL

a vial.

2.

Add 2 mL of nanopure water to each vial

3.

Add 1 mL phenol solution to each vial and shake.

4.

Add 1 mL sodium nitroprusside solution to each vial and shake.

5.

Add 2 mL of oxidizing solution to each vial and shake.

6.

Let the color develop (in a dark place) for 6.5 hr

*

and then determine the absorbance at 640 nm wavelength.

*

(

from

From a series of measurements over 8 h, it was found that results stabilized after 6.5 hr.)

Phosphate

Determination of dissolved phosphate, particularly in rapidly deposited organic carbon–rich sediments, is important in the shipboard analytical program. Phosphate concentrations may vary considerably, and it is therefore advisable to obtain a preliminary idea of the concentration ranges to be expected. This can most easily be accomplished by taking samples in the region of maximum alkalinities. Typically if alkalinities are >30 mM, dissolved phosphate concentrations may be >100 µM; thus, only very small sample aliquots will be needed to establish the concentration range.

This method is, in essence, the colorimetric method from Strickland and Parsons (1968) as modified by Presley (1971) for DSDP pore fluids. Orthophosphate reacts with Mo(VI) and Sb(III) in an acidic solution to form an antimony-phosphomolybdate complex. Ascorbic acid reduces this complex to form a blue color, and absorbance is measured spectrophotometrically at 885 nm.

It is important to note that the concentrations in the final test solution cannot exceed ~10 µM. Thus, for open-ocean (low sedimentation rate, low organic carbon) sediments, one might need to do the determination on 2 mL of sample (expected range 0–10 µM), but in typical continental margin settings, where concentrations can exceed 100–200 µM, a 0.1 or 0.2 mL sample aliquot must be used. The concentration range must be established prior to running samples, and it is highly advisable to make standards that cover the range of concentrations to be expected.

Note: Samples with high silica concentrations may give a false increase in measured concentration of phosphate (http://dx.doi.org/10.1007/BF02071829; S. Noriki, Silicate correction in the colorimetric determination of phosphate in seawater, 1983).

Reagent Solutions

A: Sulfuric Acid solution (H2SO4) (1000 mL) – stable indefinitely

Carefully add 10 mL concentrated H2SO4 to ~ 600 ml of nanopure water

Phosphate Standard (0.01 M)

make once an expedition

Dry potassium phosphate monobasic (KH2PO4) in oven at 100°C for two hours; keep in a desiccator while it cools before weighing.

Dissolve 1.361 g dried KH2POin a 1000 mL volumetric flask.

  CAUTION: Mixing sulfuric acid and water produces heat. Take appropriate precautions. Allow to cool to room temperature before bringing flask to volume (1000 mL)

Bring to volume with nanopure water.

  • To make 25 mL - Dissolve 34.025 mg dried KH2PO in a 25 mL volumetric flask. Bring to volume with nanopure water.

B. Antimony Potassium Tartrate solution (1000 mL) – make once an expedition; store in amber glass @ 4°C

In

Ammonium Molybdate

stable indefinitely

Dissolve 2 g ammonium molybdate tetrahydrate ([NH4]6Mo7O24·4H2O) in a 1000 mL volumetric flask

, dissolve 0.102 g of antimony potassium tartrate trihydrate (KSbC

. Bring to volume with nanopure water.

  • Store in polyethylene @ 4°C

Sulfuric Acid

stable indefinitely

Add 10 mL concentrated sulfuric acid (H2SO4) to ~600 mL nanopure water in a 1000 mL volumetric flask. Bring to volume with nanopure water.

  • Caution: Mixing sulfuric acid and water produces heat. Take appropriate precautions. 

Antimony Potassium Tartrate

make once an expedition

Dissolve 102 mg antimony potassium tartrate trihydrate (KSbC4H4O7·3H2O) in

~ 600 mL of

a 1000 mL volumetric flask. Bring to volume with nanopure water.

(

  • If using antimony potassium tartrate hemihydrate [KSbC4H4O7·½H2O], dissolve
0.09 g.) Bring flask to volume (1000 mL) with nanopure water.
  • 90 mg. 
  • Store in amber glass @ 4°C

Ascorbic Acid

make fresh weekly

Dissolve 3.5g ascorbic acid (C6H8O6) in a 1000 mL volumetric flask. Bring to volume with nanopure water.

  • If

C. Ammonium Molybdate solution (1000 mL) – stable indefinitely; store in polyethylene @ 4°C

In a 1000 mL volumetric flask, dissolve 2 g of ammonium molybdate tetrahydrate ([NH4]6Mo7O24·4H2O) in ~ 600 mL of nanopure water. Bring flask to volume (1000 mL) with nanopure water.

D. Ascorbic Acid solution (1000 mL) – make fresh weekly; store in amber glass @ 4°C

In a 1000 mL volumetric flask, dissolve 3.5 g of ascorbic acid (C6H8O6) in ~ 600 mL nanopure water. Bring flask to volume (1000 mL) with nanopure water. Note: If
  • the reagent is discolored upon creation, the dry ascorbic acid is probably oxidized and must be replaced.
E. Mixed reagent (250 mL) –

Mixed Reagent

make fresh every 6 hours

Add

Mix the following solutions

to an appropriate container. Do not add the ascorbic acid reagent until immediately before use.

. Mix well after each addition. Can adjust the volumes to be suitable for the number of samples as long as the proportions of each reagent are maintained.

  • 50 mL Ammonium Molybdate
  • 125 mL Sulfuric Acid
  • 25 mL Antimony Potassium Tartrate
  • 50 mL Ascorbic Acid

Standards

Add standard to a 50 mL volumetric flask and bring to volume with nanopure water.

concentration

(µM)

volume of phosphate standard

(mL)

volume of nanopure water

(mL)

0

0

50

5

0.025

49.975

10

0.050

49.950

15

0.075

49.925

20

0.100

49.900

40

0.200

49.800

60

0.

Mix well after each addition to prevent the solution from darkening.

solution

volume (mL)

Ammonium molybdate (Reagent C)

50

Sulfuric acid (Reagent A)

125

Antimony potassium tartrate (Reagent B)

25

Ascorbic acid (Reagent D)

(immediately before use)

50

Phosphate standard: 0.01 M Phosphorus (1000 mL) – make fresh once an expedition; store @ 4°C

Dry potassium phosphate monobasic (KH2PO4) in oven at 100°C for two hours; keep in a desiccator while it cools before weighing.

In a 1000 mL volumetric flask, dissolve 1.361 g oven-dried KH2PO4 in nanopure water, then bring flask to volume (1000 mL) with nanopure water.

To make 25 mL of the standard (25mL is enough): in a 25mL volumetric flask, dissolve 34.025 mg dried KH2PO4 in nanopure water; bring flask to volume (25 mL) with nanopure water. 

Standards

concentration

(µM)

volume of phosphate standard

(mL)

volume of nanopure water

(mL)

0

0

50

5

0.025

49.975

10

0.050

49.950

15

0.075

49.925

20

0.100

49.900

40

0.200

49.800

60

0.

300

49.700

80

0.400

49.600

100

0.500

49.500

1500.75049.250

200

1.000

49.000

300

1.500

48.500

Procedure

  • Add 2 mL nanopure water to a 8 mL vial.
  • Note:  Use a smaller aliquot of sample if the result exceeds the linear range of the spectrophotometer, making up the volume with nanopure water. (For example, for a 300 µL aliquot of a sample, add 300 µL nanopure water.)

    1.

    Transfer 600 µL sample/standard to a vial.

    2.

    Add 2 mL nanopure water to each vial.

    3
    Add 600 µL sample or standard
    .

    Add 4 mL mixed reagent to each vial and shake.

    Shake well
    4.

