You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 3 Next »

GC3-Natural Gas Analysis User Guide: Drilling Safety Monitoring


Manual Information




Author(s):

C. Bennight

Reviewer(s):

L. Brandt, C. Neal, K. Marsaglia

Editor(s):

K. Graber, L. Peters

Management Approval (Name, Title, Date):

D.J. Houpt, Supervisor of Analytical Systems, 9/24/2010

Audience:

Scientists, Laboratory Technicians

Origination date:

5/12/2008

Current version:

Version 1.0 9/24/2010

Revised:

 

Domain:

Chemistry

System:

Gas Chromatography

Keywords:

Hydrocarbons, Natural Gas, Headspace


Changes to User Guide


Summarize requested modifications to this user guide in an e-mail and/or annotate the PDF file and e-mail change requests to techdoc@iodp.tamu.edu.


User Guide Contents




Topic

See page…

Introduction

Apparatus, Reagents, & Materials

Instrument Calibration/Calibration Verification

Sample Preparation & Analysis

Quality Assurance/Quality Control

LIMS Integration

Health, Safety, & Environment

Maintenance & Troubleshooting (HP6890GC)


Introduction


Overview


Natural gas analysis for hydrocarbons and hydrogen sulfide (H2S) is required to avoid natural gas and oil escaping from the hole and is part of the ship's standard drilling safety plan.
The absolute quantity of hydrocarbons is the primary safety risk during shipboard operations. Gas monitoring via gas chromatography is a means of quantifying the hydrocarbon risk. H2S is another significant risk factor for individuals working in the area. Emergency monitors on the drill floor provide early detection of H2S, while later quantification is performed on the natural gas analyzer (NGA).


Hydrocarbon Gases


Hydrocarbon generation in sediments is a result of thermal decomposition (maturation) of biogenic organic matter. C1–C4 hydrocarbons may be generated in significant quantities in sediment via two processes:


  • Biogenic: biogenic hydrocarbons, typically characterized by methane, are produced in a sulfate-free environment via the reduction of dissolved bicarbonate.
  • Thermogenic: thermogenic hydrocarbons are produced in sediments in direct proportion to temperature. C5 and other heavier hydrocarbons are almost always the result of thermal generation of hydrogen-rich organic matter at temperatures typically ~100°C or greater.


Hydrogen Sulfide


Sulfate-reducing bacteria produce hydrogen sulfide in euxinic sediments. This may occur in a relatively shallow part of the sediment. Thermochemical sulfate reduction of sulfate by hydrocarbons in reservoirs occurs under high temperature (>127°C ~ 140°C).


Theory of Method


Two instruments monitor gases in core headspace and core void samples:


  • GC3: Agilent 6890 gas chromatograph (GC) with flame ionization detector (FID). This instrument measures C1–C3 hydrocarbons:
  • Methane (CH4)
  • Ethene (C2H4)
  • Ethane (C2H6)
  • Propene (C3H6)
  • Propane (C3H8)
  • NGA: Agilent 6890 GC with FID and thermal conductivity detector (TCD). This instrument measures C1–C7 hydrocarbons as well as some additional compounds:
  • Methane (CH4)
  • Ethene (C2H4) + Ethane (C2H6)
  • Propene (C3H6)
  • + Propane (C3H8)
  • n-Butane (C4H10)
  • iso-Butane (CH3-C3H7)
  • n-Pentane (C5H12)
  • iso-Pentane (CH3-C4H9)
  • n-Hexane (C6H14)
  • iso-Hexane (CH3-C5H11)
  • n-Heptane (C7H16)
  • iso-Heptane (CH3-C6H13)
  • Nitrogen (N2)
  • Oxygen (O2)
  • Carbon dioxide (CO2)
  • Carbon disulfide (CS2)
  • Hydrogen sulfide (H2S)


The FID column on the NGA cannot separately quantify ethene/ethane and propene/propane, and they are reported as combined values. The TCD column does separate these components.


Apparatus, Reagents, & Materials


Instruments


The GC3 and NGA systems are both based on an Agilent 6890 GCs. These systems were further customized with specialized gas injection inlets and various column, detector, and valving systems for gas monitoring.


GC 3


The GC3 system is equipped with a 1/6 inch VALCO union injector with 2 µm screen and an electronically switched 10 port VALCO valve. The column is an 80/100 mesh, 8 ft HayeSep "R" packed column (2.0 mm ID x 1/8 inch OD).
The detector is an FID.


