Adjust the automatic image processing settings (Image Processing section on the User Guide)

NOTE: Ask the PP technitian technician to help with these steps.

...

    Manual: Type in the sample label ID (i.e., 362-U1480A-1H-1-A), LIMS Text ID (i.e., SHLF7851761) and length of section. Use the whole round ID when imaging whole rounds. Click USE ME to update the IDs and enable the MEASURE button. Any entry will be accepted, but that does not guarantee the data will upload. The Text ID information must match an actual section in order for the data        to upload.

...

Figure 5.- Warning message indicating that the source is not ready to ramp up.

NOTE: If this message pops-up and both doors are closed, call the Physical Properties technitiantechnician.


6. Click MEASURE

7. The motion control pusher will move the section under above the sourcedetector

8. When section is over the detector, the x-ray source will ramp up and a window with its status will appear (Figure 6). Once mA and kV are stable, the window will close and image acquisition will begin. User may abort at any time by clicking the ABORT button.

...

10. Source ramps down, a window with the message 'X-RAY OFF'  (Figure 8) will appear and the red light will turn off.

...

12. Click PUSHER END SEQUENCE.  The source will ramp up back up and the motion control system will continue to push the section through the x-ray system.

NOTE: If the detector grid pattern becomes visible in the images ask the Physical Properties technitian  technician  to recalibrate the detector. 

...

14. Each thirty minutes raw an processed images will be copied automatiacally to DATA(\\CLEVELAND)(S:)\data1\7.3 Petrophysics_XRAY\mirror. Any Be careful before delete a image on the X-Ray computer, any change on the X-Ray computer will produce the same change on 'data1'.

...