    After a few minutes a blue color develops, which remains stable for a few hours.

    It is best to make the readings at 885 nm ~ 30 min after addition of the mixed reagent.


    Silica

    Silicon is routinely measured on the

    Note: Use a smaller aliquot of sample if the result exceeds the linear range of the spectrophotometer, making up the volume with nanopure water. (For example, for a 300 µL aliquot of a sample, add 300 µL nanopure water.)

    Silica

    Silicon is routinely measured on the ICP, so measurement by spectroscopic analysis can be considered an alternate method.

    Dissolved silica determinations are of great importance in interstitial waters. Often they represent the lithology of the sediments, and the concentrations can vary substantially, especially if highly dissolvable phases such as biogenic opal-A, volcanic ash, or smectite are present. Thus, a wide range of concentrations can be expected, typically from 50 to 1200 µM or higher (especially in hydrothermally affected sediments). The method below usually covers the range, although greater dilutions may be appropriate if sediments or sample sizes necessitate this.

    This method is based on the production of a yellow silicomolybdate complex. The complex is reduced by ascorbic acid to form molybdenum blue, measured at 812 nm. The blue complex is very stable, which will enable delayed reading of the samples.

    Reagent Solutions

    A: Sulfuric Acid stock solution (500 mL) – make fresh every month; store in polyethylene

    Slowly pour 250 mL of concentrated sulfuric acid into a 500 mL volumetric flask containing ~ 200 mL of nanopure water. (Caution: Mixing concentrated

    Silica Standard (3000 µM)

    make once an expedition

    Dry sodium silicofluoride in a vacuum desiccator overnight to remove excess water.

    Do not heat.

    Sulfuric Acid

    make fresh monthly

    Slowly add 250 mL concentrated sulfuric acid to ~200 mL nanopure water in a 500 mL volumetric flask. Allow to cool to room temperature, then bring flask to volume with nanopure water. 

    • Caution: Mixing sulfuric acid and water produces heat. Take appropriate precautions.
    )
    Allow to cool to room temperature before diluting to 500 mL with nanopure water.

    B. Ammonium Molybdate Tetrahydrate solution (500 mL) – make fresh every month; store in dark polyethylene @ 4°C

    In a 500 mL volumetric flask, dissolve 4 g of
    •  

    Synthetic Seawater

    make fresh monthly

    Dissolve 25 g sodium Chloride in ~800 mL nanopure water in a 1000 mL volumetric flask. Add and disolve 8 g magnesium sulfate heptahydrate (MgSO4·7H2O). Bring flask to volume with nanopure water

    Ammonium Molybdate

    make fresh monthly

    Dissolve 4 g ammonium molybdate tetrahydrate ((NH4)6MO7O24·4H2O) in ~300 mL

    of

    nanopure water in a 500 mL volumetric flask. Add 12 mL

    of

    concentrated hydrochloric acid (HCl)

    acid, mix, and make up to 500 mL

    . Bring flask to volume with nanopure water.

    Note: If a white precipitate forms, make fresh reagent.

     

    Metol Sulfite

    make fresh monthly

    C: Metol Sulfite solution (500 mL) – make fresh every month; store in amber glass, tightly sealed @ 4°C

    Dissolve 6.0 g anhydrous sodium sulfite

    ,

    (Na2SO3

    ,

    ) in a 500 mL volumetric flask. Add 10 g Metol (p-methylaminophenol sulfate [(C7H10NO)2SO4])

    and then

    . Bring flask to volume with nanopure water

    to make the volume to 500 mL

    . When the Metol has dissolved, filter the solution through a Whatman No. 1 filter paper. 

    • Note:
    This
    •  This solution may deteriorate quite rapidly.
    D:

    Oxalic Acid

    solution (500 mL) –

    make fresh

    every month; store in glass

    monthly

    Add 50g

    In a 500 mL volumetric flask, shake 50 g of analytical-grade

    oxalic acid dihydrate [(C2H4O2)·2H2O]

    in ~ 300

    to ~300 mL

    of

    nanopure water in a 500 mL volumetric flask. Shake well, and bring

    to 500 mL

    flask to volume with nanopure water. Let stand overnight. Decant saturated solution of oxalic acid from crystals before use.

    E: Reducing solution (150 mL) –

    Reducing

    make fresh daily

    Mix
    • mix 50 mL
    reagent C (
    • Metol Sulfite solution
    )
    • with 30
    mL reagent D (oxalic acid solution). Add
    • mL Oxalic Acid solution.
    • add slowly, with mixing, 30
    mL reagent A (50% sulfuric acid solution) and bring
    • mL Sulfuric Acid solution
    • add 40 mL nanopure water to bring total volume to 150 mL
    with nanopure water.

    Standards

    Add standard to a 50 mL volumetric flask and bring to volume

    F: Synthetic seawater (1000 mL) – make fresh monthly; store in polyethylene

    In a 1000 mL volumetric flask dissolve 25 g sodium chloride in ~ 800 mL nanopure water, then add 8 g magnesium sulfate heptahydrate (MgSO4·7H2O). Dilute to 1000 mL

    with nanopure water.

    concentration

    G. 3000 µM Silica Primary Standard (1000 mL) – make fresh every 2 months; store in polyethylene

    Note: When sodium silicofluoride (Na2SiF6) is dissolved in water, it hydrolyzes to form reactive dissolved silica. Place ~ 1 g of Na2SiF6 in an open vial and place in a vacuum desiccator overnight to remove excess water. Do not heat.

    In a 1000 mL polyethylene volumetric flask, dissolve 0.5642 g dried sodium silicofluoride in ~ 800 mL nanopure water. Dissolution is slow; allow at least 3 minutes. Dilute the solution to 1000 mL with nanopure water.

    Standards

    Using a 50 mL volumetric flask, add the following amounts of primary standard to approximately 30 mL of nanopure water and then bring to a total of 50 mL. Store in polyethylene containers.

    concentration (µM)

    volume of primary standard (mL)

    volume of nanopure water (mL)

    30

    0.5

    49.5

    60

    1

    49.0

    120

    2

    48.0

    240

    4

    46.0

    360

    6

    44.0

    480

    8

    42.0

    600

    10

    40.0

    900

    15

    35.0

    1200

    20

    30.0

    Procedure

    Make sure that all reagents are prepared ahead of time. The method has a time factor built in, and therefore it is of great importance to have all necessary reagents ready to go.

    1. Pipette 4 mL of nanopure water into the vials (3.8 mL for standards and blanks).
    2. For standards and blanks (nanopure water), pipette 200 µL of synthetic seawater into the vials.
    3. Pipette 200 µL of sample/standard/blank into the vials.
    4. Record time.
    5. Add 2 mL of molybdate solution (reagent B) to the vials. A yellow color will develop; allow to mature for exactly 15 minutes (± 15 s).
    6. Add 3 mL of the reducing solution (reagent E).
    7. Cap the vials and let stand for at least 3 hours.
    8. Read absorbances on the spectrophotometer at 812 nm.
    Procedure notes
    • Do not handle more than about thirty samples at a time in order to ensure that the 15 min time limit can be adhered to. Make sure that there are no large fluctuations in room temperature.
    • Do not use synthetic seawater in dilutions of the primary standard. This could cause the decrease in reactive silica in a few hours as a result of polymerization reactions.
    • The reason for adding 200 µL of synthetic seawater to the standards is to maintain a reasonably uniform salt content in relation to the samples, this suppressing a potential salt effect on the method.
    • It is important to wait at least three hours for the blue color to develop; the higher the concentration, the longer the time. The color remains stable for many hours, and reading after 4–5 hours may, in fact, be a good idea. Again, consistency in time limits is advisable.

