  1. Open the Sequence Editor (Error! Reference source not found.) by selecting DAQ > Sequence Editor from the IMS panel menu (Error! Reference source not found.).
  2. Select a sequence action on the left side of the window. In-line and off-line treatments are mutually exclusive. Once a treatment type is selected, the other treatment type will be disabled.
    • Measure only should be used to make an NRM measurement.
    • In-line AF Demag allows users to assign a degaussing step with a field ranging from 1 to 80 mT.

      Figure 14- Offline Treatment Window
    • Off-line Treatment will trigger a dialog box (Error! Reference source not found.) to open during the measurement process prompting the user to perform an offline treatment. The user may enter the type of offline treatment and comments regarding the treatment. The types of treatments available are shown in Error! Reference source not found.. Click Continue Sequence to continue with the measurement and Abort to end the sequence.
  3. Users may copy an existing sequence from a file by clicking the Copy From File button.
    • A new window with a list of existing sequences will be displayed (Error! Reference source not found.).
    • Select an existing sequence and press the Ok…Copy button.
    • Once selected, a sequence will be displayed in the sequence action list and the user may edit the sequence (Error! Reference source not found.).
  4. Click Add to List to place a new step into the sequence action list.

    Image Modified

    Figure 15- Copy Sequence File Window

    • In-line AF Demag steps will be automatically ordered by increasing value.

    • Each in-line AF Demag step automatically has a measurement step.

  5. Once the list is completed, click Ok Save and name the sequence

    • The sequence will be saved with a new name or the user may save over an existing sequence.

Defining the Sample Presets

...