Operating Instructions

WARNING: CARE MUST BE TAKEN WHEN SEALING AMPULES BECAUSE GLASS REMAINS HOT AFTER SEALING

  1. Ensure valves are shut
  2. Connect the gas and compressed air and set pressures
  3. Open the gas and air valve and light flame
  4. Adjust gas and air to obtain a flame with inner cones of approximately 5 mm and the ends of the flame just overlapping
    • Note: If the flames curl up there is too much gas
  5. Adjust the height of the ampule rest so the flames play onto the narrow part of the neck of the ampule where the seal is to be formed
  6. Place the ampule in the flames and rotate the base of the ampule slowly
  7. When the glass stem starts to glow red, use tweezers to hold the top of the ampule whilst still rotating the base. Draw off the top as the glass softens. Practice will be needed to form a good seal. 