    Sulfide

    The sulfide method is based off a method developed by Cline in 1969. This method called for very large volumes of water (50 mL). This method was modified on the BONUS Baltic Gas expedition in 2011 to work with sample volumes in the 1-5mL range.
    The method is a bit tricky in that the reagent concentrations change depending on what concentration range your samples fall in:

    (µM)

    volume of primary standard

    (mL)

    volume of nanopure water

    (mL)

    0050

    30

    0.5

    49.5

    60

    1

    49.0

    120

    2

    48.0

    240

    4

    46.0

    360

    6

    44.0

    480

    8

    42.0

    600

    10

    40.0

    900

    15

    35.0

    1200

    20

    30.0

    Procedure

    Make sure that all reagents are prepared ahead of time. The method has a time factor built in, and therefore it is of great importance to have all necessary reagents ready to go.

    Do not handle more than about thirty samples at a time in order to ensure that the 15 min time limit can be adhered to. Make sure that there are no large fluctuations in room temperature.

    Do not use synthetic seawater in dilutions of the primary standard. This could cause the decrease in reactive silica in a few hours as a result of polymerization reactions.

    The reason for adding 200 µL of synthetic seawater to the standards is to maintain a reasonably uniform salt content in relation to the samples, this suppressing a potential salt effect on the method.

    It is important to wait at least three hours for the blue color to develop; the higher the concentration, the longer the time. The color remains stable for many hours, and reading after 4–5 hours may, in fact, be a good idea. Again, consistency in time limits is advisable.


    1.

    add 4 mL of nanopure water to each vial (3.8 mL for standards).

    2.

    for standards, pipette 200 µL of synthetic seawater to each vial.

    3.add 200 µL sample/standard to each vial.
    4.Record time.
    5.

    add 2 mL ammonium molybdate solution.

    6.a yellow color will develop; allow to mature for exactly 15 minutes (± 15 s).
    7.add 3 mL reducing solution.
    8.Let color develop for at least 3 hours. Read absorbances at 812 nm.

    Sulfide

    The sulfide method is based off a method developed by Cline in 1969. This method called for very large volumes of water (50 mL). This method was modified on the BONUS Baltic Gas expedition in 2011 to work with sample volumes in the 1-5mL range.
    The method is a bit tricky in that the reagent concentrations change depending on what concentration range your samples fall in:

    • High range: 6-80 uM
    • Low range: 1-10 uM.

    All samples need to be preserved during splitting with a 1% zinc acetate solution. It can require a relatively large amount of sample in order to do this analysis. The high range method uses less sample (~500 µL), but if the sulfide level is below 6 µM, it won't be enough. It may still be worth screening samples with the high-range method in order to conserve interstitial water sample volume. The low-range method requires 4 mL (8x the amount of sample) but is sensitive down to 1 µM. The scientists will have to determine if consumption of 4 mL, possibly 4.5 mL, of sample is worth obtaining the sulfide concentration.

    ReagentSolutions

    Zinc Acetate (1%)

    For Standard:

    • Dissolve 10 g zinc acetate dehydrate into ~600 mL nanopure water in a 1000 mL volumetric flask. Add 1 mL concentrated acetic acid and bring flask to volume with nanopure water. Mix well.

    For Splits:

    • Dissolve 1 g zinc acetate dehydrate into ~60 mL nanopure water in a 100 mL volumetric flask. Add 100 µL concentrated acetic acid and bring flask to volume with nanopure water. Mix well.

    Zinc-Sulfide Standard Suspension (1 mM)

    • Before weighting sodium sulfide nonahydrate (Na2S·9H2O), wash the crystals with nanopure water and dry with kimwipe to remove oxidation products.
    • Dissolve 200 mg of Na2S·9H2O in 1000 mL 1% Zinc Acetate solution.
    • Store in plastic @ 4°C. Shake vigorously each time before using.
    Zinc-Sulfide Standard (100 µM)Dilute 10 mL of Zinc-Sulfide Standard Suspension to 100 mL with nanopure water in a 100 mL volumetric flask.
    High-Range Sulfide Diamine 

    For the high-range method: 

    • add 500 mL concentrated HCl to 500 mL nanopure water. Mix thoroughly and allow to cool to room temperature. Add 4 g N,N-dimethyl-p-phenylenediamine sulfate and 6.0 g iron chloride hexahydrate (FeCl3·6H2O). Mix well.
    • Store in amber Nalgene bottle at 4°C

    Low-Range Sulfide Diamine

    For the low-range method:

    • add 250 mL concentrated HCl to 250 mL nanopure water. Mix thoroughly and allow to cool to room temperature. Add 500 mg of N,N-dimethyl-p-phenylenediamine sulfate and 750 mg iron chloride hexahydrate (FeCl3·6H2O). Mix well.
    • Store in amber Nalgene bottle at 4°C
    Dilution Reagent

    This reagent is used to dilute too-dark samples into the range of color covered by the standard curve, while keeping a constant concentration of the diamine reagent. Create this reagent as appropriate for the high-range or low-range method:

    • For high-range method:  add 800 µL of High-Range Sulfide Diamine to 10 mL of nanopure water.
    • For low-range method:  add 4 mL of Low-Range Sulfide Diamine to 50 mL nanopure water.

    High-Range Method 

    Sample Preservation

    The IW splits for sulfide should be preserved immediately with the 1% zinc acetate solution to prevent the loss of dissolved sulfide.

    Add 40 µL of 1% zinc acetate to 0.5 mL sample aliquots.

    Standards

    Make dilutions of the 100 µM zinc-sulfide standard for the standards to create 0.5 mL of standard at each level.

    concentration

    (µM)

    zinc-sulfide standard

    (µL)

    nanopure water

    (µL)

    0

    0

    500

    6

    30

    470

    20

    100

    400

    40

    200

    300

    60

    300

    200

    80

    400

    100


    Procedure

    1.Shake well the zinc acetate preserved sample.
    2.add 500 µL sample/standard to each vial
    3.add 40 µL high-range diamine reagent to each sample/standard and shake.
    4.

    let color develop (in a dark place) for 30 minutes. Measure absorbance at 670 nm.

    • If any samples are darker blue than the highest standard, dilute the sample with the dilution reagent until it is within the color range of the standards.

    Low-Range Method

    Sample Preservation

    The IW splits for sulfide should be preserved immediately with the 1% zinc acetate solution to prevent the loss of dissolved sulfide.

    Add 800 µL of 1% zinc acetate to 4 mL sample aliquots.

    Standards

    Make dilutions of the 100 µM zinc-sulfide standard for the standards to create 4 mL of standard at each level.

    concentration

    (µM)

    zinc-sulfide standard

    (µL)

    nanopure water

    (mL)

    004000
    1403.960
    2.51003.900
    52003.800
    7.53003.700
    104003.600

    Procedure

    1.Shake well the zinc acetate preserved sample.
    2.add 4 mL sample/standard to each vial
    3.add 320 µL low-range diamine reagent to each sample/standard and shake.
    4.

    let color develop (in a dark place) for 30 minutes. Measure absorbance at 670 nm.

    • If any samples are darker blue than the highest standard, dilute the sample with the dilution reagent until it is within the color range of the standards

    Analyzing Samples


    Preparing the Cary Spectrophotometer

    1. Turn on the unit and let it warm up for at least three hours at the wavelength in question.
    2. In Varian > Cary WinUV start the Advanced Read application. 
    3. Click the Setup button and select the Options tab.
      • Confirm in the Beam Mode area that Double Beam is selected and Normal is also selected.

    SPS3 Autosampler

    The Agilent Technologies SPS3 autosampler facilitates sample introduction into the Cary Spectrophotometer with minimal operator interaction.  


    Image AddedImage Added 

    Figure 1 : Sample inlet tube and flow cell

    The SPS3 is connected to the PC via a RS232 cable to a Keyspan USB converter and the SPS3 is connected to the Cary unit via a RSA sample inlet tube (Figure 1).

