1.

On the Alkalinity/Standard tab, ensure dispensing rate, drift span, and standard ratio are set correctly (Figure 7).

2.

Ensure the water bath temperature is set to 25°C and pipette sample in titration vessel.

3.

After rinsing the electrode with DI water and gently blotting dry with a Kimwipe, insert the electrode tip into sample (not touching the bottom of the cup).

4.

Click SAMPLE.

5.

Select measurement type from the list, enter operator name (this must be the same as LIMS user ID), text_ID, and sample volume (Figure 8). Click OK.

6.

After taking the initial mV and pH readings, the software prompts the user to insert the acid dispensing tip and click GO (Figure 9). If you make a mistake, click ABORT and no acid will be added to the sample. After clicking GO, the sample will be titrated to completion.

7.

Click Done/Save to upload data (see LIMS Integration).








Figure 8. Measuring Alkalinity. The plot on the left side is the signal coming directly from the electrode in real time.

...

Data reports are not available at the instrument, but data can be viewed on the View Datalog tab (Figure 11). Also, the software appends the alkalinity results to the DAT file C:\ALKALINITY\LOG\ALKALINITYSAMPLES.DAT. The best way to view the alkalinity values to be used in graphs and reports is via LIMS Reports. It is also highly recommended to record the alkalinity and pH values in the blue laboratory notebook to protect against inadvertent data loss.


Figure 11. Datalog.

Image Removed






Figure 12. Viewing alkalinity data in LIMS.

...