Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Figure 59 - Motion Profiles Window


GOSCAN Hardware Setup for IMS

To use the GOSCAN Flat Panel Detector with IMS it is necessary to make the following hardware configuration in the Window’s Device Manger and in NIMAX.

 

Device Manager

  1. Login as administrator
  2. In the window, select the Intel 1350 Gigabit Network Connection (that is the NIC that the GOSCAN is connected to) (Figure 60).
  3. Right click and select properties.
  4. Click advance tab
  5. Select Jumbo Packet and set value to 9014 Bytes.
  6. Click OK

 

 

 Image Added

Figure 60- Device Manager window and jumbo packet setup

 

NI MAX Configuration:

  1. Open NIMAX

2. Select Device and Interfaces>Network Devices> Teledyne DALSA GoScan 1510HR (Figure 61)

3. Select the Acquisition Attribute tab

4. Set the values as shown in Figure 62.


5. Make sure that Packet Size is 9004.  Should be able to set this value 9014 (Jumbo Packet Size).  From experience, this would cause instability in the acquisition.


6. Click Save.

 Image Added


Figure 61- NIMAX device manager


Image Added

Figure 62- Acquisition Attributes window.  Set the packet size to 9004.



 

 

Programmer’s Note: the value “1727014677” is the alias name used by the LabVIEW code:

 

NIMAX summary

Most values shown are the default values.  Those highlight are critical to the IMS acquisition.

MAX Configuration Report

3/19/2019 6:12:21 PM

System Settings

Hostname

PC53236

DNS Name

PC53236.ship.iodp.tamu.edu

Vendor

HP

Model

HP Z240 SFF Workstation

Serial Number

2UA7121YXV

Firmware Version

N51 Ver. 01.52

Hardware Revision


Operating System

Microsoft Windows 10 Enterprise

System Start Time

3/4/2019 7:24:09 PM

Description

PC53236 - 64 bit Workstationt Image

System Configuration Web Access

Local Only

System Resources

Total Physical Memory

31.8 GB

Free Physical Memory

26.2 GB

Total Virtual Memory

36.5 GB

Free Virtual Memory

30.7 GB

Primary Disk Capacity

476 GB

Primary Disk Free Space

130 GB

CPU Model

Intel(R) Core(TM) i7-6700 CPU @ 3.40GHz

 

Teledyne DALSA GoScan 1510 HR "1727014677"

Settings

Name

1727014677

Hostname

169.254.5.141

Vendor

Teledyne DALSA

Model

Teledyne DALSA GoScan 1510 HR

Serial Number

DC60A1378609

 

Attribute Categories

Attribute Tables

AcquisitionAttributes

Property

Value

BitsPerPixel

Use hardware value

ImageDecoderCopyMode

Auto

IncompleteBufferMode

Ignore

OutputImageType

Grayscale (U16)