    Image Added

    Figure 2 : SPS3 Reservoir peristaltic pump

    The reservoir peristaltic pump (Figure 2) allows nanopure water to be pumped through the system between samples to flush lines. Pump speed is controlled by a dial above the pump. This pump fills the reservoir but does not change the flow into the Cary's flow cell; that is controlled by the spectrometer's peristaltic pump.  

    Image Added 

    Figure 3 : SPS3 Sample trays

    Sample trays allow up to 180 samples to be run (Figure 3). First vial is a "zero" (nanopure water).

    Software setup

    Advanced Reads setup

    Image Added

    Figure 4 : Advanced Reads main screen


    In the Advanced Reads software, click the Setup button (figure 4).


    Image Added

    Figure 5: Cary (Enter wavelength)

    On the Cary tab (figure 5), the wavelength can be changed to the necessary wavelength for the analysis type. 

    Ave Time (sec) is the length of time over which readings will be taken and averaged.  The default setting is 0.1 seconds; increasing this time is beneficial for improving precision and reducing the effect of noise on the resultant measurement.

    1 second is a good minimum setting for Ave Time. There are diminishing returns for precision as time increases. Setting it higher than 5 seconds provides almost no more benefit at the expense of taking more time.


    Image Added

    Figure 6 : Setup (Entering sample information)

    On the samples tab (figure 6), the samples to be run can be entered.

    Set the number of samples to the total combined amount of samples and standards. Enter sample names, including the standards.

    An alternative is that a text file containing sample names can be imported by clicking import names. A text file with textID for sample names can be filled quickly with a barcode scanner (one sample per line). 


    Image Added

    Figure 7 : Setup (Sipper settings [Cary peristaltic pump])

    On the Acessories2 tab (figure 7), make sure sipper on is checked and internal RSA is selected. Set fill/return to 32 seconds and delay to 8 seconds.


    Image Added

    Figure 8 : Setup (Rinse/Sample trays)

    On the samplers tab (figure 8) make sure use sampler is checked along with rinse. Verify that SPS3 Autosampler is selected and displayed. Click configure for the next window.




    Image Added

    Figure 9 : Setup (Rinse/Rack type)

    Click on a tray area and then select the 5x12 positions rack type from the sample rack type drop down menu.

    To position your samples on the autosampler. Click on zero, then click on the first tube position (top right). Then click on sample and click on the second position. This will then populate the sample rack with the samples that were entered on the sample tab.

    Verify that the display matches the loading of samples on the sample rack.


    Image Added

    Figure 10 : Setup (Comm Setup/RS232 settings)

    Com port settings (figure 10) can be reached by clicking comm setup (figure 9). This is just a reference for the proper com port setup in case the setup is lost or forgotten by the software.

    Running samples

      • Prepare samples according to the protocol outlined in the above sections for the appropriate analyte.  
      • Engage the tubing on the peristaltic pump for both the Spectrophotometer and the Autosampler. Make sure that the waste lines go into a receptacle.
      • Click on START in the main Advanced Reads window (Figure 4).
      • A window will pop-up showing the samples entered in the previous steps (Figure 12). Click OK.
      • Save the data to a file for later viewing (Figure 13).


    Image Added

    Figure 12 : Samples ready for analyses


    Image Added

    Figure 13 : Assigning a filename for the results


      • Observe where the samples need to go (Figure 14). Click OK.

    Image Added

    Figure 14 : How to load the vials in the sample racks


      • View the results in real time (Figure 15).

    Image Added

    Figure 15 : Viewing results in real-time


      • Save the results to a .csv file.
        • Select File > Export report (*.csv)
        • Enter the location/file name.
      • Copy/paste into a calibration spreadsheet for manipulation and subsequent upload via Spreadsheet Uploader.
        • Using the results from the reads of the standards, create a calibration curve from the plot of the concentration vs. absorbance values. Use this equation to extrapolate the sample concentrations from their corresponding absorbance value. This can be done in the same spreadsheet as created above in the Advanced Reads application. This sheet can be loaded into the LIMS Spreadsheet Loader application as outlined in the LIMS Integration section.

    Shutting down the Instrument

    Aspirate approximately eight cycles of nanopure water, release the tubing on the peristaltic pump, turn off power to the unit and exit from the software. Clean any spills that may have occurred.
    Empty the waste container into a gallon size ziploc bag of corn cob absorbent in a fume hood. Label the bag "CHEM LAB WASTE - DO NOT OPEN". Let evaporate for a week and place in burnables bag with other burnables. 

    QAQC


    QA/QC for analysis consists of calibration verification using check standards, blanks and replicate samples.

    QA/QC Types

    Check Standard

    A check standard for each set of analytes is run ~ every fifteen analyses depending on batch size. Check standards consist of a standard from approximately the midpoint of the calibration curve.
    The check standard result is evaluated against the threshold for % variance limits for calibration verification standard against true value:

        • Within ±10%: calibration is verified and sample analysis can continue.
        • Outside of ±10%: check reagent solutions and rerun all samples in the corresponding analytical batch.

    Blank

    A blank is run with every batch to determine if high background levels are interfering with accurate sample results.

    Replicate Samples

    During each batch, a single sample should be run in duplicate and the variation of the results compared.

    Data Available in LORE


    Interstitial Waters Standard Report 

    • Exp: expedition number
    • Site: site number
    • Hole: hole number
    • Core: core number
    • Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
    • Sect: section number
    • A/W: archive (A) or working (W) section half.
    • Top offset on section (cm): position of the upper edge of the sample, measured relative to the top of the section.
    • Bottom offset on section (cm): position of the lower edge of the sample, measured relative to the top of the section.
    • Top depth CSF-A (m): position of observation expressed relative to the top of the hole.
    • Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
    • Sampling tool: tool used to collect sample
    • Data columns: header lists parameter measured and concentration units, followed by wavelength (for ICP-AES) and then analysis method.

    Expanded IC Report

    • Exp: expedition number
    • Site: site number
    • Hole: hole number
    • Core: core number
    • Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
    • Sect: section number
    • A/W: archive (A) or working (W) section half.
    • text_id: automatically generated unique database identifier for a sample, visible on printed labels
    • sample_number: sample number of sample. text ID with sample type prefix removed.
    • label_id: id combining exp, site, hole, core, type, sect, A/W, parent sample name (if any), sample name
    • sample_name: name of sample
    • x_sample_state:
    • x_project: expedition project the sample is uploaded under. typically the same as Exp.
    • x_capt_loc: 
    • location: location sample was taken
    • x_sampling_tool: tool used to collect sample
    • changed_by: name of person who uploaded sample
    • changed_on: date and time sample was uploaded
    • sample_type: type of sample. typically LIQ, for liquid.
    • x_offset: top offset of parent sample where sample was taken in m
    • x_offset_cm: top offset of parent sample where sample was taken in cm
    • x_bottom_offset_cm: bottom offset of parent sample where sample was taken in cm
    • x_diameter: 
    • x_idmp: 
    • x_orig_len: 
    • x_length: length of sample in m
    • x_lengeth_cm:  length of sample in cm
    • status:
    • old_status:
    • original_sample: 
    • parent_sample:
    • standard:
    • login_by: name of person logged into LIMS application used for this test
    • sampled_date: 
    • legacy:
    • test changed_on: date of last edit of analysis
    • test date_started: date analysis was started
    • test group_name: 
    • test status:
    • test old_status:
    • test test_number: unique number associated with the instrument measurement steps that produced these data
    • test date_received: date analysis was uploaded to LIMS
    • test instrument: instrument used to perform analysis
    • test analysis: analysis type
    • test x_project: project test was assigned to
    • test version: 
    • test order_number:
    • test replicate_test:
    • test replicate_count:
    • rest sample_number: sample number for sample the analysis was performed on
    • Top depth CSF-A (m): position of observation expressed relative to the top of the hole.

    • Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole.

    • Top depth CSF-B (m):

    • Bottom depth CSF-B (m):

    • analyte: the analyte measured for this test
    • concentration (uM):  concentration of analyte in uM
    • ssup_assman_id:  link to download the batch of data uploaded with spreadsheet uploader
    • ssup_filename: filename of the batch of data uploaded with spreadsheet uploader
    • sample description: observations recorded about the sample itself

    • test test_comment: observations about a measurement or the measurement process; some measurement observations may be under Result comments

    • result comments: observations about a measurement or the measurement process; some measurement observations may be under Test Comments


    Data collected is transferred to the LIMS database using IODP's Spreadsheet Loader application.

    This is best done by entering the results into an Excel spreadsheet in a format similar to the pattern below, keeping the appropriate number of columns blank, and omitting the column headers.

    Then run the Spreadsheet Loader application. Go to File > Load and it should import something like below (Figure 16).

    Before uploading, click Edit > Validate Sheet to validate the data.  

    To upload into the database, go to  Lims > Upload and status messages will appear in the blank window.


    Image Added

    Figure 16 : Uploading results with Spreadsheet Uploader

    Maintenance and Troubleshooting


    Cleaning

    Any spills in the sample compartment should be immediately wiped up and any deposits on the sample compartment windows should also be removed. The exterior surfaces should be cleaned with a soft cloth and, if necessary, this cloth can be dampened with water or a mild detergent.

    Source Lamps

    Instructions for how to change and align both the visible and UV lamps are included in the Help provided with the software. Care must be taken when removing lamps. Touching the glass envelope will reduce its efficiency. NEVER touch the glass surface of a new lamp. Always handle a lamp by its base, using a soft cloth.

    Fuses

    To replace a fuse, disconnect the unit from the power supply, and replace the blown fuse with one of the type and rating as outlined in the hardware specifications section.

    1. Disconnect the instrument from the power supply.
    2. Undo the fuse cap by pressing the cap and turning it counter-clockwise.
    3. Carefully pull out the cap. The fuse should be held in the fuse cap.
    4. Check that the fuse is the correct type and is not damaged. If necessary, replace it.
    5. Place the fuse into the cap, push the cap in and then turn the cap clockwise.
    6. Reconnect the instrument to the power supply.

    Cary Win UV Help

    Varian provides extensive help resources, available from the software CD. After installing the help utilities, go to START > All Programs > Varian > Cary WinUV > Cary Help.
    Cary Help offers troubleshooting information, maintenance information like how to replace lamps, aligning lamps and mirrors, and replacing fuses and cleaning the flowcells.

    Contact Information

    Varian Instruments

    1 800 926 3000

    customer.service@varianinc.com

    References


    Gieskes, J.M., Gamo, T., and Brumsack, H., 1991. Chemical methods for interstitial water analysis aboard JOIDES Resolution, ODP Tech. Note, 15. doi:10.2973/odp.tn.15.1991.

    Kastner, M., Elderfield, H., Martin, J.B., Suess, E., Kvenvolden, K.A., and Garrison, R.E., 1990. Diagenesis and interstitial water chemistry at the Peruvian margin—major constituents and strontium isotopes. In Suess, E., von Huene, R., et al., Proc. ODP, Sci. Results, 112: College Station, TX (Ocean Drilling Program), 413–440. doi:10.2973/odp.proc.sr.112.144.1990

    Noriki, S. 1983. Silicate correction in the colorimetric determination of phosphate in seawater. J. Oceanograph. Soc. Japan, 39(6):324–326. doi:10.1007/BF02071829

    Presley, B.J., 1971. Techniques for analyzing interstitial water samples: Appendix Part 1: determination of selected minor and major inorganic constituents. In Winterer, E.L., et al., Init. Repts. DSDP, 7(2): Washington, DC (U.S. Govt. Printing Office), 1749–1755. doi:10.2973/dsdp.proc.7.app1.1971

    Solorzano, L., 1969. Determination of ammonia in natural waters by phenol-hypochlorite method. Limno. Oceanogr., 14:799–801.

    Strickland, J.D.H., and Parsons, T.R., 1968. A manual for sea water analysis. Bull. Fish. Res. Board Can., 167.

    Appendix


    Hardware

    The Varian Cary 100 is a double-beam, dual-chopper, monochromator UV-Vis spectrophotometer, centrally controlled by a PC. It has a high-performance R928 photomultiplier tube, tungsten halogen visible source with quartz window, and deuterium arc ultraviolet source. 


    Name

    Agilent Technologies Cary UV-Vis Spectrophotometer

    Model

    Cary-100

    Serial number

    UV1110M021

    Dimensions

    26 × 26 × 13 in (unpacked)

    Weight

    99 lb (unpacked)

    Monochromator

    Czerny-Turner 0.28 m

    Grating

    30 × 35 mm, 1200 lines/mm, blaze angle 8.6° at 240 nm

    Beam Splitting System

    Chopper (30 Hz)

    Detectors

    R928 PMT

    UV-Vis Limiting Resolution (nm)

    0.189

    Wavelength Range (nm)

    190–900

    Wavelength Accuracy (nm)

    0.02 at 656.1 nm; 0.04 at 486.0 nm

    Wavelength Reproducibility (nm)

    0.008

    Signal Averaging (s)

    0.033–999

    Spectral Bandwidth (nm)

    0.20–4.00 nm, 0.1 nm steps, motor driven

    Spectral Bandwidth Accuracy (nm)

    @ 0.2: 0.193; @ 2.0: 2.03

    Photometric Accuracy (Abs)

    @ 0.3 Abs (Double Aperture method): 0.00016

    Photometric Range (Abs)

    3.7

    Photometric Display

    (Abs) ± 9.9999; (%T) ± 200.00

    Photometric Reproducibility

    (Abs; NIST 930D filters)

    2 s signal averaging time @ 590 nm, 2 nm SBW
    Maximum deviation at 1 Abs
    Std dev for 10 measurements


    <0.0008
    <0.00016

    2 s signal averaging time @ 546.1 nm, 2 nm SBW
    Maximum deviation at 0.5 Abs
    Std dev for 10 measurements


    <0.0004
    <0.00008

    Photometric Stability (Abs/hr)
    500 nm, 1 s signal averaging time

    2 h warmup
    <0.0003

    Photometric Noise (Abs, RMS)
    500 nm, 1 s signal averaging time

    2 nm SBW
    0.000030 at 0 Abs; 0.00014 at 3 Abs, 1.5 Abs RBA

    Baseline Flatness (Abs)
    200–850 nm, smoothing = 21, baseline corrected

    0.00022

    Sample Compartment Beam Separation (mm)

    110




    Image Added


    Figure 17 : Schematic of Cary Spectrophotometer

    Electrical

    Power supply (VAC)

    100, 120, 220, or 240 ± 10%

    Frequency (Hz)

    50 or 60 ± 1 with 400 VA power consumption

    Fuses (100–120 VAC)

    T5 AH 250 V, IEC 127 sheet 5, 5 × 20 mm ceramic

    COM port (rear)

    IEEE 488

    PC port

    USB


    Replacement Parts

    Item

    Part number

    Instrument fuse, 5 A time lag, ceramic, M205

    1910009100

    Peristaltic pump tubing replacement kit

    9910052900

    Visible source lamp

    5610021700

    Deuterium lamp

    5610021800

    Dissolution cell, 715 µL, 10 mm

    6610015200

    Thumbscrew kit

    9910064100

    Spares kit: accessory locating pin, accessory fastening screws, instrument feet, instrument cover snap cap washer, snap cap, ACB cover plate, socket covers for ACB

    9910064300

    Pumps

    A double-action peristaltic pump services the feed and waste. 
    Image Added
    (1) plastic sleeve. (2) metal hook tubing. (3) waste outlet.