NGA


The NGA gas chromatograph is equipped with 2 detectors:


  • Flame ionization detector (FID)
  • Thermal conductivity detector (TCD)


The TCD flow path travels through a 6 ft x 2.0 mm ID stainless steel (SS) column packed with Poropak T (50/80 mesh), a 3 ft x 2.0 mm ID SS column packed with molecular sieve 13x (60/80 mesh), and 6 ft x 2.0 mm ID SS column packed with 80/100 mesh HayeSep R (acid washed).
The FID flow path traverses a 60 m x 0.25 mm ID capillary column with 0.25 µm DB-1 film.


Instrument Calibration/Calibration Verification


Overview


Before unknown samples can be analyzed for headspace gases, each GC system must have a valid calibration curve and the calibration curve must have been verified using a calibration verification standard.


1

Create/refresh calibration curve (start at least 1 day before reaching site) (see Creating a Calibration Curve).

2

Verify calibration (Running a Calibration Verification Standard).

3

Perform a work flow test (Running a Gas Sample).




Creating a Calibration Curve


1

Prepare 5–7 registered standard gases.

2

Activate GC3/NGA LIMS uploader located at Start > Program Files > IODP > MegaUploadaTron. The uploader must be activated before the calibration is run.

3

In the ChemStation Main menu, click Run Control > Sample Info.

4

Fill in the specific fields on the screen as follows:

  • Operator name: LIMS user account (your last name)
  • Sample name: name of the standard (e.g., STD_D) and the replicate number (STD_D-1, STD_D-2, etc.)
  • Comment: text ID of the standard (scan the label)
    Click OK to close screen.

5

Slowly inject 5000 µL of the first standard gas and observe the floating ball in the flow meter move upward.
Keep the outflow rate on the flow meter <80 mL/min.

6

When the ball in the flow meter indicates flow has fallen to just above 0 (is about to hit 0), press the Start button on the control panel of the GC.

7

When the run has finished, open the Data Analysis screen in ChemStation and click Calibration.

8

On the Main ChemStation menu, select Calibration > Recalibrate.

9

On the Recalibration screen, select Level # and Replace (or Average) as applicable for that level.

10

Repeat Steps 5–9 for 3 replicate standards (CH4: A 25%, B = 50%, C 75%, D = 99%).

11

Click OK to change the calibration value. For NGA calibration, the same standard can be applied to both the appropriate TCD and FID level; you do not need separate standards for TCD and FID.




Running a Calibration Verification Standard


1

Ensure the uploader is activated and the CV standard is registered in LIMS.

2

Click Run Control in the main menu of ChemStation and select Sample Info.

3

Fill in the specific section on the window as follows:

  • Operator name: LIMS user account (your last name)
  • Sample name: common name for standard (e.g., STD_D-1)
  • Comment: text ID of the standard (scan the label)
    Click OK to close the sample info screen.

4

Prepare the CV standard at approximately the mid-point concentration of the curve.

5

Slowly inject 5000 µL of the standard gas, keeping the outflow rate <80 mL/min.

6

Press Start on the GC control panel when the flow meter is just above 0.

7

When the run is finished, the report will automatically display the values. Click Upload in the uploader to submit the data to LIMS.




Running a Blank


1

To run a blank, in the Main menu click RunControl > Sample Info.

2

Fill in the following fields:

  • Operator name: your last name
  • Sample name: "BLANK"
  • Comment: text ID of the blank (scan the label)
    Click OK to close window and save information.

3

Prepare laboratory air (5000 µL) and inject it into the GC in the same fashion as the standards above when the ChemStation software shows Ready.

4

Press the Start button on the GC control panel to start the run.

5

Confirm the chromatogram on the screen shows no peaks. If peaks are present, the system contamination must be found (injector, detector, sample loop, etc.).




Running a Gas Sample


1

Ensure the uploader has been activated.

2

Click Run Control in the main menu of ChemStation and select Sample Info.

3

Fill in the specific section on the window as follows:

  • Operator name: LIMS user account (your last name)
  • Sample name: Exp/site/hole/core/coretype/section/interval (e.g., 324-U1351A 5H4 32-35)
  • Comment: text ID of sample (scan label)
    Click OK to close sample info screen.

4

Prepare a headspace or void gas sample.

5

Slowly inject 5000 µL of the gas sample, keeping the maximum gas outflow <80 mL/min.

6

Press Start on the GC control panel when the ball on the flow meter is just above 0.




Sample Preparation & Analysis


Overview


There are two primary sample types used for natural gas analysis.


  • Headspace gas, which is obtained from core samples by heating a sample to ~70°C.
  • Void gas collected with a vacuum vial.


