A common program is an increase in temperature of 3°C/min to a target temperature of 900°C and a hold of one hour. Then ramp up at a rate of 3°C/min to a target temperature of 1025°C and hold for four hours. This ramp cycle is already programmed into the furnace and corresponds to 'Program 1'. To check or edit a program see additional guides attached to the furnace itself. It is also possible to run the furnace manually without a ramp up cycle. Discuss with scientists their preference.
Either enter the desired temperature manually or select your program. If you are running the furnace manually, enter the desired temperature; no other buttons or steps are needed. If selecting a program, press and hold 'Run' (Figure 45).
...
- Rinse crucibles with DI water.
- If beads are stuck to the bottom, sonicate with DI water for 30 min or more.
- Place crucibles in HNO3 10% bath for 12 hr. If you notice any signs of residue, leave in the acid bath for longer.
- Clean a Tupperware container with isopropyl alcohol. Lay down sheets of paper towel and a large kimwipes.
- Take crucibles out of the acid bath and rinse with DI water 3 times. Place crucible in the container. If the crucibles require polishing, see Polishing the Platinum Crucibles.
- Cover all crucibles with a large Kimwipe and place in the drying oven in the Chemistry Laboratory. Leave overnight.
- When dry remove crucibles and bring into the XRD laboratory. Wrap each crucible in a small Kimwipe and and place back in the safe. Lock the safe after all crucibles have been put back.
...
- Wait until the temperature returns to 1020°C., then leave the samples in for 6 min.
- Remove each sample crucible with the long tongs. When the sample is removed, turn the crucible 45° to one side.
...