Figure 28. Fused glass bead. 

  • An analytical procedural blank is prepared identically to the samples, with the exception that only 0.4 g of flux (the pre-weighed onshore amount) is fused and dissolved. An additional 0.1 g of flux is NOT added to mimic the TDS of the 0.5 g mix of sample + flux because this would provide an inaccurate quantitation of the impurities introduced by the amount of flux used in preparation of the unknowns.

Weighing the Sample

Note: This process is typically done by the chemistry technicians.
Weighing the sample is a critical step. The sample weight should be as close to 100 mg as possible. Inaccuracies in the weight will show up in the analytical results. Print small labels for each sample and place on your small, clear capped vial. On the lid label a sticker with the core, section, and interval.

...