Occasionally, cores that come on deck have voids with large amounts of free gas. Free gas must be sampled using a sampling device that penetrates the liner and provides a channel for the gas to be drawn into a gas-tight syringe, vacuum vial, or gas sampling bag.


Sampling Tools


  • Sample coring tool (metal cylinder)
  • Sample coring plunger
  • Puncture tool (to penetrate plastic liner)
  • Headspace vial
  • Headspace gasket with crimp top
  • Crimping tool
  • Permanent marker for labeling




Sampling Procedure and Gas Sample Preparation


Headspace Gas


Collect samples from a freshly cut core section at a position within 0.5 inch of the inner side of the core liner (where sample has not been disturbed by contact with drilling fluid or core liner). In addition, the sample must be taken prior to the use of acetone or any other organic solvent in the catwalk area.
The curator authorizes the sampling plan before coring; therefore, the chemistry specialist must know the catwalk sampling plan before taking samples.


Collecting a Headspace Gas Sample




1

Locate a freshly sectioned core (consult with the curator).

2

Gently push the sample coring tool into the core section slightly inward of the edge.

3

Gently pull out the tool. If the sample recovery (% of coring tool with sample) is >80% (~5–7 cm3), proceed; otherwise repeat Steps 1 and 2.

4

Place the open end of the sample coring tool over a clean headspace gas vial and use the plunger to push the sediment into the vial.

5

Immediately place a gasket with a crimp top over the vial and crimp shut.

6

After sealing the vial, immediately write down the sampling interval, location, and any other information for the sample that was just taken. Generate a proper label and apply it to the vial as soon as possible.

7

Place the vial with the sample in a 70°C oven for 30 min to degas the sediment (use timer).

8

Inject extracted gas sample into the GC using syringe (see
Running a Sample).




Collecting a Void Gas Sample




1

Use the puncture tool to make a hole in the core liner to make a channel for the gas.

2

Quickly collect a free gas sample from the small hole with a syringe.

3

Immediately introduce the gas sample into the GC instruments in the same manner as the headspace samples.




Running a Sample




1

Start GC and operation system at least 1 day before reaching site (the system should be fully calibrated and ready for analysis) (see Advanced User Guide).

2

Ensure LIMS uploader is running.

3

Inject 5 mL of headspace gas after the sample has heated in the oven for 30 min.

4

Click Upload if the uploader is not in automatic mode.




Quality Assurance/Quality Control


Overview


QA/QC for GC3/NGA analysis consists of instrument calibration and continuing calibration verification using check standards, instrument blanks, and replicate samples.


Analytical Batch


An analytical batch is a method-defined number of samples with which QC samples including calibration verification, blank, and replicate samples are run. Samples are implicitly grouped into batches based on the spacing between CV samples.


QC Samples


Blank


  • The blank determines the level of contamination originating from the laboratory environment (air) and sample path in the GC (injection port with screen, sample loop, and separation column).
  • Run a blank with each batch of samples by injecting 5 mL of ambient laboratory air into the GC using the same syringe used to inject headspace gas samples.
  • All calibrated values other than O2 and N2 should be nondetectable in the blank. If aberrant peaks appear, bake the column for 8 hr and repeat the blank analysis.


Calibration Sample


  • Five to seven levels of calibration samples (standard gases) are used to create a calibration curve, which is saved with the measurement data (see Instrument Calibration/Calibration Verification).
  • Correlation coefficient values for calibration curves should be 0.99 or better, except O2 and N2, which should be 0.95 or better.


Calibration Verification (CV) Sample


  • Select one of the 5–7 calibration samples from the calibration curve for the calibration verification sample.
  • Run a CV sample at least every shift that samples are taken (see Instrument Calibration/Calibration Verification).
  • The CV should fall within 3% of the calibrated value; O2 and N2 should be within 10% of the calibrated value.


Control Limits


For a system to be considered in control, all QA/QC samples (blank and calibration verification) must be in control.


In Control


A QA/QC sample is in control when the sample analysis result is within a certain tolerance of acceptable limits (see above). Calibration verification samples should be within acceptable limits of the actual value calculated against the calibration curve (see Calibration Verification (CV) Sample) and blanks should be within acceptable limits of background levels of headspace hydrocarbons and gases (see Blank). When the system is in control, as indicated by acceptable results on QA/QC samples, analytical results for unknown samples are considered to be reliable.


Out of Control


If the control limits are exceeded, the instrument system is considered out of control and all samples in the current analytical batch are invalid and must be rerun after the system is proved to be in control.


