Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Vsyringe = Volume of the secondary standard extracted via the autosampler or manual syringe.



PFMD Stock Level A (400,000 ng/

...

mL)
  1. Pipette 12.25 ml of Optima grade Hexanes (that is what we currently use as of X395C) into a 20 mL headspace vial.
  2. Cap the vial with a PTFE septum (cream-colored "shiny" side facing the vial) and a magnetic crimp cap.
  3. Using a 10 uL cemented needle syringe, add 2.5 µL of neat PFMD into the vial. Let the needle remain the vial for a few moments after the injection to ensure all PFMD from the needle tip has evaporated into the vial. 

...