Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Cary Spectrophotometer: User Guide


Manual Information




Author(s)

E. Moortgat

Revision

E. MoortgatD. Houpt

Reviewer(s)

C. Bennight, M. Bertoli, L. Brandt

Management Approval

D. Houpt, Supervisor of Analytical Systems

Audience

Laboratory Technicians and Scientists

Origination date

12/12/2011

Current version

Version 1.# (12/17/2015)V3741T

FebJuly 20187

Domain

Chemistry (IW)

Analysis

Spectrophotometry

...




...

Introduction


The principles of spectroscopic analysis rely on Beer's law. The principle of Beer's law is that passing light of a known wavelength through a sample of known thickness and measuring how much of the light is absorbed at that wavelength will provide the concentration of the unknown, provided that the unknown is in a complex that absorbs significantly at the chosen wavelength.
Beer's law, expressed for liquids, can be represented as an equation, where A = absorbance, k = an experimentally determined constant, b = path length, and c = concentration:
A = kbc.
Thus, concentration can be determined.
IODP's Agilent Cary 100 double-beam UV-Vis (ultraviolet–visible) spectrophotometer is ideal for shipboard routine and research laboratory work. The system measures analytes in interstitial water obtained from sediment cores using standard colorimetric methodology.
Anchor_Ref307906522_Ref307906522


Hardware and Materials


The Varian Cary 100 is a double-beam, dual-chopper, monochromator UV-Vis spectrophotometer, centrally controlled by a PC. It has a high-performance R928 photomultiplier tube, tungsten halogen visible source with quartz window, and deuterium arc ultraviolet source (Figure 1). More details can be found below.


Name

Agilent Technologies Cary UV-Vis Spectrophotometer

Model

Cary-100

Serial number

UV1110M021

Dimensions

26 × 26 × 13 in (unpacked)

Weight

99 lb (unpacked)

Monochromator

Czerny-Turner 0.28 m

Grating

30 × 35 mm, 1200 lines/mm, blaze angle 8.6° at 240 nm

Beam Splitting System

Chopper (30 Hz)

Detectors

R928 PMT

UV-Vis Limiting Resolution (nm)

0.189

Wavelength Range (nm)

190–900

Wavelength Accuracy (nm)

0.02 at 656.1 nm; 0.04 at 486.0 nm

Wavelength Reproducibility (nm)

0.008

Signal Averaging (s)

0.033–999

Spectral Bandwidth (nm)

0.20–4.00 nm, 0.1 nm steps, motor driven

Spectral Bandwidth Accuracy (nm)

@ 0.2: 0.193; @ 2.0: 2.03

Photometric Accuracy (Abs)

@ 0.3 Abs (Double Aperture method): 0.00016

Photometric Range (Abs)

3.7

Photometric Display

(Abs) ± 9.9999; (%T) ± 200.00

Photometric Reproducibility

(Abs; NIST 930D filters)

2 s signal averaging time @ 590 nm, 2 nm SBW
Maximum deviation at 1 Abs
Std dev for 10 measurements


<0.0008
<0.00016

2 s signal averaging time @ 546.1 nm, 2 nm SBW
Maximum deviation at 0.5 Abs
Std dev for 10 measurements


<0.0004
<0.00008

Photometric Stability (Abs/hr)
500 nm, 1 s signal averaging time

2 h warmup
<0.0003

Photometric Noise (Abs, RMS)
500 nm, 1 s signal averaging time

2 nm SBW
0.000030 at 0 Abs; 0.00014 at 3 Abs, 1.5 Abs RBA

Baseline Flatness (Abs)
200–850 nm, smoothing = 21, baseline corrected

0.00022

Sample Compartment Beam Separation (mm)

110




Image Modified
Figure 1. Schematic of Cary Spectrophotometer.


Electrical


Power supply (VAC)

100, 120, 220, or 240 ± 10%

Frequency (Hz)

50 or 60 ± 1 with 400 VA power consumption

Fuses (100–120 VAC)

T5 AH 250 V, IEC 127 sheet 5, 5 × 20 mm ceramic

COM port (rear)

IEEE 488

PC port

USB






Replacement Parts


Item

Part number

Instrument fuse, 5 A time lag, ceramic, M205

1910009100

Peristaltic pump tubing replacement kit

9910052900

Visible source lamp

5610021700

Deuterium lamp

5610021800

Dissolution cell, 715 µL, 10 mm

6610015200

Thumbscrew kit

9910064100

Spares kit: accessory locating pin, accessory fastening screws, instrument feet, instrument cover snap cap washer, snap cap, ACB cover plate, socket covers for ACB

9910064300




Pumps


A double-action peristaltic pump services the feed and waste (Figure 2).
Image Modified
Figure 2. Peristaltic pump. (1) plastic sleeve. (2) metal hook tubing. (3) waste outlet.