    Credits

    This document originated from Word documents Cary_UG_374_draft and CarySPS3_QSG_374_draft that were written by E. Moortgat (20111212). Credits for subsequent changes to this document are given in the page history.

    LIMS Component Table

    ANALYSISTABLENAMEABOUT TEXT
    SPECSAMPLEExpExp: expedition number
    SPECSAMPLESiteSite: site number
    SPECSAMPLEHoleHole: hole number
    SPECSAMPLECoreCore: core number
    SPECSAMPLETypeType: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
    SPECSAMPLESectSect: section number
    SPECSAMPLEA/WA/W: archive (A) or working (W) section half.
    SPECSAMPLEtext_idText_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples.
    SPECSAMPLEsample_numberSample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table.
    SPECSAMPLElabel_idLabel identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples.
    SPECSAMPLEsample_nameSample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter.
    SPECSAMPLEx_sample_stateSample state: Single-character identifier always set to "W" for samples; standards can vary.
    SPECSAMPLEx_projectProject: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
    SPECSAMPLEx_capt_locCaptured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
    SPECSAMPLElocationLocation: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
    SPECSAMPLEx_sampling_toolSampling tool: sampling tool used to take the sample (e.g., syringe, spatula)
    SPECSAMPLEchanged_byChanged by: username of account used to make a change to a sample record
    SPECSAMPLEchanged_onChanged on: date/time stamp for change made to a sample record
    SPECSAMPLEsample_typeSample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ)
    SPECSAMPLEx_offsetOffset (m): top offset of sample from top of parent sample, expressed in meters.
    SPECSAMPLEx_offset_cmOffset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm)
    SPECSAMPLEx_bottom_offset_cmBottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm)
    SPECSAMPLEx_diameterDiameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master
    SPECSAMPLEx_orig_lenOriginal length (m): field for the original length of a sample; not always (or reliably) populated
    SPECSAMPLEx_lengthLength (m): field for the length of a sample [as entered upon creation]
    SPECSAMPLEx_length_cmLength (cm): field for the length of a sample. This is a calculated field (length, converted to cm).
    SPECSAMPLEstatusStatus: single-character code for the current status of a sample (e.g., active, canceled)
    SPECSAMPLEold_statusOld status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample
    SPECSAMPLEoriginal_sampleOriginal sample: field tying a sample below the CORE level to its parent HOLE sample
    SPECSAMPLEparent_sampleParent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR)
    SPECSAMPLEstandardStandard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T)
    SPECSAMPLElogin_byLogin by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created])
    SPECSAMPLElogin_dateLogin date: creation date of the sample
    SPECSAMPLElegacyLegacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition
    SPECTESTtest changed_onTEST changed on: date/time stamp for a change to a test record.
    SPECTESTtest statusTEST status: single-character code for the current status of a test (e.g., active, in process, canceled)
    SPECTESTtest old_statusTEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test
    SPECTESTtest test_numberTEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table.
    SPECTESTtest date_receivedTEST date received: date/time stamp for the creation of the test record.
    SPECTESTtest instrumentTEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input)
    SPECTESTtest analysisTEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table)
    SPECTESTtest x_projectTEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
    SPECTESTtest sample_numberTEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table
    SPECCALCULATEDTop depth CSF-A (m)Top depth CSF-A (m): position of observation expressed relative to the top of the hole.
    SPECCALCULATEDBottom depth CSF-A (m)Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole.
    SPECCALCULATEDTop depth CSF-B (m)Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
    SPECCALCULATEDBottom depth CSF-B (m)Bottom depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
    SPECRESULTanalyteRESULT analyte: analyte name of the property being measured (e.g., phosphate, ammonium)
    SPECRESULTconcentration (µM)RESULT concentration (uM): concentration of the analyte
    SPECRESULTssup_asman_idRESULT spreadsheet uploader ASMAN_ID: serial number for the ASMAN link for the spreadsheet uploader file
    SPECRESULTssup_filenameRESULT spreadsheet uploader filename: file name for the spreadsheet uploader file
    SPECSAMPLEsample descriptionSAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments"
    SPECTESTtest test_commentTEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments"
    SPECRESULTresult commentsRESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments"
    • High range: 6-80 uM
    • Low range: 1-10 uM.

    All samples need to be preserved during splitting with a 1% zinc acetate solution. It can require a relatively large amount of sample in order to do this analysis. The high range method uses less sample (~500 µL), but if the sulfide level is below 6 µM, it won't be enough. It may still be worth screening samples with the high-range method in order to conserve interstitial water sample volume. The low-range method requires 4 mL (8x the amount of sample) but is sensitive down to 1 µM. The scientists will have to determine if consumption of 4 mL, possibly 4.5 mL, of sample is worth obtaining the sulfide concentration.

    Reagent Solutions

    A: 1% zinc acetate (w/v)

    For Standard: Prepare 1 L of 1% zinc acetate by dissolving 10 g zinc acetate dehydrate into about 600 mL DI water in a 1 L volumetric flask. Add 1 mL of concentrated acetic acid, bring up to volume, and mix well.
    For Splits: Prepare 100 mL of 1% zinc acetate by dissolving 1.0 g zinc acetate dehydrate into about 60 mL DI water in a 100 mL volumetric flask. Add 100 µL of concentrated acetic acid, bring up to volume, and mix well. 

    B: 1 mM zinc-sulfide standard suspension

    Before weighing the sodium sulfide nonahydrate, wash the crystals off with nanopure water and dry with a clean Kim Wipe to remove the oxidation products from the crystal before weighing. Try to select very small crystals (or break them up) to facilitate accurate weighing. Dissolve 0.20 g of Na2S·9H2O in 1 L of 1% zinc acetate solution (Reagent A). Store in plastic and cool. Shake vigorously each time before using.

    Note: It is very difficult to weigh 0.20 g of this reagent accurately because of the size of the crystals. Weigh 0.20 g within 5 mg and then record the mass used on the bottle. The Baltic Gas methods paper (Ferdelman et al., 2011) describes the use of sodium thiosulfate to standardize the Zn-S suspension, but this has not been the standard practice.

    C: 100 µM zinc-sulfide standard

    Dilute 10 mL of 1 mM zinc-sulfide standard suspension (Reagent B) up to 100 mL with nanopure water. This is the stock standard for the dilution of the standard curve.

    D: High-Range Sulfide Diamine Reagent

    For the high-range (6-80 µM) method, add 500 mL of concentrated HCl to 500 mL of nanopure water. Mix thoroughly and allow to cool to room temperature. Add 4.0 g of N,N-dimethyl-p-phenylenediamine sulfate and 6.0 g iron chloride hexahydrate (FeCl3·6H2O), mix well, and refrigerate in an amber Nalgene bottle.

    E: Low-Range Sulfide Diamine Reagent

    For the low-range (1-10 µM) method, add 250 mL of concentrated HCl to 250 mL of nanopure water. Mix thoroughly and allow to cool to room temperature. Add 0.50 g of N,N-dimethyl-p-phenylenediamine sulfate and 0.75 g iron chloride hexahydrate (FeCl3·6H2O), mix well, and refrigerate in an amber Nalgene bottle.

    F: Dilution Reagent

    This reagent is used to dilute too-dark samples into the range of color covered by the standard curve, while keeping a constant concentration of the diamine reagent. Create this reagent as appropriate for the high-range or low-range method:

    For the high-range method, add 800 µL of Reagent D to 10 mL of nanopure water. 
    For the low-range method, add 4 mL of Reagent E to 50 mL of nanopure water.

