Prepare reagents

Prepare the following reagents according to the recipes in the user guides. Remember to put your initials and the date on all reagents made.

    • Alkalinity/pH/chloride titrations
    • Coulometer
      • IC
      • ICP
      • SPEC
          • HCl solution, KOH and AgNO3 if needed
        • IC 
          • all solutions except for anion eluent, which should be made shortly before analysis
          • pourable calibration standards
        • ICP
          • 2% and 3% TM nitric acid solutions, also 10% if doing hard rock
          • internal standard
          • primary in-house cocktail
          • pourable in-house and IAPSO calibration standards
        • SPEC
          • Ammonium
            • Ammonium standard and calibration standards, alkaline solution
          • Phosphate 
            • Phosphate primary standard (dry monobasic potassium phosphate in oven overnight) + calibration standards
            • Sulfuric acid solution
            • Antimony potassium tartrate solution
            • Ammonium molybdate solution


      Check gas bottles

      Check the pressure of the oxygen (ChemLab), helium and argon bottles (Tween Pallet Stores). Ensure the bottles are above 200 psi; otherwise, swap them out with fresh bottles. When a Pallet Stores bottle array is empty, tear the yellow tags to EMPTY, disconnect the manifold and connect it to full bottles. It is good practice to always have a full gas bottle array ready to go. 

      ...