Computer Control


An external computer workstation provides control, communication, and data analysis.
Anchor_Ref352228586
_Ref352228586


Software


Agilent provides a full suite of applications for various user requirements; for our needs we use the Simple Reads application. Simple Reads is used to perform simple absorbance readings of single samples. There is, however, a built-in programming language, Applications Development Language (ADL), which allows complete customization of Cary WinUV to specific applications.

...


Methods


The described methods are based on ODP Technical Note 15, Chemical Methods for Interstitial Water Analysis Aboard the JOIDES Resolution, Aug 1991; J.M. Gieskes, T. Gamo, and H. Brumsack.

...

C: Sodium Nitroprusside/Sodium Nitroferricyanide solution (1200 mL) – make fresh daily

Wiki Markup
Dissolve 75 mg sodium nitroprusside (Na{~}2{~}\[Fe(CN){~}5{~}NO\]) in 100 mL nanopure water. *Note:* do not mix sodium nitroprusside with acid, as HCN fumes can will be liberated.!

D: Alkaline solution (500 mL) – make fresh monthly





D: Alkaline solution (500 mL) – make fresh monthly


Dissolve 7.5 g trisodium citrate (Na3C6H5O7) and 0.4 g sodium hydroxide (NaOH) Dissolve 7.5 g trisodium citrate (Na3C6H5O7) and 0.4 g sodium hydroxide (NaOH) in 500 mL nanopure water.

...

F: Ammonium standard: 0.10 M Ammonium (1000 mL) – make fresh monthly


a.

Dry ammonium chloride (NH4Cl) in oven overnight, then dissolve 5.345 g dried NH4Cl in nanopure water; bring to 1000 mL in a volumetric flask.

b.

Alternatively, use non-dried NH4Cl and determine the chloride content of the standard solution by means of chloride titration. For reasonable accuracy, use a 500 µL aliquot (Cl = 0.1 NM) to obtain almost the same concentration of Cl as in IAPSO reference standard seawater.




Standards


50 mL batches are stable for 1 month. (Note that the 50 µM solution is prepared from the 1000 µM standard, and the rest of the standards are made from the 100 mM primary stock standard.)


concentration
(µM)

volume of nanopure water
(mL)

volume of standard
(mL)

note

0

50.00

0

 

50

47.50

2.50

(dilute from 1000 µM std)

100

49.95

0.05

(dilute from 100 mM std)

200

49.90

0.10

(dilute from 100 mM std)

400

49.80

0.20

(dilute from 100 mM std)

600

49.70

0.30

(dilute from 100 mM std)

800

49.60

0.40

(dilute from 100 mM std)

1000

49.50

0.50

(dilute from 100 mM std)

1500

49.25

0.75

(dilute from 100 mM std)

2000

49.00

1.00

(dilute from 100 mM std)

3000

48.50

1.50

(dilute from 100 mM std)




Procedure


Concentrations of ammonium may differ occur at different sites. Typically in areas with strong evidence of organic carbon diagenesis (e.g., organic carbon–rich sediments), high concentrations of NH4+ can be expected. In that case, sample aliquots must be made appropriately small or sample dilution may be required. The range can be established by using a sample near the alkalinity maximum. Once the range has been determined, prepare standards that cover this range. In this manner, samples and standards are treated in a similar way.
The high range of the instrument is 3.0 Absorbance units (Abs). The results are linear up to this point. If the readings are expected to be higher than 3.0 Abs, use a sample volume of 30 µL instead of 100 µL and divide measured concentrations by 0.3 to arrive at the final sample concentration.
Care should be taken to use clean vials, preferably not those used for Si and PO4 determinations, in which ammonium molybdate is used as a reagent. Note: The order of dilution matters, so do not change this order. Shake samples after EACH addition.


1.

Transfer 200 µL of sample to an 8 mL glass vial.

2.

Add 2 mL of nanopure water to each vial and shake.

3.

Add 1 mL phenol-alcohol solution to each vial and shake.

4.

Add 1 mL sodium nitroprusside to each vial and shake.

5.

Add 2 mL of oxidizing solution to each vial and shake.

6.

Let the color develop (in a dark place) for 6.5 hr* and then determine the absorbance at 640 nm wavelength.


*(from a series of measurements over 8 h, it was found that results stabilized after 6.5 hr.)


...