OverwriteMode

Get Newest

PacketSize

9004

ReceiveTimestampMode

System Time

ShiftPixelBits

false

SwapPixelBytes

false

Timeout

5000


AcquisitionAttributes    AdvancedEthernet    BandwidthControl

Property

Value

ActualPeakBandwidth

1000.00000

DesiredPeakBandwidth

1000.00000


AcquisitionAttributes          AdvancedEthernet            Controller

Property

Value

DestinationMode

Unicast

DestinationMulticastAddress

239.192.0.1


AcquisitionAttributes   AdvancedEthernet   EventParameters

Property

Value

MaxOutstandingEvents

50


AcquisitionAttributes   AdvancedEthernet  FirewallTraversal

Property

Value

Enabled

true

KeepAliveTime

30

Supported

true


AcquisitionAttributes  AdvancedEthernet  ResendParameters

Property

Value

MaxResendsPerPacket

25

MemoryWindowSize

1024

MissingPacketTimeout

2

NewPacketTimeout

100

ResendBatchingPercentage

10

ResendResponseTimeout

2

ResendThresholdPercentage

5

ResendTimerResolution

1

ResendsEnabled

true


AcquisitionAttributes  AdvancedEthernet  TestPacketParameters

Property

Value

MaxTestPacketRetries

1

TestPacketEnabled

true

TestPacketTimeout

250


AcquisitionAttributes                                       AdvancedGenicam

Property

Value

CommandTimeout

100

EventsEnabled

true

EventsSupported

true

IgnoreCameraValidationErrors

false



AcquisitionAttributes       ChunkDataDecoding

Property

Value

ChunkDataDecodingEnabled

true

MaximumChunkCopySize

64



CameraAttributes     deviceInformation

Property

Value

DeviceVendorName

Teledyne DALSA

DeviceModelName

GoScan 1510 HR

DeviceID

1727014677

DeviceVersion

209.21

HardwareRevision

IL1.0

TimingVersion

12.119

deviceMacAddress

DC:60:A1:37:86:09

DeviceUserID

1727014677

UserSetDefaultSelector

Factory Setting

UserSetSelector

Factory Setting

UserSetLoad


UserSetSave


DeviceTemperature

50.0000000


CameraAttributes       detectorControl

Property

Value

SettingsRevision

2

Standby

Off

ReadOutMode

Full Resolution

ReadOutDirectionY

Bottom to Top

ReadOutDirectionX

Right to Left

TriggerMode

Free Running

SoftwareTrigger


SoftwareTrigIntTime

50000

ExtendedExposure

10000

Offset

1.95%

FullWell

High Full Well

ROIStartH

0

ROIStopH

1031

ROIStartV

0

ROIStopV

1549


CameraAttributes                          Diagnose

Property

Value

ImageSourceSelector

Sensor Data

SensorDigitalTestPatternValue

0

TestImageIncrement

1

PowerCheck3V5DIG

false

PowerCheck5V4

false

PowerCheck1V2

false

PowerCheck1V5

false

PowerCheck2V5

false

PowerCheck3V3

false

PowerCheck3V5ANA

false

PowerCheck1V8

false

ResetHighReadbackAbs

3.49500000

SelectHighReadbackAbs

3.49375000

VrReadbackAbs

3.19875000

NWellReadbackAbs

3.49375000

AdcRefReadbackAbs

2.79875000

AnaT2ReadbackAbs

0.0387500000

PSupReadbackAbs

11.4541485

Opamp5VReadbackAbs

4.99875000

RefreshDiagnoseBits


transferQueueCurrentBlockCount

0


CameraAttributes                                  ImageFormatControl

Property

Value

PixelPacking

Mono16

PixelFormat

Monochrome 16-Bit Packed

PixelCoding

Mono

PixelSize

16 Bits/Pixel

Width

1032

Height

1550

Mono12PackedBitSelection

LSB


CameraAttributes                                  ImageProcessingControl

Property

Value

SUMEnable

Off

SUMNofFrame

1 Frame



CameraAttributes >> TransportLayerControl

Property

Value

deviceUPnPDiscoveryMode

Active

GevLinkSpeed

1000

devicePacketResendBufferSize

12.9375000

GevIPConfigurationStatus

DHCP

GevPrimaryApplicationIPAddress

169.254.116.129

deviceCCP

Control Access

GevDiscoveryAckDelay


GevGVCPHeartbeatDisable



CameraAttributes >> FileAccessControl

Property

Value

FileSelector

Firmware

FileOperationSelector

Open

FileOperationExecute


FileOpenMode

Write

FileAccessOffset


FileAccessLength

4

FileOperationStatus

Success

FileOperationResult

0

FileSize

0


CameraInformation

Property

Value

BaseAddress

0x0

BusType

Ethernet

HostIPAddress

169.254.116.129

HostMACAddress

00:0A:CD:2F:13:8C

HostNetworkInterfaceName

{1E2FC758-4010-4217-A057-038CD47B344A}

IPAddress

169.254.5.141

ModelName

GoScan 1510 HR

PrimaryURLString

Local:TeledyneDALSA_CD45M21_02_2457f43f_1CA40.0000.xml;b000;31a8e

SecondaryURLString

http://www.teledynedalsa.com/public/mv/software/genicam/GenICamFile.xml

SerialNumberHigh

0xDC60

SerialNumberLow

0xA1378609

VendorName

Teledyne DALSA


StatusInformation

Property

Value

AcqInProgress

false

HandledEventCount

0

IncompleteBufferCount

0

LastBufferCount

0

LastBufferNumber

4294967292

LostBufferCount

0

LostEventCount

0

LostPacketCount

0

ReceivedResendPackets

0

RequestedResendPacketCount

0