    Sample Preservation

    The IW splits for sulfide should be preserved immediately with the 1% zinc acetate solution (Reagent A) to prevent the loss of dissolved sulfide.

      • For 0.5 mL aliquots (high range method), fix the sulfide with 40 µL of 1% zinc acetate. 
      • For 4.0 mL aliquots (low-range method), fix the sulfide with 800 µL of 1% zinc acetate solution.

    Procedure

    High-Range Method

    Make dilutions of the 100 µM zinc-sulfide standard (Reagent C) for the standards to create 0.5 mL of standard at each level.

    standard level

    reagent C ( µL)

    nanopure water (µL)

    concentration (µM)

    blank

    0

    500

    0.0

    6

    30

    470

    6.0

    20

    100

    400

    20

    40

    200

    300

    40

    60

    300

    200

    60

    80

    400

    100

    80

    1. Well shake the zinc acetate-preserved sample
    2. Pipette the sample (0.5 mL) into the vial.
    3. Add 40 µL of the high-range diamine reagent (Reagent D) to the sample, blank, and standards and shake each vial. 
    4. Place the samples, blank, and standards, into the dark for 30 minutes. The standards should turn blue. If the visual appearance of any of the samples is darker blue than the highest standard, dilute the sample in the dilution reagent (Reagent F) until it is within the color range of the standards.

    Low-Range Method

    Make dilutions of the 100 µM zinc-sulfide standard (Reagent C) for the standards to create 4 mL of standard at each level. Add the specified volume of Reagent C and nanopure water as shown in the table below to create the standards.

    concentration (µM)

    reagent C (µL)

    nanopure water (µL)

    0

    0

    4000

    1

    40

    3960

    2.5

    100

    3900

    5

    200

    3800

    7.5

    300

    3700

    10

    400

    3600

    1. Well shake the zinc acetate-preserved sample. 
    2. Pipette the sample (4 mL) into the vial. 
    3. Add 320 µL of the low-range diamine reagent (Reagent E) to the sample, blank, and standards, and shake each vial. 
    4. Place the samples, blank, and standards into the dark for 30 minutes. The standards should turn blue. If the visual appearance of any of the samples is darker blue than the highest standard, dilute the sample in the dilution reagent (Reagent F) until it is within the color range of the standards.
    Note: From the Exp. 366 Methods chapter

    Hydrogen sulfide (expected to be HS- at the pH of most IW samples) concentrations were analyzed following the method of Cline (1969) with modifications as adapted for small volumes of pore fluids by Ferdelman et al., 2011. Initially, 500 µL sample was fixed with 40 µL of a 1% zinc acetate solution, and reagents used were for a range of 6 to 80 µM. However, because most samples were below detection for the first set of samples, 4 ml sample was fixed with 800 µL zinc acetate solution, and the analyses were conducted following the lowest range (1 to 3 µM) outlined in Cline (1969), which had a linear range for the calibration curve at least up to 10 µM. The zinc acetate fixed sample was vigorously shaken and 320 µL of a diamine solution consisting of 0.5 g N,N-dimethyl-p-phenylenediamine sulfate and 0.75 g ferric chloride (FeCl3 * 6H2O) per 500 ml nanopure water, was added. The solution was shaken, and left for 30 minutes in the dark, then measured by spectrophotometry at 670 nm. If the blue color of the sample was visually darker than that of the highest calibration standard (10 µM), the sample was diluted with nanopure water until a lighter color was observed.

    Analyzing Samples

    Preparing the Cary Spectrophotometer

    1. Engage the tubing on the peristaltic pump. Make sure that the waste line goes into a receptacle.
    2. Turn on the unit and let it warm up for at least three hours at the wavelength in question.
    3. Confirm that the instrument Beam Mode is correctly configured. 
      1. In Varian > Cary WinUV start the Advanced Read application. 
      2. Click the Setup button and select the Options tab. 
        • Confirm in the Beam Mode area that Double Beam is selected and Normal is also selected.

    SPS3 Autosampler

    The Agilent Technologies SPS3 autosampler facilitates sample introduction into the Cary Spectrophotometer with minimal operator interaction. 

     
    The SPS3 is connected to the PC via a RS232 cable to a Keyspan USB converter and the SPS3 is connected to the Cary unit via a RSA sample inlet tube (Figure 1).

    Image RemovedImage Removed 

    Figure 1 : Sample inlet tube and flow cell

    The reservoir peristaltic pump (Figure 2) allows nanopure water to be pumped through the system between samples to flush lines. Pump speed is controlled by a dial above the pump. This pump fills the reservoir but does not change the flow into the Cary's flow cell; that is controlled by the spectrometer's peristaltic pump.  

    Image Removed

    Figure 2 : SPS3 Reservoir peristaltic pump

    Sample trays allow up to 180 samples to be run (Figure 3). First vial is a "zero" (nanopure water).

    Image Removed 

    Figure 3 : SPS3 Sample trays

    Software setup

    Advanced Reads setup

    Image Removed

    Figure 4 : Advanced Reads main screen

    Image Removed

    Figure 5 : Setup (Entering sample information)

    Image Removed

    Figure 6 : Setup (Sipper settings [Cary peristaltic pump])

    Image Removed

    Figure 8 : Setup (Rinse/Sample trays)

    Image Removed

    Figure 9 : Setup (Rinse/Rack type)

    Image Removed

    Figure 10 : Setup (Comm Setup/RS232 settings)

    Click on Custom Racks (Figure 9) to configure sample racks.

    Image Removed

    Figure 11 : Setup (Custom Sample Rack Editor)

    Running samples

    1. Prepare samples according to the protocol outlined in the above sections for the appropriate analyte. 
    2. Click on START in the main Advanced Reads window (Figure 4). 
    3. A window will pop-up showing the samples entered in the previous steps (Figure 12). Click OK.
    4. Save the data to a file for later viewing (Figure 13).

    Image Removed

    Figure 12 : Samples ready for analyses

    Image Removed

    Figure 13 : Assigning a filename for the results

    5. Observe where the samples need to go (Figure 14). Click OK.

    Image Removed

    Figure 14 : How to load the vials in the sample racks

    6. View the results in real time (Figure 15).

    Image Removed

    Figure 15 : Viewing results in real-time

    7. Save the results to a .csv file.

      • Select File → Export report (*.csv)
      • Enter the location/file name.

    8. Copy/paste into a calibration spreadsheet for manipulation and subsequent upload via Spreadsheet Uploader.

    Using the results from the reads of the standards, create a calibration curve from the plot of the concentration vs. absorbance values. Use this equation to extrapolate the sample concentrations from their corresponding absorbance value. This can be done in the same spreadsheet as created above in the Advanced Reads application. This sheet can be loaded into the LIMS Spreadsheet Loader application as outlined in the LIMS Integration section.

    Shutting down the Instrument

    Aspirate approximately eight cycles of nanopure water, release the tubing on the peristaltic pump, turn off power to the unit and exit from the software. Clean any spills that may have occurred.
    Empty the waste container and rinse with tap water.

    QAQC

    QA/QC for analysis consists of calibration verification using check standards, blanks and replicate samples.

    QA/QC Types

    Check Standard

    A check standard for each set of analytes is run ~ every fifteen analyses depending on batch size. Check standards consist of a standard from approximately the midpoint of the calibration curve.
    The check standard result is evaluated against the threshold for % variance limits for calibration verification standard against true value:

      • Within ±10%: calibration is verified and sample analysis can continue.
      • Outside of ±10%: check reagent solutions and rerun all samples in the corresponding analytical batch.

    Blank

    A blank is run with every batch to determine if high background levels are interfering with accurate sample results.