B. Antimony Potassium Tartrate solution (1000 mL) – make fresh every two months; store in amber glass @ 4°C

Wiki MarkupIn a 1000 mL volumetric flask, dissolve 0.102 g of antimony potassium tartrate trihydrate (KSbC{~}4{~}H{~}4{~}O{~}7{~}- 3H{~}2{~}7·3H2O) in ~600 mL of nanopure water. (If using antimony potassium tartrate hemihydrate \ [KSbC{~}4{~}H{~}4{~}O{~}7{~}- ½H{~}2{~}O\], dissolve 7·½H2O], dissolve 0.09 g.) Dilute to 1000 mL with reagent water.



C. Ammonium Molybdate solution (1000 mL) – stable indefinitely; store in polyethylene @ 4°C

...

In a 1000 mL volumetric flask, dissolve 2 g of ammonium molybdate tetrahydrate (\[NH{~}4{~}\]{~}6{~}Mo{~}7{~}O{~}24{~}- 4H{~}2{~}24·4H2O) in ~600 mL of nanopure water. Dilute to 1000 mL with reagent water.


D. Ascorbic Acid solution (1000 mL) – make fresh every week; store in amber glass @ 4°C
In a 1000 mL volumetric flask, dissolve 3.5 g of ascorbic acid (C6H8O6) in ~600 mL nanopure water. Dilute to 1000 mL with nanopure water. Note: If the reagent is discolored upon creation, the dry ascorbic acid is probably oxidized and must be replaced.

...

Add the following solutions to an appropriate container. Do not add the ascorbic acid reagent until immediately before use. Mix well after each addition to prevent the solution from darkening.


solution

volume (mL)

Ammonium molybdate

50

Sulfuric acid

125

Antimony potassium tartrate

25

Ascorbic acid (immediately before use)

50

Antimony potassium tartrate

25




F. Primary standard: 0.01 M Phosphorus (1000 mL) – make fresh every 4 weeks; store @ 4°C

...

Dry monobasic potassium phosphate monobasic (KH2PO4) in oven at 100°C for two hours; keep in a desiccator while it cools before weighing.
In a 1000 mL volumetric flask, dissolve 1.361 g oven-dried KH2PO4 in ~600 mL nanopure water, then dilute this solution to 1000 mL with nanopure water.


Standards


concentration (µM)

volume of primary standard (mL)

volume of reagent water
(mL)

0

0

100

5.0

0.050

99.95

10.0

0.100

99.90

15.0

0.150

99.85

20.0

0.100

49.90

40.0

0.200

49.80

60.0

0.300

49.70

80.0

0.400

49.60

100

0.500

49.50

200

1.000

49.00

300

1.500

48.50




Procedure


1.

Add 2 mL nanopure water to an 8 mL glass vial.

2.

Add 600 µL sample or standard.

3.

Add 4 mL mixed reagent.

4.

Shake well.

5.

After a few minutes a blue color develops, which remains stable for a few hours. It is best to make the readings at 885 nm ~30 min after addition of the mixed reagent.

Note:

Use a smaller aliquot of sample if the result exceeds the linear range of the spectrophotometer, making up the volume with nanopure water. (For example, for a 300 µL aliquot of a sample, add 300 µL nanopure water.)




Silica


Note: Silica Silicon  Silicon is routinely measured on the ICP, so measurement by spectroscopic analysis is can be considered an alternate method.
Dissolved silica determinations are of great importance in interstitial waters. Often they represent the lithology of the sediments, and the concentrations can vary substantially, especially if highly dissolvable phases such as biogenic opal-A, volcanic ash, or smectite are present. Thus, a wide range of concentrations can be expected, typically from 50 to 1200 µM or higher (especially in hydrothermally affected sediments). The method below usually covers the range, although greater dilutions may be appropriate if sediments or sample sizes necessitate this.
This method is based on the production of a yellow silicomolybdate complex. The complex is reduced by ascorbic acid to form molybdenum blue, measured at 812 nm. The blue complex is very stable, which will enable delayed reading of the samples.

...

C: Metol Sulfite solution (500 mL) – make fresh every month; store in amber glass, tightly sealed @ 4°C

...

Dissolve 6.0 g anhydrous sodium sulfite, Na{~}2{~}SO{~}3{~}, in a 500 mL volumetric flask. Add 10 g Metol (p-methylaminophenol sulfate \ [(C{~}7{~}H{~}10{~}NO){~}2{~}SO{~}4{~}\]) and then nanopure water to make the volume to 500 mL. When the Metol has dissolved, filter the solution through a Whatman No. 1 filter paper. *Note:* This solution may deteriorate quite rapidly and erratically.





