The IODP x-ray system is composed of a 120 kV, 1 mA constant potential x-ray source and a detector unit. The source is a Teledyne ICM CP120B portable x-ray generator with a 0.8 by 0.5 mm focal spot. The beam angle is 50 by 50 degrees generating a directional cone onto the detector, which is distanced 65 cm from the source. The detector is a Go-Scan 1510 HR unit composed of an array of CMOS sensors with an active area of 102 by 153 mm and a resolution of 99 microns.
Manual Configuration
You must manually configure Teledyne ICM CP120B prior to closing the shielding. This needs to be done whenever you replace the source. From the control panel on the source do:
- Insert Key and turn on power. Wait for system to initialize.
- Press Menu
- Press either the plus or minus buttons unit you see "Settings"
- Press Enter and then press the plus key once for each configuration
- Auto Config = Manual Mode
- Keep Alive = Disabled
- English (don't care)
- Fil(iment) Always On = Enable
- Bluetooth = some code (ignore)
- Trig(ger) = Security Enable
- Com Status = Enable
- Baud Rate = 9600
- Long Time = Enable (don't care)
- Menu Step = Quick (don't care)
- Factory = Disabled
- Back Light = Enable (don't care)
- External Trig = Disabled
- Start Button = Disabled
- Pre-warning = 0 sec
- Press Escape
Shielding
The shielding consists of four sections as shown in Figure 1. The Source Shield houses the x-ray source and detector. It is constructed with 6-mm of lead with an internal baffle on the unload side of the box. The exterior corners are covered with 6-mm lead strips and lead vinyl (½-mm) covering the sides. The GRA-MS shield section is constructed with 2-mm of lead with an internal baffle on the load side.
...
3/19/2019 6:12:21 PM
System Settings
Hostname | PC53236 |
DNS Name | |
Vendor | HP |
Model | HP Z240 SFF Workstation |
Serial Number | 2UA7121YXV |
Firmware Version | N51 Ver. 01.52 |
Hardware Revision | |
Operating System | Microsoft Windows 10 Enterprise |
System Start Time | 3/4/2019 7:24:09 PM |
Description | PC53236 - 64 bit Workstationt Image |
System Configuration Web Access | Local Only |
System Resources
Total Physical Memory | 31.8 GB |
Free Physical Memory | 26.2 GB |
Total Virtual Memory | 36.5 GB |
Free Virtual Memory | 30.7 GB |
Primary Disk Capacity | 476 GB |
Primary Disk Free Space | 130 GB |
CPU Model | Intel(R) Core(TM) i7-6700 CPU @ 3.40GHz |
Teledyne DALSA GoScan 1510 HR "1727014677" |
Settings
Name | 1727014677 |
Hostname | 169.254.5.141 |
Vendor | Teledyne DALSA |
Model | Teledyne DALSA GoScan 1510 HR |
Serial Number | DC60A1378609 |
Attribute Categories
Attribute Tables
AcquisitionAttributes | |
Property | Value |
BitsPerPixel | Use hardware value |
ImageDecoderCopyMode | Auto |
IncompleteBufferMode | Ignore |
OutputImageType | Grayscale (U16) |
OverwriteMode | Get Newest |
PacketSize | 9004 |
ReceiveTimestampMode | System Time |
ShiftPixelBits | false |
SwapPixelBytes | false |
Timeout | 5000 |
AcquisitionAttributes AdvancedEthernet BandwidthControl | |
Property | Value |
ActualPeakBandwidth | 1000.00000 |
DesiredPeakBandwidth | 1000.00000 |
AcquisitionAttributes AdvancedEthernet Controller | |
Property | Value |
DestinationMode | Unicast |
DestinationMulticastAddress | 239.192.0.1 |
AcquisitionAttributes AdvancedEthernet EventParameters | |
Property | Value |
MaxOutstandingEvents | 50 |
AcquisitionAttributes AdvancedEthernet FirewallTraversal | |
Property | Value |
Enabled | true |
KeepAliveTime | 30 |
Supported | true |
AcquisitionAttributes AdvancedEthernet ResendParameters | |
Property | Value |
MaxResendsPerPacket | 25 |
MemoryWindowSize | 1024 |
MissingPacketTimeout | 2 |
NewPacketTimeout | 100 |
ResendBatchingPercentage | 10 |
ResendResponseTimeout | 2 |
ResendThresholdPercentage | 5 |
ResendTimerResolution | 1 |
ResendsEnabled | true |
AcquisitionAttributes AdvancedEthernet TestPacketParameters | |
Property | Value |
MaxTestPacketRetries | 1 |
TestPacketEnabled | true |
TestPacketTimeout | 250 |
AcquisitionAttributes AdvancedGenicam | |
Property | Value |
