Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

BONDING BILLETS TO FROSTED SLIDES: USER GUIDE



 

 

Author:

E. Fisher

Reviewer(s):

 


Editor(s):

S. Boehm, S. Frank

Management Approval (Name, Title, Date):

B. Mills (Lab Officer)

Audience:

Technicians

Origination date:

4/27/2013

Current version:

7/30/2018

Revised: 


Domain:

Thin Section Lab

System:

Bonding Jig

Keywords:

Bonding Samples

...

oduction

This user guide will help thin section technicians bond the flattened billets to frosted slides. For more information refer to the Logitech Thin Rock Section Production manual, available in the Thin Section Lab both on a Logitech thumb drive in the bookcase and computer under the "Manual" folder on the desktop.

PROCEDURE

PREPARING FOR RESIN

  1. Set the billets flattened side up on heat pad and spray them with isopropyl alcohol.  This will help them dry.
  2. Clean frosted slides off with isopropyl alcohol, blow them dry, and measure them again.

...

          b. Take note of how thick the slides are and write it on the sample tray. The samples will need to be 30 microns on top of                   the of the slide thickness.

      3. Place the slides on the heat pad as well.

MIXING RESIN

  1. Mix one full pump of EpoTek 301 part A and part B into a weighing dish.
    1. The ratio is 3 to 1. The pumps are preset to this.
    2. Make sure pump is primed first or there will not be the right ratio.
  2. Stir the resin slowly with wooden stick until there are no trails of the two mixing (Fig. 1) and go slow to avoid bubbles.



                          Figure 1

Make sure no wooden splinters fall wood particles fall into the resin; pick them out if they do.

The resin should stay good for about 45 minutes. If it starts to get gooeyget too thick, place  place on the hot plate for a few seconds and it will become runny become thin again for a very short time.

PUTTING RESIN ON THE BILLETS

  1. Once the samples are completely dry clean the surface off with a kimwipe kim-wipe and isopropyl alcohol.
  2. Place the billet on a Teflon puck.
  3. With the wooden stick slowly draw an X with the resin on the sample (Fig. 2).



                                   Figure 2

          a. If the first stroke is discolored it means the resin is soaking in the vesicles.  This is one way to tell if surface impregnation is needed.

...

 5. Let it sit for a minute or two and watch if any bubbles form.  If there are bubbles get them out with the metal point tool (Fig. 3).



                                Figure 3

SURFACE IMPREGNATION

  1. If the billet is soaking up resin, it is a good idea to surface impregnate it.
  2. Hook up the vacuum hose to the impregnation stand (Fig. 4).
  3. Place the billet under the bell jar (Fig. 4).
  4. Check the vacuum oil level. Refill if it starts to run low.
  5. Turn on vacuum pump by turning on the power strip connected to the wall (Fig. 4).
    Image Modified
                                                     

...

  1.                                                Figure 4

6. Close the valve on the upper right side of the Logitech VS2 (Fig. 5) and make sure the valve on the impregnation stand is open (parallel with the vacuum hose) (Fig. 4).

                                             Figure                                                             Figure 5

    1. Watch the gauge on the left side; it should be in the green around 25 in/Hg.
    2. Press on the bell jar to get a good seal.

...

13. After the billet is surface impregnated it can be immediately bonded to a frosted slide.
 

BONDING FROSTED SLIDES TO THE BILLETS

  1. Make sure the frosted slides are clean.
  2. Hold the frosted side up to the sample to measure how large it is. Draw two horizontal lines with resin to outline the size of the sample (Fig. 6).


                                Figure 6

3. Fill in the space between the two lines with resin.  Put more resin on the right side.

4. Place the slide with resin side down onto the sample from left to right (Fig. 7).  Lay down very slowly to prevent making bubbles.


Figure                                    Figure 7

5. Press on the middle of the slide moving any bubbles out to the sides (Fig. 8).



Figure                                 Figure 8

     a. If there are lots of bubbles take off the slide. Wipe off the resin and clean the sample off with isopropyl alcohol then start over again.

...

     a. Squeeze the spring and remove the Teflon puck to begin lowering.


                       Figure 9

7.  Let the resin cure for about one hour.

...

9. When cured remove sample and replace Teflon pucks.

GRAIN MOUNTS

  1. Dry the grains on the hot plate for an hour or more in a weighing dish.
  2. Mix up EpoTek resin.
  3. Place dry grains into a clean plastic mold and cover them with resin.
    1. It is best to use a small mold and to fill it at least half way.
  4. Label the mold with the sample number.
  5. Place the mold on the hot plate on top of a Teflon puck.
  6. Let it cure until it is hard all the way through-it may take the whole day.
  7. Peel away the mold and make sure to label the sample.
  8. Once the resin is dry the grain mount can be treated as a hard rock billet.
  9. When flattening the grain mount it is best to do it by hand. Make sure enough material is taken off so the grains are cut in half.
  10. When bonding the grain mount to a slide treat it as a hard rock.

IMPREGNATING A BILLET

  1. Impregnation should be done on samples that are soft, water soluble or vesicular billets (ex. pumas, sandstone, mudstone, and clay stone).
    1. These samples should all be dried first. Refer to the Freeze Drier user guide.
  2. Once the samples are dry mix up 3 pumps of both part A and B of EpoTek resin.
  3. Place the billet into a clean plastic mold that is best fit it.
  4. Label the mold with the sample number.
  5. Cover the billet with resin (Fig. 10).



                                Figure 10

6. Hook up the vacuum hose to the impregnation stand (Fig. 4).

7. Place the mold with the billet under the bell jar (Fig. 11).



                                                                                                                        Figure 11

8. Check the vacuum oil level. Refill if it starts to run low.

...

22. Treat the impregnated billet as a hard rock when bonding it to a slide.

Credit

Original document created Emily Fisher  

4/27/2013

Archive