D: Oxalic Acid solution (500 mL) – make fresh every month; store in glass

Wiki MarkupIn a 500 mL volumetric flask, dissolve shake 50 g of analytical-grade oxalic acid dihydrate \ [(C{~}2{~}H{~}4{~}O{~}2{~})- 2H{~}2{~}O\] with in ~3500 mL of nanopure water and allow to stand overnight. Shake well and bring to 500 mL volume with nanopure water. Let stand overnight. Decant saturated solution of oxalic acid from crystals before )·2H2O] in 500 mL of nanopure water and allow to stand overnight. Let stand overnight. Decant saturated solution of oxalic acid from crystals before use.





E: Reducing solution (150 mL) – make fresh daily

...

Note: When sodium silicofluoride (Na2SiF6) is dissolved in water, it hydrolyzes to form reactive dissolved silica. Place 1 g of Na2SiF6~ in an open vial and place in a vacuum desiccator overnight to remove excess water. Do not heat.
In a 1000 mL polyethylene volumetric flask, dissolve 0.5642 g dried sodium silicofluoride in ~800 mL nanopure water. Dissolution is slow; allow at least 3 minutes. Dilute the solution to 1000 mL with nanopure water.


Standards


Using a 50 mL volumetric flask, add the following amounts of primary standard to approximately 30 mL of nanopure water and then bring to a total of 50 mL. Store in polyethylene containers.


concentration(µM)

volume of primary standard (mL)

volume of nanopure water (mL)

30

0.5

49.5

60

1.0

49.0

120

2

48.0

240

4

46.0

360

6

44.0

480

8

42.0

600

10

40.0

900

15

35.0

1200

20

30.0




Procedure


Make sure that all reagents are prepared ahead of time. The method has a time factor built in, and therefore it is of great importance to have all necessary reagents ready to go.


1

.Label 20+ mL vials (well washed with nanopure water)

.

12.

Measure 48 mL of nanopure water into the vials (7.63.8 mL for standards and blanks).

23

2.

For standards and blanks (nanopure), pipette 2400 µL of synthetic seawater into the vials.

34

3.

Pipette 4200 µL of sample/standard/blank into the vials.

45

4.

Record time.

56

5.

Add 24 mL of molybdate solution (reagent B) to the vials. A yellow color will develop; allow to mature for exactly 15 minutes (± 15 s).

67

6.

Add 36 mL of the reducing solution (reagent E).

78

7.

Cap the vials and let stand for at least 3 hours.

89

8.

Read absorbances on the spectrophotometer at 812 nm.


Procedure notes


Do not handle more than about thirty samples at a time in order to ensure that the 15 min time limit can be adhered to. Make sure that there are no large fluctuations in room temperature.

Do not use synthetic seawater in dilutions of the primary standard. This could cause the decrease in reactive silica in a few hours as a result of polymerization reactions.

The reason for adding 400 µL of synthetic seawater to the standards is to maintain a reasonably uniform salt content in relation to the samples, this suppressing a potential salt effect on the method.

It is important to wait at least three hours for the blue color to develop; the higher the concentration, the longer the time. The color remains stable for many hours, and reading after 4–5 hours may, in fact, be a good idea. Again, consistency in time limits is advisable.


...


_Ref352228645Sulfide
The sulfide method is based off a method developed by Cline in 1969. This method called for very large volumes of water (50mL). This method was modified on the BONUS Baltic Gas expedition in 2011 to work with sample volumes in the 1-5mL range.
The method is a bit tricky in that the reagent concentrations change depending on what concentration range your samples fall in:

  • High range: 6-80 uM
  • Low range: 1-10 uM.

All samples need to be preserved during splitting with a 1% Zinc Acetate Solution. It can require a relatively large amount of sample in order to do this analysis.
The high range method uses less sample (~500 µL), but if the sulfide level is below 6 µM, it won't be enough. It may still be worth screening samples with the high-range method in order to conserve interstitial water (IW) sample volume. The low-range method requires 4 mL (8x the amount of sample) but is sensitive down to 1 µM. The scientist will have to determine if consumption of 4 mL, possibly 4.5 mL, of sample is worth obtaining the sulfide concentration.
Reagent Solutions


A: 1% zinc acetate (w/v)
For Standard: Prepare 1 L of 1% zinc acetate by dissolving 10 g zinc acetate dehydrate into about 600 mL DI water in a 1 L volumetric flask. Add 1 mL of concentrated acetic acid, bring up to volume, and mix well.
For Splits: Prepare 100 mL of 1% zinc acetate by dissolving 1.0 g zinc acetate dehydrate into about 60 mL DI water in a 100 mL volumetric flask. Add 100 µL of concentrated acetic acid, bring up to volume, and mix well. Set aside for use on the splits.