CommandTimeout | 100 |
EventsEnabled | true |
EventsSupported | true |
IgnoreCameraValidationErrors | false |
AcquisitionAttributes ChunkDataDecoding | |
Property | Value |
ChunkDataDecodingEnabled | true |
MaximumChunkCopySize | 64 |
CameraAttributes deviceInformation | |
Property | Value |
DeviceVendorName | Teledyne DALSA |
DeviceModelName | GoScan 1510 HR |
DeviceID | 1727014677 |
DeviceVersion | 209.21 |
HardwareRevision | IL1.0 |
TimingVersion | 12.119 |
deviceMacAddress | DC:60:A1:37:86:09 |
DeviceUserID | 1727014677 |
UserSetDefaultSelector | Factory Setting |
UserSetSelector | Factory Setting |
UserSetLoad | |
UserSetSave | |
DeviceTemperature | 50.0000000 |
CameraAttributes detectorControl | |
Property | Value |
SettingsRevision | 2 |
Standby | Off |
ReadOutMode | Full Resolution |
ReadOutDirectionY | Bottom to Top |
ReadOutDirectionX | Right to Left |
TriggerMode | Free Running |
SoftwareTrigger | |
SoftwareTrigIntTime | 50000 |
ExtendedExposure | 10000 |
Offset | 1.95% |
FullWell | High Full Well |
ROIStartH | 0 |
ROIStopH | 1031 |
ROIStartV | 0 |
ROIStopV | 1549 |
CameraAttributes Diagnose | |
Property | Value |
ImageSourceSelector | Sensor Data |
SensorDigitalTestPatternValue | 0 |
TestImageIncrement | 1 |
PowerCheck3V5DIG | false |
PowerCheck5V4 | false |
PowerCheck1V2 | false |
PowerCheck1V5 | false |
PowerCheck2V5 | false |
PowerCheck3V3 | false |
PowerCheck3V5ANA | false |
PowerCheck1V8 | false |
ResetHighReadbackAbs | 3.49500000 |
SelectHighReadbackAbs | 3.49375000 |
VrReadbackAbs | 3.19875000 |
NWellReadbackAbs | 3.49375000 |
AdcRefReadbackAbs | 2.79875000 |
AnaT2ReadbackAbs | 0.0387500000 |
PSupReadbackAbs | 11.4541485 |
Opamp5VReadbackAbs | 4.99875000 |
RefreshDiagnoseBits | |
transferQueueCurrentBlockCount | 0 |
CameraAttributes ImageFormatControl | |
Property | Value |
PixelPacking | Mono16 |
PixelFormat | Monochrome 16-Bit Packed |
PixelCoding | Mono |
PixelSize | 16 Bits/Pixel |
Width | 1032 |
Height | 1550 |
Mono12PackedBitSelection | LSB |
CameraAttributes ImageProcessingControl | |
Property | Value |
SUMEnable | Off |
SUMNofFrame | 1 Frame |
CameraAttributes >> TransportLayerControl | |
Property | Value |
deviceUPnPDiscoveryMode | Active |
GevLinkSpeed | 1000 |
devicePacketResendBufferSize | 12.9375000 |
GevIPConfigurationStatus | DHCP |
GevPrimaryApplicationIPAddress | 169.254.116.129 |
deviceCCP | Control Access |
GevDiscoveryAckDelay | |
GevGVCPHeartbeatDisable |
CameraAttributes >> FileAccessControl | |
Property | Value |
FileSelector | Firmware |
FileOperationSelector | Open |
FileOperationExecute | |
FileOpenMode | Write |
FileAccessOffset | |
FileAccessLength | 4 |
FileOperationStatus | Success |
FileOperationResult | 0 |
FileSize | 0 |
CameraInformation | |
Property | Value |
BaseAddress | 0x0 |
BusType | Ethernet |
HostIPAddress | 169.254.116.129 |
HostMACAddress | 00:0A:CD:2F:13:8C |
HostNetworkInterfaceName | {1E2FC758-4010-4217-A057-038CD47B344A} |
IPAddress | 169.254.5.141 |
ModelName | GoScan 1510 HR |
PrimaryURLString | Local:TeledyneDALSA_CD45M21_02_2457f43f_1CA40.0000.xml;b000;31a8e |
SecondaryURLString | http://www.teledynedalsa.com/public/mv/software/genicam/GenICamFile.xml |
SerialNumberHigh | 0xDC60 |
SerialNumberLow | 0xA1378609 |
VendorName | Teledyne DALSA |
StatusInformation | |
Property | Value |
AcqInProgress | false |
HandledEventCount | 0 |
IncompleteBufferCount | 0 |
LastBufferCount | 0 |
LastBufferNumber | 4294967292 |
LostBufferCount | 0 |
LostEventCount | 0 |
LostPacketCount | 0 |
ReceivedResendPackets | 0 |
RequestedResendPacketCount | 0 |
Trouble Shooting
Issue | Solution |
---|---|
An error message occurs when the user opens the software indicating an invalid calibration for the dark and white images, but the dark and white images are available on the computer. | Verify that the .ini files are referring to valid calibration images that exist in the correct locations. If the files exist, then check the USB connections to the computer, particularly the USB for the X-ray source. |
Archive Versions
...