Liquid Scintillation Counter Quick User Guide
X328T (Sept-Oct 2010)
Chieh Peng


Setup


Computer/Laptop is connected through comp port #2 on the counter


Startup

  1. Insert "Wallac 1414 WinSpecial Ins.Prog "3.5" disc into the front port
  2. Log on to the computer using daq/daq
  3. Open "1414 Winspecial" software on the computer
  4. Turn on the scintillation counter, the power switch is on the back
  5. Winspecial main window status line should show "Loading" > "counter is busy"> "counter online"
  6. Now the counter is ready


Sample Loading

  1. Two types of sample racks, can be used in combination
  2. Locate the barcode on the side of the sample rack. This should be your first rack, so the instrument knows where to "start" the measurement
  3. Barcode should be facing the right hand side, when operator is facing the counter. The first rack should go in the right side of the instrument closer to the back

Image Modified

To Run Sample/Standard

  1. 5 Sub-windows show up in the Winspecial main window

...

  1. Instrument will calculate base on the half-life and give you the current strength in comparison to the measurement result.
  2. Easy counter cam measure 3H, 14C, 125I, 32P, 35S, and 45Ca. We only carry the first 2 standards.
  3. Easy counter test value, please refer to software manual P67.


Safety

  1. Store all the standards in the grey plastic boxes when they are not being used.
  2. Wash hands with soap after touching the standards and other potential contaminated samples.
  3. Always follow RadVan user protocol.