...

B: 1 mM zinc-sulfide standard suspension


Before weighing the sodium sulfide nonahydrate, wash the crystals off with DI water and dry with a clean Kim Wipe to remove the oxidation products from the crystal before weighing. Try to select very small crystals (or break them up) to facilitate accurate weighing. Dissolve 0.20 g of Na2S·9H2O in 1 L of 1% zinc acetate solution (Reagent A). Store in plastic and cool. Shake vigorously each time before using.

Note: It is very difficult to weigh 0.20 g of this reagent accurately because of the size of the crystals. Weigh 0.20 g within 5 mg and then record the mass used on the bottle. The Baltic Gas methods paper (Ferdelman et al., 2011) describes the use of sodium thiosulfate to standardize the Zn-S suspension, but this has not been the standard practice.


C: 100 µM zinc-sulfide standard


Dilute 10 mL of 1 mM zinc-sulfide standard suspension (Reagent B) up to 100 mL with DI water. This is the stock standard for the dilution of the standard curve.


D: High-Range Sulfide Diamine Reagent


For the high-range (6-80 µM) method, add 500 mL of concentrated HCl to 500 mL of DI water. Mix thoroughly and allow to cool to room temperature. Add 4.0 g of N,N-dimethyl-p-phenylenediamine sulfate and 6.0 g iron chloride hexahydrate (FeCl3·6H2O), mix well, and refrigerate in an amber Nalgene bottle.


E: Low-Range Sulfide Diamine Reagent

...

Make dilutions of the 100 µM zinc-sulfide standard (Reagent C) for the standards to create 0.5 mL of standard at each level. Add the specified volume of Reagent C and DI water as shown in the table below to create the standards.



Standard Level

µL 100 µM
zinc-sulfide


µL DI Water

Concentration (µM)

Blank

0

500

0.0

6

30

470

6.0

20

100

400

20

40

200

300

40

60

300

200

60

80

400

100

80


Shake the zinc acetate-preserved sample well before pipetting into it. For this method, the sample volume should be 0.5 mL. Add 40 µL of the high-range diamine reagent (Reagent D) to the sample, blank, and standards, and shake each vial. Place the samples, blank, and standards, into the dark for 30 minutes; the standards should turn blue. If the visual appearance of any of the samples is darker blue than the highest standard, dilute the sample in the dilution reagent (Reagent F) until it is within the color range of the standards.

...

Make dilutions of the 100 µM zinc-sulfide standard (Reagent C) for the standards to create 4 mL of standard at each level. Add the specified volume of Reagent C and DI water as shown in the table below to create the standards.



Standard Level

µL 100 µM
zinc-sulfide


µL DI Water

Concentration (µM)

Blank

0

4000

0.0

1

40

3960

1.0

2.5

100

3900

2.5

5

200

3800

5.0

7.5

300

3700

7.5

10

400

3600

10


Shake the zinc acetate-preserved sample well before pipetting into it. For this method, the sample volume should be 4 mL. Add 320 µL of the low-range diamine reagent (Reagent E) to the sample, blank, and standards, and shake each vial. Place the samples, blank, and standards into the dark for 30 minutes; the standards should turn blue. If the visual appearance of any of the samples is darker blue than the highest standard, dilute the sample in the dilution reagent (Reagent F) until it is within the color range of the standards.

...

Analyzing Samples


Preparing the Instrument


1.

Engage the tubing on the peristaltic pump. Make sure that the waste line goes into a receptacle.

2.

Turn on the unit and let it warm up for at least three hours.

3.

Confirm that the instrument Beam Mode is correctly configured. This shouldn't change, but to confirm, in Varian > Cary WinUV start the Advanced Read application. Click the Setup button and select the Options tab. Confirm in the Beam Mode area that Double Beam is selected and Normal is also selected. (Figure 3)

4.

To start the measuring computer application, select Varian > Cary WinUV > Simple Reads.

5.

Clean the 'fill' tubing with a Kimwipe and aspirate nanopure water about six times, without returning. This is done by depressing the fill button on the front of the Cary unit. (Figure 4)


Note: The Simple Reads application must be running to be able to operate the fill/return features on the Cary unit.


Image Modified
Figure 3. Setup Screen.

...

...



Image Modified
Figure 4. Peristaltic Pump.

...



Running Samples


Preparing Samples

...

Setting up Instrument Parameters


1.

Click the Setup button to display the Setup dialog box (Figure 5).

2.

Select Read at Wavelength and enter the wavelength for the analysis along with 1 s measurement time.

2.

