Coulometer: Quick Start Guide

Table of Contents
maxLevel3

Manual Information


Author(s):

C. Bennight

Reviewer(s):

D. Houpt, K. Fujine, L. Brandt

Revised by:

D. Houpt

Manager Approval (Name, Title, Date):

D.J. Houpt (Supervisor of Analytical Services)

Audience:

Scientists, Laboratory Technicians

Origination date:

3/1/2009

Current version:

V371T

July 2017

Previous Revisions:

V1.2

2/20/2014, V 1.0

9/08/2010

Domain:

Chemistry

System:

Coulometry

...

Motion compensation software developed in house allows the user to weigh the mass of chemicals and samples at sea. Reagents and samples >250 mg must be measured on the Mettler-Toledo XS204 balance (Figure 5). Reagents and samples ≤ 250 mg must be measured on the Cahn balance (Figure 6).


Figure 5. Mettler-Toledo Dual Balance Control Software.

Figure 6. Cahn Balance Control Software.

Laboratory Supplies

Apparatus

...





Figure 9. Coulometer Software

  1. Click Measure (Figure 9).
  2. Quickly add 5mL of 2N HCl to the sample using the acid dispenser (within 10 seconds).
  3. Once measuring is complete, click Save and record the %CaCO3, % carbon, and carbon mass in the log book. The %CaCO3 standard should give a %CaCO3 value of 100%. See the technician if this value is not acquired after running 3 standards.
  4. Wait 10-15 minutes for the system to restabilize between samples.
  5. Repeat steps 10-16 for the samples.

...