    Replicate Samples

    During each batch, a single sample should be run in duplicate and the variation of the results compared.

    LIMS Integration

    Results are stored in the LIMS database associated with an analysis code and an analysis component.

    Analysis

    Component Name

    Component value

    Unit

    SPEC

    analyte

    ammonium

    NONE

    phosphate

     NONE

    silica

     NONE

    concentration

    concentration of analyte

    µM

    Data collected is transferred to the LIMS database using IODP's Spreadsheet Loader application.

    This is best done by entering the results into an Excel spreadsheet in a format similar to the pattern below, keeping the appropriate number of columns blank, and omitting the column headers.

    Then run the Spreadsheet Loader application. Go to File > Load and it should import something like below (Figure 16).

    To upload into the database, go to  Lims > Upload and status messages will appear in the blank window.

    Image RemovedFigure 16 : Uploading results with Spreadsheet Uploader

    Maintenance and Troubleshooting

    Cleaning

    Any spills in the sample compartment should be immediately wiped up and any deposits on the sample compartment windows should also be removed. The exterior surfaces should be cleaned with a soft cloth and, if necessary, this cloth can be dampened with water or a mild detergent.

    Source Lamps

    Instructions for how to change and align both the visible and UV lamps are included in the Help provided with the software. Care must be taken when removing lamps. Touching the glass envelope will reduce its efficiency. NEVER touch the glass surface of a new lamp. Always handle a lamp by its base, using a soft cloth.

    Fuses

    To replace a fuse, disconnect the unit from the power supply, and replace the blown fuse with one of the type and rating as outlined in the hardware specifications section.

    1. Disconnect the instrument from the power supply.
    2. Undo the fuse cap by pressing the cap and turning it counter-clockwise.
    3. Carefully pull out the cap. The fuse should be held in the fuse cap.
    4. Check that the fuse is the correct type and is not damaged. If necessary, replace it.
    5. Place the fuse into the cap, push the cap in and then turn the cap clockwise.
    6. Reconnect the instrument to the power supply.

    Cary Win UV Help

    Varian provides extensive help resources, available from the software CD. After installing the help utilities, go to START > All Programs > Varian > Cary WinUV > Cary Help.
    Cary Help offers troubleshooting information, maintenance information like how to replace lamps, aligning lamps and mirrors, and replacing fuses and cleaning the flowcells.

    Contact Information

    Varian Instruments

    1 800 926 3000

    customer.service@varianinc.com

    References

    Gieskes, J.M., Gamo, T., and Brumsack, H., 1991. Chemical methods for interstitial water analysis aboard JOIDES Resolution, ODP Tech. Note, 15. doi:10.2973/odp.tn.15.1991.

    Kastner, M., Elderfield, H., Martin, J.B., Suess, E., Kvenvolden, K.A., and Garrison, R.E., 1990. Diagenesis and interstitial water chemistry at the Peruvian margin—major constituents and strontium isotopes. In Suess, E., von Huene, R., et al., Proc. ODP, Sci. Results, 112: College Station, TX (Ocean Drilling Program), 413–440. doi:10.2973/odp.proc.sr.112.144.1990

    Noriki, S. 1983. Silicate correction in the colorimetric determination of phosphate in seawater. J. Oceanograph. Soc. Japan, 39(6):324–326. doi:10.1007/BF02071829

    Presley, B.J., 1971. Techniques for analyzing interstitial water samples: Appendix Part 1: determination of selected minor and major inorganic constituents. In Winterer, E.L., et al., Init. Repts. DSDP, 7(2): Washington, DC (U.S. Govt. Printing Office), 1749–1755. doi:10.2973/dsdp.proc.7.app1.1971

    Solorzano, L., 1969. Determination of ammonia in natural waters by phenol-hypochlorite method. Limno. Oceanogr., 14:799–801.

    Strickland, J.D.H., and Parsons, T.R., 1968. A manual for sea water analysis. Bull. Fish. Res. Board Can., 167.

    Appendix

    Hardware

    The Varian Cary 100 is a double-beam, dual-chopper, monochromator UV-Vis spectrophotometer, centrally controlled by a PC. It has a high-performance R928 photomultiplier tube, tungsten halogen visible source with quartz window, and deuterium arc ultraviolet source. 

    Name

    Agilent Technologies Cary UV-Vis Spectrophotometer

    Model

    Cary-100

    Serial number

    UV1110M021

    Dimensions

    26 × 26 × 13 in (unpacked)

    Weight

    99 lb (unpacked)

    Monochromator

    Czerny-Turner 0.28 m

    Grating

    30 × 35 mm, 1200 lines/mm, blaze angle 8.6° at 240 nm

    Beam Splitting System

    Chopper (30 Hz)

    Detectors

    R928 PMT

    UV-Vis Limiting Resolution (nm)

    0.189

    Wavelength Range (nm)

    190–900

    Wavelength Accuracy (nm)

    0.02 at 656.1 nm; 0.04 at 486.0 nm

    Wavelength Reproducibility (nm)

    0.008

    Signal Averaging (s)

    0.033–999

    Spectral Bandwidth (nm)

    0.20–4.00 nm, 0.1 nm steps, motor driven

    Spectral Bandwidth Accuracy (nm)

    @ 0.2: 0.193; @ 2.0: 2.03

    Photometric Accuracy (Abs)

    @ 0.3 Abs (Double Aperture method): 0.00016

    Photometric Range (Abs)

    3.7

    Photometric Display

    (Abs) ± 9.9999; (%T) ± 200.00

    Photometric Reproducibility

    (Abs; NIST 930D filters)

    2 s signal averaging time @ 590 nm, 2 nm SBW
    Maximum deviation at 1 Abs
    Std dev for 10 measurements

    <0.0008
    <0.00016

    2 s signal averaging time @ 546.1 nm, 2 nm SBW
    Maximum deviation at 0.5 Abs
    Std dev for 10 measurements

    <0.0004
    <0.00008

    Photometric Stability (Abs/hr)
    500 nm, 1 s signal averaging time

    2 h warmup
    <0.0003

    Photometric Noise (Abs, RMS)
    500 nm, 1 s signal averaging time

    2 nm SBW
    0.000030 at 0 Abs; 0.00014 at 3 Abs, 1.5 Abs RBA

    Baseline Flatness (Abs)
    200–850 nm, smoothing = 21, baseline corrected

    0.00022

    Sample Compartment Beam Separation (mm)

    110

    Image RemovedFigure 17 : Schematic of Cary Spectrophotometer

    Electrical

    Power supply (VAC)

    100, 120, 220, or 240 ± 10%

    Frequency (Hz)

    50 or 60 ± 1 with 400 VA power consumption

    Fuses (100–120 VAC)

    T5 AH 250 V, IEC 127 sheet 5, 5 × 20 mm ceramic

    COM port (rear)

    IEEE 488

    PC port

    USB

    Replacement Parts

    Item

    Part number

    Instrument fuse, 5 A time lag, ceramic, M205

    1910009100

    Peristaltic pump tubing replacement kit

    9910052900

    Visible source lamp

    5610021700

    Deuterium lamp

    5610021800

    Dissolution cell, 715 µL, 10 mm

    6610015200

    Thumbscrew kit

    9910064100

    Spares kit: accessory locating pin, accessory fastening screws, instrument feet, instrument cover snap cap washer, snap cap, ACB cover plate, socket covers for ACB

    9910064300

    Pumps

    A double-action peristaltic pump services the feed and waste. 
    Image Removed
    (1) plastic sleeve. (2) metal hook tubing. (3) waste outlet.

    Credits

    This document originated from Word documents Cary_UG_374_draft and CarySPS3_QSG_374_draft that were written by E. Moortgat (20111212). Credits for subsequent changes to this document are given in the page history.