Click OK. The new wavelength is shown in the Abscissa status display at the top right of the application window.

3.

To zero the current ordinate value, click the Zero button at the left side of the screen. This is best done after aspirating with nanopure water a few times. Do not re-zero the instrument during a measurement batch run.

4.

In the Y Mode group, select the ordinate mode required (Abs).

5.

To clear the contents (text) of the Report area, click on the Clear report button.



Image Modified
Figure 5. Setup Dialog Box.


Measuring the Samples


1.

Insert the aspiration tube completely into the sample vial.

2.

Depress the fill button on the front of the Cary. Make sure that the peristaltic pump aspirates the liquid. Do this a second time.

3.

Monitor the absorbance value in the top left (above the Setup button) of the application window. When the value stabilizes, or wait at least 5 to 8 seconds, CONSISTENTLY, click the Read button to measure the sample at the specified wavelength.

4.

The result of the sample read appears in the Report area and includes the ordinate reading obtained and the wavelength at which the reading was measured (Figure 6).

5.

Depress the return button on the front of the unit to return the aspirated sample to the sample container for further analysis if required.

6.

Remove the aspiration tubing from the sample vial and wipe clean with a Kimwipe.

7.

Aspirate nanopure water in between samples to flush out the line.

8.

Continue sample readings. Once all samples have been read, the data can be saved and/or printed.
Check each uptake for consistent/inconsistent liquid volume extractions. Also check for bubbles in the lines and worn peristaltic tubing. This could lead to carry-over or tube voids in the view window. Replace pump tubing immediately.

10.

Click on the Print… button to display the Windows Print dialog box to choose the printer and pages to print of the current displayed report (Figure 7).

11.

To save data, select Save Data As… from the top File menu option. Save the data in an RTF format.

12.

This file may be opened in Excel, using space-delimited option, and formatted per the requirements of the
LIMS Integration section (i.e., text ids, analytes, and concentrations will have to be added).




READ button Image Modified
Figure 6. Measuring Samples.

Image Modified
Figure 7. Print/Save Samples.


Data Analysis


Using the results from the reads of the standards, create a calibration curve from the plot of the concentration vs. absorbance values. Use this equation to extrapolate the sample concentrations from their corresponding absorbance value. This can be done in the same spreadsheet as created above in the Simple Reads application. This sheet can be loaded into the LIMS Spreadsheet Loader application as outlined in the LIMS Integration section.

...

Aspirate approximately eight cycles of nanopure water, release the tubing on the peristaltic pump, turn off power to the unit and exit from the software. Clean any spills that may have occurred.
Empty the waste container and rinse with tap Anchor_GoBack_GoBack water.
Anchor_Ref307906541
_Ref307906541


QAQC


QA/QC for analysis consists of calibration verification using check standards, blanks and replicate samples.

...

If the control limits are exceeded (generally 3 standard deviations from the mean), the instrument system is considered out of control and all samples in the current analytical batch are invalid and must be rerun after the system is proved to be in control.

...


LIMS Integration


Results are stored in the LIMS database associated with an analysis code and an analysis component.


Analysis

Component Name

Component value

Unit

SPEC

analyte

ammonium

 

 

phosphate

 

 

 

silica

 

 

concentration

concentration of analyte

µM



Data collected is transferred to the LIMS database using IODP's Spreadsheet Loader application. This is best done by entering the results into an Excel spreadsheet in a format similar to the pattern below, keeping the appropriate number of columns blank, and omitting the column headers. Then run the Spreadsheet Loader application. Go to File > Load and it should import something like below. To upload into the database, go to Lims > Upload and status messages will appear in the blank window (Figure 8).


Image Modified
Figure 8. Spreadsheet uploader.

...


Maintenance and Troubleshooting


Cleaning


Any spills in the sample compartment should be immediately wiped up and any deposits on the sample compartment windows should also be removed.
The exterior surfaces should be cleaned with a soft cloth and, if necessary, this cloth can be dampened with water or a mild detergent.

...

To replace a fuse, disconnect the unit from the power supply, and replace the blown fuse with one of the type and rating as outlined in the hardware specifications section.


1.

Disconnect the instrument from the power supply.

2.

Undo the fuse cap by pressing the cap and turning it counter-clockwise.

3.

Carefully pull out the cap. The fuse should be held in the fuse cap.

4.

Check that the fuse is the correct type and is not damaged. If necessary, replace it.

5.

Place the fuse into the cap, push the cap in and then turn the cap clockwise.

6.

Reconnect the instrument to the power supply.


