- Installing and calibrating light array (see Light Installation).
- Setting black and white saturation gain levels (see Maximum Dynamic Sensor Range).
- Setting color balancebalance Author: Where is this covered?
- Making camera corrections (see Routine Camera Adjustments).
- Setting track velocity (see Track Speed Example).
- Adjusting knee slope, if needed (see Procedure: Iterative Adjustment).
- Performing QA/QC to confirm camera settings (see Quality Assurance/Quality Control).
...
The laboratory technician calibrates the system when needed by adjusting camera settings and analyzing an imaged Kodak Q13 grayscale standard.
Calibration adjustments are performed using an iterative process of imaging a ruler placed crosswise on the track then counting the pixels between millimeter marks, moving the camera up or down, and repeating. See See <LINK> [Author: Where is this procedure documented?] for details on this procedure.
...
Sample and Analysis Components
Analysis Code | Component | Description |
IMAGES | adapter_mag |
|
adapter_tube |
| |
comment |
| |
compound_field_of_view |
|
contrasting_method |
| |
dimension_horizontal |
|
dimension_vertical |
| |
file_create_date |
| |
filename |
|
filesize |
| |
filters |
| |
illumination_type |
| |
light_source |
|
objective |
|
objective_field_of_view |
|
objective_mag |
|
ppi |
|
total_magnification |
| |
tube_field_of_view |
|
Uploading Data
After section is imaged, select the Ends of Material Points > Save, and a virtual label with section identification will attach to the saved image while the concatenation endpoints are recorded to the LIMS.
...
Common problems encountered when using the core imager and their possible causes and solutions:
Issue | Possible Causes | Solution |
Actuator squeal | NA | Lightly tap the actuator housing to silence the noise |
Image too dark | Manual F-stop on the camera closed down | Have technician adjust F-stop aperture |
Exposure time is too low | Increase exposure time |
Focused lights are not aimed at the correct spot | Adjust lights |
Track is |
“stuck” | Run was aborted with the software abort switch | Reset software and run sample again |
Run was aborted with the hardware abort switch | Reset hardware and run sample again |
Gantry flag has tripped the end-of-travel limit switch | Adjust gantry flag and run sample again |
Current limit on motors was exceeded | Check the Galil AMP-19520 for LED error indicators. Call track technician or ET to reset the motor controller |
Torque limit on motors was exceeded. <we need to check how Labview handles this!>[djh1] |
| |
Image indicates that camera was triggered erratically OR no image acquired | Camera was left in Free Run mode in MAX | Set camera to Externally Triggered for normal operation |
Linear encoder head has failed | Call an ET to verify/repair |
Lens cap is on | Remove cap and repeat image capture procedure. |
...
Scheduled Maintenance
Frequency | Task |
---|---|
Daily | Ensure that the color standards, ruler, and barcode imager lens are free from dust, smudges, and crumbs. |
Weekly | Using a mirror, ensure that there are no fingerprints or smudges on the camera lens. Call the imaging specialist if the lens needs cleaning. Do not attempt to clean it yourself! |
Monthly | Check socket head cap screws in the camera and lights mounting plates for looseness. |
Every Expedition |
|
Annually |
|
...