Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Use the jig labeled "Frost".
  2. Clean jig and Logitech scale by wiping them with a kimwipe and isopropyl alcohol.
    1. Refer to the Logitech LP Set Up user guide for more information on scale use.
  3. Set the scale on the jig gently (Fig. 1) and wiggle the scale.
  4. The jig should be set to 1217 (just before the 17 is better).
    1. Use the middle gauge on the scale.
      1. The small dial (Fig. 1) on the bottom left side should read 12.
      2. The large face (Fig. 1) should read 17.
        1. If the jig is not set, put the Tommy bar into the holes on the main post in the middle of the jig to adjust it (Fig. 1).
          1. Rotating the Tommy bar clockwise will lower the chuck face causing the numbers on the scale to increase.
          2. Rotating the Tommy bar counterclockwise will raise the chuck face, causing the numbers on the scale to decrease.



Image Added Large FaceSmall DialMain PostTommy Bar Image Removed
Figure 1


LAPPING UNFROSTED SLIDES

...