Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

When the imaging is complete, IMS will automatically process the raw images and output them to the AUX_DATA folder.  These are not the primary processed images, but are available for quick image evaluation.  For more information on IMS image processing, see the Automatic Image Processing section.

 

X-Rays OFF

Figure # 19 will be displayed when the X-ray Off command is sent to the CP160CP120.  The window will be displayed until the kV has been confirmed to be at zero.

You will also see this window flash on during X-Ray START when the CP160 CP120 fails to respond to the start command.  The X-ray off will reset communications with the CP160 and bring it to a known state which will clear any communication issue.

...

2. No images appear after selecting Process Images: The bit size or pixel length are not set correctly.  Check the command window for error messages and check the settings in the processing settings window.

 

File Formats and File Uploading

IMS automatically outputs the raw and processed images.  Both file types are .tif images.

...