LIMS Integration


LIMS Components


Analysis

Component

Unit

Description

GC3

dat_asman_id

Serial number of chromatographic data file in digital asset management database (ASMAN)

 

dat_filename

File name of chromatographic data file containing measurements

 

run_test

Test number of related calibration or QA/QC test

 

propene

ppmv

Relative concentration of propene in the sample

 

propane

ppmv

Relative concentration of propane in the sample

 

ethene

ppmv

Relative concentration of ethene in the sample

 

ethane

ppmv

Relative concentration of ethane in the sample

 

methane

ppmv

Relative concentration of methane in the sample

GC3_QAQC

dat_asman_id

Serial number of chromatographic data file in ASMAN

 

dat_filename

File name of chromatographic data file containing measurements

 

run_test

Test number of related calibration or QA/QC test

 

propene

ppmv

Relative concentration of propene in the sample

 

propane

ppmv

Relative concentration of propane in the sample

 

ethene

ppmv

Relative concentration of ethene in the sample

 

ethane

ppmv

Relative concentration of ethane in the sample

 

methane

ppmv

Relative concentration of methane in the sample

GC3_QCAL

mtd_asman_id

Serial number of chromatographic method in ASMAN

 

mtd_filename

File name of the chromatographic method file containing measurements

 

ethene_corr2

R2

Ethene calibration coefficient

 

ethene_intercept

Intercept of ethene calibration curve

 

ethene_slope

Slope of ethene calibration curve

 

ethane_corr2

R2

Ethane calibration coefficient

 

ethane_intercept

Intercept of ethane calibration curve

 

ethane_slope

Slope of ethane calibration curve

 

propene_corr2

R2

Propene calibration coefficient

 

propene_intercept

Intercept of propene calibration curve

 

propene_slope

Slope of propene calibration curve

 

propane_corr2

R2

Propane calibration coefficient

 

propane_intercept

Intercept of propane calibration curve

 

propane_slope

Slope of propene calibration curve

 

methane_corr2

R2

Methane calibration coefficient

 

methane_intercept

Intercept of methane calibration curve

 

methane_slope

Slope of methane calibration curve

NGAFID

dat_asman_id

Serial number of chromatographic data file in ASMAN

 

dat_filename

File name of chromatographic data file containing measurements

 

run_test

Test number of related calibration or QA/QC test

 

iso_butane

ppmv

Concentration of iso_butane in a sample

 

iso_heptane

ppmv

Concentration of iso_heptane in a sample

 

iso_hexane

ppmv

Concentration of iso_hexane in a sample

 

iso_pentane

ppmv

Concentration of iso_pentane in a sample

 

n_butane

ppmv

Concentration of n_butane in a sample

 

n_heptane

ppmv

Concentration of n_heptane in a sample

 

n_hexane

ppmv

Concentration of n_hexane in a sample

 

n_pentane

ppmv

Concentration of n_pentane in a sample

 

ethane_ethene

ppmv

Concentration of ethane + ethene in a sample

 

propane_propene

ppmv

Concentration of propane + propene in a sample

 

methane

ppmv

Concentration of methane in a sample

NGAFID_QA

dat_asman_id

Serial number of chromatographic data file in ASMAN

 

dat_filename

File name of chromatographic data file containing measurements

 

run_test

Test number of related calibration or QA/QC test

 

iso_butane

ppmv

Concentration of iso_butane in a sample

 

iso_heptane

ppmv

Concentration of iso_heptane in a sample

 

iso_hexane

ppmv

Concentration of iso_hexane in a sample

 

iso_pentane

ppmv

Concentration of iso_pentane in a sample

 

n_butane

ppmv

Concentration of n_butane in a sample

 

n_heptane

ppmv

Concentration of n_heptane in a sample

 

n_hexane

ppmv

Concentration of n_hexane in a sample

 

n_pentane

ppmv

Concentration of n_pentane in a sample

 

ethane_ethene

ppmv

Concentration of ethane + ethene in a sample

 

propane_propene

ppmv

Concentration of propane + propene in a sample

 

methane

ppmv

Concentration of methane in a sample

NGAFID_QC

mtd_asman_id

Serial number of chromatographic method in ASMAN

 

mtd_filename

File name of the chromatographic method file containing measurements

 

iso_butane_corr2

R2

Iso-butane calibration coefficient

 

iso_butane_intercept

Intercept of iso-butane calibration curve

 

iso_butane_slope

Slope of iso-butane calibration curve

 

iso_heptane_corr2

R2

Iso-heptane calibration coefficient

 

iso_heptane_intercept

Intercept of iso-heptane calibration curve

 

iso_heptane_slope

Slope of iso-heptane calibration curve

 

