– | Make sure the probe storage container is filled with 3 M KCl solution and the reservoir is filled with 0.1 M HCl solution. |
– | Ensure no air bubbles are present in the acid dispensing line. Press DOS on the body of the titrator to push acid through the line and remove air bubbles. |
– | Change out the Drierite trap when ~50% of the color turns from blue to pink. |
– | If you are hearing In case of a bubbling noise, top up your temperature-controlled water reservoirbath. Tap Use tap water is fine. Check titration vessel for air bubble and eliminate if present. |
...