Cary Win UV Help


Varian provides extensive help resources, available from the software CD. After installing the help utilities, go to START > All Programs > Varian > Cary WinUV > Cary Help.
Cary Help offers troubleshooting information, maintenance information like how to replace lamps, aligning lamps and mirrors, and replacing fuses and cleaning the flowcells.

...


...

Contact Information


Sales and ServiceVarian Instruments2700 Mitchell Dr.Walnut Creek, CA 94598 Phone:1 800 926 3000 E-mail: customer.service@varianinc.com

...

...


Health & Safety


Ultraviolet Radiation


The deuterium in the instrument emits hazardous ultraviolet radiation that can cause serious damage to eyes. NEVER look directly at the lamp.
Ozone can be generated by radiation from the source lamp. Exposure to ozone can result in irritation to exposed severe radiation damage to the skin, eyes, and upper respiratory system. The maximum permissible exposure level is 0.1 ppm. Be sure to work with the deuterium lamp using adequate ventilation.

...

Ammonium Chloride


Health effects


Inhalation:

irritant

Ingestion

irritant

Contact:

irritant


Incompatible materials


Concentrated acids, strong bases, silver salts, potassium chlorate, ammonium nitrate, hydrogen cyanide

...

Ammonium Molybdate Tetrahydrate


Health effects


Inhalation:

irritant

Ingestion:

irritant

Contact:

irritant


Incompatible materials


Alkali metals

...

Hazards


Poison


Health effects


Inhalation:

irritant

Ingestion:

toxic

Contact:

irritant


Incompatible materials


Alkali metals and their carbonates, perchloric acid, reducing agents

...

Ascorbic Acid


Health effects


Inhalation:

mild irritant

Ingestion:

mild irritant

Contact:

mild irritant


Incompatible materials


Strong oxidizers, alkalis and alkali hydroxides, iron, copper

...

Bleach (Sodium Hypochlorite)


Health effects


Inhalation:

irritant

Ingestion:

irritant

Contact:

irritant


Stability


Decomposes in air and sunlight

...

Hazards


Flammable


Health effects


Inhalation:

irritant

Ingestion:

irritant

Contact:

irritant


Incompatible materials


Oxidizing agents, acids, alkali metals, ammonia, peroxides, silver nitrate

...

Poison, corrosive


Health effects


Inhalation:

corrosive, may be fatal

Ingestion:

corrosive, may be fatal

Contact:

corrosive, may cause permanent damage


Incompatible materials


Strong bases, metals, metal oxides, hydroxides, amines, carbonates, alkaline materials and strong nitric acid.

...

Magnesium Sulfate heptahydrate


Health effects


Inhalation:

minor irritant

Ingestion:

minor irritant

Contact:

minor irritant


Stability


Stable but loses some moisture on exposure to dry air at room temperatures

...

Combustable, toxic, possible sensitizer


Health effects


Inhalation:

harmful

Ingestion:

harmful

Contact:

harmful


Incompatible materials


Strong oxidizing agents, acids, acid halides

...

Poison, corrosive


Health effects


Inhalation:

may cause severe burns

Ingestion:

toxic, may be fatal

Contact:

may cause severe burns


Stability


May become unstable under heat

...

Hazards


Carcinogen


Health effects


Inhalation:

slightly hazardous, non-corrosive for lungs

Ingestion:

VERY hazardous

Contact:

VERY hazardous

Skin contact:

corrosive, sensitizer, permeator

Eye contact:

corrosive


Stability


Stable but air and light sensitive

...

Potassium Phosphate, Monobasic


Health effects


Inhalation:

irritant

Ingestion:

irritant

Contact:

irritant


Sodium Chloride


Health effects


Inhalation:

mild irritant

Ingestion:

mild irritant in large doses

Contact:

irritant with eye contact




Incompatible materials


Lithium, bromine trifluoride

...

Sodium Sulfite


Health effects


Inhalation:

mild, cough, sore throat

Ingestion:

mild

Contact:

mild


Stability


Stable


Incompatible materials

...

Hazards


Toxic


Health effects


Inhalation:

toxic

Ingestion:

toxic, may be lethal

Contact:

irritant


Stability


Hygroscopic


Incompatible materials

...

Hazards


Toxic


Health effects


Inhalation:

toxic

Ingestion:

toxic

Contact:

irritant


Incompatible materials


Metals

...

Poison, danger, corrosive


Health effects


Inhalation:

strong irritant

Ingestion:

corrosive, may be fatal

Contact:

corrosive, may cause severe burns


Stability


Reacts violently with water

...

Trisodium Citrate


Health effects


Inhalation:

irritant

Ingestion:

irritant

Contact:

irritant


Incompatible materials


Strong oxidizers

...