iso_hexane_corr2

R2

Iso-hexane calibration coefficient

 

iso_hexane_intercept

Intercept of iso-hexane calibration curve

 

iso_hexane_slope

Slope of iso-hexane calibration curve

 

iso_pentane_corr2

R2

Iso-pentane calibration coefficient

 

iso_pentane_intercept

Intercept of iso-pentane calibration curve

 

iso_pentane_slope

Slope of iso-pentane calibration curve

 

n_butane_corr2

R2

n-butane calibration coefficient

 

n_butane_intercept

Intercept of n-butane calibration curve

 

n_butane_slope

Slope of n-butane calibration curve

 

n_heptane_corr2

R2

n-heptane calibration coefficient

 

n_heptane_intercept

Intercept of n-heptane calibration curve

 

n_heptane_slope

Slope of n-heptane calibration curve

 

n_hexane_corr2

R2

n-hexane calibration coefficient

 

n_hexane_intercept

Intercept of n-hexane calibration curve

 

n_hexane_slope

Slope of n-hexane calibration curve

 

n_pentane_corr2

R2

n-pentane calibration coefficient

 

n_pentane_intercept

Intercept of n-pentane calibration curve

 

n_pentane_slope

Slope of n-pentane calibration curve

 

ethane_ethene_corr2

R2

Ethane + ethene calibration coefficient

 

ethane_ethene_intercept

Intercept of ethane + ethene calibration curve

 

ethane_ethene_slope

Slope of ethane + ethene calibration curve

 

propane_propene_corr2

R2

Propane + propene calibration coefficient

 

propane_propene_intercept

Intercept of propane + propene calibration curve

 

propane_propene_slope

Slope of propane + propene calibration curve

NGAFID_QC

methane_corr2

R2

Methane calibration coefficient

 

methane_intercept

Intercept of methane calibration curve

 

methane_slope

Slope of methane calibration curve

NGATCD

dat_asman_id

Serial number of chromatographic data file in ASMAN

 

dat_filename

File name of chromatographic data file containing measurements

 

run_test

Test number of related calibration or QA/QC test

 

carbon_dioxide

ppmv

Concentration of carbon dioxide in a sample

 

ethane

ppmv

Concentration of ethane in a sample

 

ethene

ppmv

Concentration of ethene in a sample

 

hydrogen_sulfide

ppmv

Concentration of hydrogen sulfide in a sample

 

methane

ppmv

Concentration of methane in a sample

 

nitrogen

ppmv

Concentration of nitrogen in a sample

 

oxygen

ppmv

Concentration of oxygen in a sample

 

propane

ppmv

Concentration of propane in a sample

 

propene

ppmv

Concentration of propene in a sample




Uploading Data to LIMS


Data are uploaded to LIMS automatically using a process explained in the GC3-NGA Advanced User Guide. If the data do not upload to LIMS, contact the laboratory technician.


Health, Safety, & Environment


Safety


  • The following parts are dangerously hot. Avoid touching these areas and cool completely to room temperature before servicing them:
  • Inlets
  • Oven
  • Detectors
  • Column nuts
  • Be careful when working behind the instrument; during cooldown cycle the oven emits hot exhaust that can cause burns.
  • Do not place temperature-sensitive items (e.g., gas cylinders, chemicals, regulators, and plastic tubing) in the path of the heated exhaust.
  • Insulation around inlets, detectors, and valve box contains refractory ceramic fibers. Avoid inhaling particles and wear personal protective equipment including gloves, safety glasses, and dust/mist respirator when working in these areas.
  • Do not leave flammable gas flows on if GC will be unmonitored for long periods of time (however, leave carrier gas on for column flow).
  • Always operate the instrument with the cover properly installed.


Maintenance & Troubleshooting (HP6890GC)


Troubleshooting


Faults


  • Beeping instrument (cancel beep by pressing Clear on the instrument keyboard)
  • One beep: instrument fault, warning, or shutdown
  • Series of beeps: gas flow cannot reach setpoint and flow will be shut down after 1–2 min
  • Continuous beep: thermal shutdown
  • Blinking setpoint on GC display
  • Control table setpoint blinking: gas flow, valve, or oven shutdown
  • Detector On/Off line blinking: pneumatics or detector failure
  • Instrument screen messages (press Clear to remove message)
  • Caution: configuration problems
  • Error: setpoint out of range or incorrect hardware
  • Popup: shutdown, fault, or warning (see error table)
  • FID will not stay lit
  • Make sure the dessicant in the hydrogen generator is not saturated with water (replace/recharge as necessary).
  • Check water level in hydrogen generator



  • No labels