12. Click PUSHER END SEQUENCE.  The source will ramp back up and the motion control system will continue to push the section through the x-ray system.

NOTE: If the detector grid pattern becomes visible in the images ask the Physical Properties technitian  to recalibrate the detector. 

13. When the imaging is complete, IMS will automatically process the raw images and output them to the C:\AUX_DATA folder. The raw images output to the C:\DATA\IN folder. For more information on IMS image processing, see the 'Automatic Image Processing' section in the User Guide.

...

14. Each thirty minutes raw an processed images will be copied automatiacally to DATA(\\CLEVELAND)(S:)\data1\7.3 Petrophysics_XRAY\mirror. Any change on the X-Ray computer will produce the same change on 'data1'.


  • ABORT button

  1. During imaging, user can click ABORT button to end the imaging process and ramp down the source.

...

  1. Once pressed, imaging will end and

...

  1. it will

...

  1. be necessary to start imaging the section from the beginning.

...

  1. A dialog window with different options will appear (Figure 9).


Image Added

Figure 9.- Abort by User dialog box. 


NOTE: The are also two Emergency Stops on the XMSL track, one for stopping the movement of the track and the other for shut down the X-Ray source.