Guidelines for Handling Chemicals


Wear gloves to protect from specific hazards.

Wash hands immediately after removing gloves, after handling chemical agents, and before leaving the laboratory area.

Wear laboratory coats and gloves only in the laboratory area.

Confine long hair and loose clothing.

Wear close-toed shoes.

Do not store or prepare food, eat, drink, chew gum, apply lip balm or cosmetics, or handle contact lenses in areas where chemicals are handled.

Label all chemical containers and reagent solution bottles.

Keep laboratory work areas clean and uncluttered.

Consult the MSDS before using any chemical.

Know the emergency procedures for handling chemicals used.

Vent into local exhaust devices vapors, fumes, mists, dusts, or gases.


Waste Management


Waste management includes proper handling or reaction byproducts, surplus and waste chemicals, and contaminated materials. Each worker is responsible for ensuring wastes are handled in a manner that minimizes personal exposure and the potential for environmental contamination.
Dispose of cleaning solutions properly.
Dispose of surplus and expended reagents properly.
Select the least hazardous chemical for the job and use only in quantities needed. Minimize waste whenever possible by reducing both the volume and physical hazard or toxicity of the material.

...

Gieskes, J.M., Gamo, T., and Brumsack, H., 1991. Chemical methods for interstitial water analysis aboard JOIDES Resolution, ODP Tech. Note, 15. doi:10.2973/odp.tn.15.1991.
Kastner, M., Elderfield, H., Martin, J.B., Suess, E., Kvenvolden, K.A., and Garrison, R.E., 1990. Diagenesis and interstitial water chemistry at the Peruvian margin—major constituents and strontium isotopes. In Suess, E., von Huene, R., et al., Proc. ODP, Sci. Results, 112: College Station, TX (Ocean Drilling Program), 413–440. doi:10.2973/odp.proc.sr.112.144.1990
Noriki, S. 1983. Silicate correction in the colorimetric determination of phosphate in seawater. J. Oceanograph. Soc. Japan, 39(6):324–326. doi:10.1007/BF02071829
Presley, B.J., 1971. Techniques for analyzing interstitial water samples: Appendix Part 1: determination of selected minor and major inorganic constituents. In Winterer, E.L., et al., Init. Repts. DSDP, 7(2): Washington, DC (U.S. Govt. Printing Office), 1749–1755. doi:10.2973/dsdp.proc.7.app1.1971
Solorzano, L., 1969. Determination of ammonia in natural waters by phenol-hypochlorite method. Limno. Oceanogr., 14:799–801.
Strickland, J.D.H., and Parsons, T.R., 1968. A manual for sea water analysis. Bull. Fish. Res. Board Can., 167.


Appendix

...


Reagent Cheat Sheet


Ammonium

 

 

 

 

 

reagent

total volume (mL)

solute

amountadded

solvent

when

A

100

95% ethanol

100 mL

NA

daily

B

101

phenol

1 mL

reagent A

daily

C

100

sodium nitroprusside

75 mg

nanopure

daily

D

500

trisodium citrate /
sodium hydroxide

7.5 g /0.4 g

nanopure

monthly

E

50

Sodium hypochlorite (bleach)

1 mL

reagent D

daily

F

1000

ammonium chloride (dried)

5.345 g

nanopure

monthly

Phosphate

 

 

 

 

 

Reagent

Total volume (mL)

Solute(s)

Amountadded

Solvent

When

A

1000

sulfuric acid

10 mL

nanopure

anytime

B

1000

antimony potassium tartrate trihydrate

0.102 g

nanopure

every two months

C

1000

ammonium molybdate tetrahydrate

2 g

nanopure

anytime

D

1000

ascorbic acid

3.5 g

nanopure

weekly

E

250

reagent Creagent Areagent Dreagent B

50 mL125 mL50 mL25 mL

 

6 hours

F

1000

potassium phosphate monobasic (dried)

1.361 g

nanopure

monthly

Silica

 

 

 

 

 

Reagent

Total volume (mL)

Solute(s)

Amountadded

Solvent

When

A

500

sulfuric acid

250 mL

nanopure

monthly

B

500

ammonium molybdate tetrahydrate

4 g

nanopure

monthly

C

500

anhydrous sodium sulfite Metol

6.0 g10 g

nanopure

monthly

D

500

oxalic acid dihydrate

50 g

nanopure

monthly

E

150

reagent Creagent D

50 mL30 mL

nanopure

daily

F

1000

sodium chloridemagnesium sulfate heptahydrate

25 g8 g

nanopure

monthly

G

1000

sodium silicofluoride (dried)

0.564 g

nanopure

bi-monthly