Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Select DAQ > Measurement Editor from the drop-down menu. The Measurement Editor window will open (Figure 3).


    Figure 3. Measurement Editor Window.

  2. Set the desired sample interval, number of points to average, leader/trailer length, and measurement speed.
  3. If the sample interval has changed, you will likely receive a warning (DAQ Valid? indicator). When the DAQ Valid? indicator is red, the user must measure the tray background.
  4. Click OK to exit the Measurement Editor window.

...