Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

StatusInformation

Property

Value

AcqInProgress

false

HandledEventCount

0

IncompleteBufferCount

0

LastBufferCount

0

LastBufferNumber

4294967292

LostBufferCount

0

LostEventCount

0

LostPacketCount

0

ReceivedResendPackets

0

RequestedResendPacketCount

0

Transferring Images to Data 1

Images are transferred to Data 1 via a script on the XMSL host PC.  The script copies the output from the XMSL to data1 automatically and is executed via a task scheduler every 30 minutes.

The script:

  • Connects to Data:\Data1\7.3 Petrophysics_XRAY directory
  • Mirrors the Data and AUX_DATA directories on the host PC (currently XMSL is PC53236) to a "Mirror" directory via an Xray eDirectory user-account

This is a "one-way" mirror operation. Whatever changes are made to DATA and Aux_DATA directories on the host PC will be mirrored in the location in data1.  To make the files permanent in the Data1 folder, copy the mirror contents to the base folder periodically.  To make the script more efficient it is also recommended to remove the already copied files from the Data and AUX_ Data folders so they are not copied multiple times.

The script will hang if a user logs into the file servers while the script runs.  No one should log in to the servers on this computer unless absolutely necessary.  Task scheduler will kill the task if the script attempts to run for more than an hour and half.

Trouble Shooting

IssueSolution

An error message occurs when the user opens the software indicating an invalid calibration for the dark and white images, but the dark and white images are available on the computer.


Verify that the .ini files are referring to valid calibration images that exist in the correct locations.  If the files exist, then check the USB connections to the computer, particularly the USB for the X-ray source.

...