The Agilent Technologies SP3 autosampler facilitates sample introduction into the Cary Spectrophotometer with minimal operator time. Yay!
The SP3 is connected to PC via RS232 cable to Keyspan USB convertor.


Image Modified
The SP3 is connected to Cary unit via RSA sample inlet tube.


Image ModifiedImage Modified



The reservoir peristaltic pump allows DI water to be pumped through the system between samples to flush lines.
Image Modified


Pump speed is controlled by a dial above the pump. This pump fills the reservoir but does not change the flow into the Cary's flow cell; that is controlled by the spectrometer's peristaltic pump. (fills reservoir but does not change flow into Cary). That is controlled by the Cary's peristaltic pump.
Sample trays allow up to 180 samples to be run. First vial is a "zero" (DI water).


Image Modified

Software setup

Advanced Reads setup

...

Image Added


Image Added

RSA (sipper) settings


 Image Added 30 sec uptake
30 sec read delay
Image Removed

SPS3 configuration

...

settings# 

Image Addedrinse time : 100 sec# Click on tray (on this screen the number for tray1 has turned red).

  1. Assign a Rack Type.
  1. To assign zero and samples, first click tube color and then click Individual tray numbers.

Image Added
Image Added

 
 5.  Anchor_GoBack_GoBack Image Removed
Image Removed
Click on Custom Racks in the previous screen to configure sample racks.


This example shows a simple 60 vial rack.
Image Modified

Running samples

  1. Prepare samples according to the protocol outlined in the Cary Spectrophotometer User Guide and let the machine warm-up for a couple of hours at the correct wavelength for the analyte in question.
  2. Click on START in the main Advanced Reads window.
  3. A window will pop-up showing the samples entered in the previous. Click OK.

...

  1. 4.Save the data to a file for later viewing.


Image Added 



Image Modified

5. View where the samples need to go. Click OK.


Image Modified

6. View the results in real-time.


Image Modified

7. Save results to a .csv file:

    1. Select File -> Export report (*.csv)
    2. Enter the location/file name.

8. Copy/paste into calibration spreadsheet for manipulation and subsequent use in Spreadsheet Uploader.